Table 1 The estimated values of the circuit components used in fitting of Nyquist plot for all six bacteria culture.
Bacteria concentration (C.F.U/mL) | E. coli | P.F. | L.P. | S.A. | B.C. | B.S. | ||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
 | \(R_{1}\) (M\(\Omega\)) | \(C_{1}\) (\(\mu\)F) | \(C_{pc}\) (\(\mu\)F) | \(R_{ct}\) (M\(\Omega\)) | \(C_{pc}\) (\(\mu\)F) | \(R_{ct}\) (k\(\Omega\)) | W (k\(\Omega\)) | \(C_{pc}\) (\(\mu\)F) | \(R_{ct}\) (k\(\Omega\)) | W (k\(\Omega\)) | \(C_{pc}\) (\(\mu\)F) | \(R_{ct}\) (k\(\Omega\)) | \(C_{pc}\) (\(\mu\)F) | \(R_{ct}\) (k\(\Omega\)) | \(C_{pc}\) (\(\mu\)F) | \(R_{ct}\) (k\(\Omega\)) |
10 | 74.30 | 1.62 | 1.27 | 8.66 | 4.00 | 15.30 | 0.69 | 2.56 | 23.9 | 1.09 | 2.50 | 9.36 | 1.57 | 35.0 | 2.50 | 8.91 |
50 | 56.00 | 2.12 | 1.37 | 8.32 | 4.26 | 14.80 | 0.81 | 2.73 | 23.10 | 1.27 | 7.13 | 10.26 | 1.54 | 35.7 | 7.13 | 8.92 |
250 | 51.6 | 2.34 | 1.36 | 7.60 | 4.77 | 15.2 | 0.82 | 3.05 | 23.7 | 1.29 | 3.39 | 10.79 | 1.53 | 35.30 | 3.39 | 8.82 |
2900 | 41.10 | 2.64 | 1.41 | 6.37 | 5.02 | 15.00 | 0.77 | 3.21 | 23.4 | 1.22 | 6.48 | 11.38 | 2.10 | 31.2 | 6.48 | 7.01 |
5400 | 34.30 | 2.7 | 1.47 | 6.21 | 4.08 | 15.20 | 0.33 | 2.61 | 23.70 | 0.52 | 5.48 | 11.43 | 2.34 | 31.8 | 5.48 | 7.95 |
50400 | 20.30 | 2.92 | 1.56 | 4.76 | 3.98 | 15.20 | 0.33 | 2.57 | 23.90 | 0.53 | 5.57 | 10.89 | 2.36 | 25.8 | 5.57 | 6.57 |
200400 | 9.66 | 3.06 | 1.81 | 2.10 | 3.62 | 15.4 | 0.34 | 2.32 | 23.99 | 0.54 | 5.98 | 10.50 | 2.84 | 21.8 | 5.98 | 5.42 |
500400 | 6.66 | 3.88 | 2.09 | 1.68 | 3.56 | 14.00 | 0.34 | 2.28 | 21.82 | 0.37 | 5.35 | 11.27 | 2.92 | 21.8 | 5.35 | 5.44 |