Table 4 RR average values of macaques detected via ECG and radar.

From: Contactless vital signs monitoring in macaques using a mm-wave FMCW radar

Macaque ID

State

ECG

Radar

Abs. error

Rel. error (%)

16104

Awake

38.67

41.33

2.66

6.88

152024

Awake

25.00

25.67

0.67

2.68

152030

Awake

22.17

20.00

2.17

9.79

152032

Awake

28.83

31.00

2.17

7.53

15092

Awake

26.00

23.67

2.33

8.96

151010

Awake

24.50

21.17

3.33

13.59

151007

Awake

21.67

21.33

0.34

1.57

151009

Awake

22.50

22.17

0.33

1.47

19080

Anesthetized

20.67

20.00

0.67

3.24

19092

Anesthetized

13.00

13.67

0.67

5.15

20005

Anesthetized

26.00

28.67

2.67

10.27

20042

Anesthetized

9.67

10.00

0.33

3.41

20313

Anesthetized

11.00

10.00

1.00

9.09