Table 1 Overview of mean lesion dimensions across the evaluated contact force range (5–30 g), comparing the very high-power short-duration and high-power short-duration protocols.
Lesion metrics | vHP-SD 90 W/4 s (n = 1201) | HP-SD-4 50 W/4 s (n = 1201) | HP-SD-15 50 W/15 s (n = 1201) | p-value vHP-SD vs. HP-SD-4 | p-value vHP-SD vs. HP-SD-15 | p-value HP-SD-4 vs. HP-SD-15 |
---|---|---|---|---|---|---|
Lesion depth1 (mean ± SD) [mm] | 3.16 ± 0.41 | 2.42 ± 0.61 | 4.49 ± 0.66 | < 0.001 | < 0.001 | < 0.001 |
Maximum lesion diameter1 (mean ± SD) [mm] | 7.34 ± 0.92 | 6.02 ± 1.00 | 9.13 ± 1.59 | < 0.001 | < 0.001 | < 0.001 |
Surface diameter1 (mean ± SD) [mm] | 4.79 ± 0.95 | 4.08 ± 0.93 | 4.45 ± 1.02 | < 0.001 | 0.008 | 0.004 |
Lesion volume1 (mean ± SD) [mm3] | 96.4 ± 28.8 | 52.0 ± 22.5 | 211.4 ± 95.3 | < 0.001 | < 0.001 | < 0.001 |
Mean power output1 (mean ± SD) [mm3] | 59.1 ± 15.6 | 37.5 ± 3.4 | 40.5 ± 7.0 | < 0.001 | < 0.001 | < 0.001 |