Table 3 Validation parameters for the detection of CO using GC-TCD.

From: Advancing forensic accuracy: mitigating methemoglobin interference in postmortem carbon monoxide analysis using sodium dithionite reduction

Sample

LOD (%)

LOQ (%)

RSD (%)

R2

Recovery (%)

Spleen A

7.84

12.95

2.66

0.9992

94.91

Spleen B

6.29

8.29

1.86

0.9991

96.73

Blood

4.2

5.27

1.77

0.9959

102.7