Fig. 6 | Scientific Reports

Fig. 6

From: Evaluation of Clareon Vivity and PureSee intraocular lenses: optical quality, depth of focus and misalignment effects

Fig. 6

The through-focus modulation transfer function at 25, 50, and 100 lp/mm measured at on- (colored lines) and off-axis (black lines). The dotted lines show the values of each lens separately; the solid lines refer to the average of two samples. Decentration lowered the optical performance as determined by the primary peak for both models, and impacted the EDoF properties, especially for Clareon Vivity at 3-mm aperture at 25 lp/mm.