Table 6 Fatigue analysis of implants with different materials, under impact force.
AL | ZR | CFP60 | TI | CFP30 | PEEK | PE | ||
---|---|---|---|---|---|---|---|---|
Life | Max | 1e10 | 1e10 | 1e10 | 1e10 | 1e10 | 1e10 | 1e10 |
Min | 0 | 1e10 | 1e10 | 1e10 | 1e10 | 0 | 0 | |
Safety Factor | Max | 15 | 15 | 15 | 15 | 15 | 15 | 15 |
Min | 0.413 | 1.46 | 1.037 | 1.44 | 1.02 | 0.388 | 0.213 | |
Equivalent Alternating Stress (MPa) | 700.62 | 449.9 | 361.36 | 289.83 | 117.5 | 146.6 | 70.29 |