Table 1 Sensitivity, and quality factor of the different quasi-periodic phcs.
From: Quasi periodic photonic crystal as gamma detector using Poly nanocomposite and porous silicon
structure | Gamma range | S (nm/Gy) | Q |
---|---|---|---|
\(\:2{\text{G}\text{T}\text{M}\left(\text{1,1}\right)}^{2}\) | 0–100 Gy | 0.55 | 77.4 |
100–200 Gy | 0.15 | 76.6 | |
\(\:3\text{G}\text{T}\text{M}\left(\text{1,1}\right)\) | 0–100 Gy | 0.55 | 33.8 |
100–200 Gy | 0.14 | 34.6 | |
\(\:2{\text{G}\text{T}\text{M}\left(\text{2,1}\right)}^{2}\) | 0–100 Gy | 0.55 | 892.2 |
100–200 Gy | 0.16 | 957.4 | |
\(\:3\text{G}\text{T}\text{M}\left(\text{2,1}\right)\) | 0–100 Gy | 0.55 | 1715.9 |
100–200 Gy | 0.16 | 1951.8 | |
\(\:2{\text{G}\text{T}\text{M}\left(\text{1,2}\right)}^{2}\) | 0–100 Gy | 0.50 | 473.0 |
100–200 Gy | 0.14 | 467.9 | |
\(\:3\text{G}\text{T}\text{M}\left(\text{1,2}\right)\) | 0–100 Gy | 0.49 | 414.8 |
100–200 Gy | 0.13 | 365.5 | |
\(\:2{\text{G}\text{T}\text{M}\left(\text{1,3}\right)}^{2}\) | 0–100 Gy | 0.41 | 1805.6 |
100–200 Gy | 0.13 | 2620.3 | |
\(\:3\text{G}\text{T}\text{M}\left(\text{1,3}\right)\) | 0–100 Gy | 0.18 | 1929.3 |
100–200 Gy | 0.09 | 2363.4 |