Table 1 Troubleshooting common issues encountered during tissue dissociation for intestinal crypt isolation via semi-automated system.

From: Automated tissue dissociation for the establishment of human intestinal organoids

Problem

Solution

Biopsy adhesion to Via Extractor

Sachet / pipette tips

Pre-coat plastic containers and equipment with 0.1%

BSA prior to addition of biopsies

Organoids fail to grow from the

dissociation mixture

1. Ensure crypts are fully released and tissue is not

under-dissociated (but avoid over dissociation)

2. Minimise exposure to EDTA solution

3. Maintain dissociation temperature at 4 °C

Organoids show impaired

growth

Adjust dissociation parameters to balance cell yield and viability,

ensuring crypts (not single cells) are isolated