Table 1 Overview of the innovative programmable (IPG) electrical stimulation parameters for the pudendal nerve (PN).

From: An innovative electrical neurostimulation approach to mimic reflexive urination control in spinal cord injury models

Serial number

Peak voltage (V)

Frequency (Hz)

Pluse duration (ms)

Latency (ms)

Total time (ms)

Inter-group durations (ms)

1

0.5

800

0.8

0.45

15

150

2

0.75

800

0.8

0.45

22.5

160

3

0.75

800

0.8

0.45

25

185

4

1

800

0.8

0.45

30

180

5

1

800

0.8

0.45

31.25

160

6

1

800

0.8

0.45

27.5

175

7

1

800

0.8

0.45

32.5

165

8

1

800

0.8

0.45

35

160

9

1

800

0.8

0.45

31.25

165

10

1

800

0.8

0.45

37.5

140

11

1

800

0.8

0.45

42.5

180

12

1

800

0.8

0.45

35

180

13

1

800

0.8

0.45

42.5

190

14

1

800

0.8

0.45

37.5

200

15

1

800

0.8

0.45

31.25

170

16

1

800

0.8

0.45

37.5

160

17

1

800

0.8

0.45

35

150

18

1

800

0.8

0.45

37.5

180

19

1

800

0.8

0.45

31.25

160

20

1

800

0.8

0.45

37.5

190

21

1

800

0.8

0.45

35

30

22

0.3

80

0.8

11.7

400

0

23

0.1

80

0.8

11.7

400

0

24

0.03

80

0.8

11.7

300

0

  1. Each IPG stimulation requires the execution of 144 critical parameters in the order specified by the serial numbers, as detailed in the table. Serial numbers, total time, and inter-group durations are adjustable, but the number of serial numbers used must be at least half of those listed. Specifically, the number of serial numbers can be selected or adjusted between 4 and 20.