Table 7 Distribution of unsuccessful Trueprep test results from January to December 2022

From: Pilot implementation result of frequency, distribution, and factors associated with unsuccessful tests and error rates at Truenat sites in Nigeria

Variable

Code

Frequency (%)

Cartridge valve error

E1

65 (6.4)

Cartridge error

E2

450 (44.1)

Cartridge clogged

E3

219 (21.4)

Cartridge not loaded

E6

66 (6.5)

Reset card error

E9

16 (1.6)

Invalid reset card

E10

22 (2.2)

Device heater plates not working

E11

19 (1.9)

Device heater plates not working

E12

2 (0.2)

Self-test failed

E13

13 (1.3)

Cartridge error –Elution

E14

54 (5.3)

Cartridge clogged-elution

E15

95 (9.3)

Total Trueprep errors

 

1021 (100)