Table 2 No mock specimens or specimen data tracked by the device were lost or misplaced; A total of 1064 automated transactions and 1501 mock specimen data were handled.
From: A multi-center evaluation of a novel IVF cryostorage device in an active clinical setting
Site | α | β | Γ |
---|---|---|---|
Mock patient cohorts | 56 | 173 | 100 |
Lost or misplaced mock specimens | 0 | 0 | 0 |