Table 3 Summary of serial cases.

From: Efficacy and safety of a novel flow-disruptor device in a rabbit aneurysm model: a preliminary study

No.

Width (mm)

Height (mm)

Neck (mm)

Size of NFD (mm)

Follow-up duration (months)

Early post-implantation in DSA (grade)

Follow-up occlusion in DSA (grade)

Protrusion

1

2.97

3.36

2.19

5 × 2.5

1

1

1

Yes

2

2.81

6.35

3.20

5 × 2.5

1

3

3

No

3

3.30

3.46

3.61

5 × 2.5

1

2

2

Yes

4

3.58

7.83

2.77

6 × 3

1

4

3

No

5

3.14

6.36

4.66

5 × 2.5

1

1

1

Yes

6

4.19

6.45

4.43

6 × 3

1

4

2

No

7

4.96

10.25

5.16

6 × 3

1

2

2

Yes

8

3.14

4.50

3.45

5 × 2.5

2

3

2

No

9

3.31

10.06

3.66

6 × 3

2

2

2

Yes

10

3.60

6.51

5.24

5 × 2.5

2

3

3

No

11

4.74

10.05

4.46

6 × 3

2

1

1

Yes

12

2.96

7.14

3.63

6 × 3

2

3

2

No

13

3.74

6.76

4.46

5 × 2.5

2

4

3

No

14

3.38

8.05

2.02

6 × 3

2

2

2

Yes

15

3.95

9.08

3.30

6 × 3

3

4

2

No

16

4.58

10.16

4.85

6 × 3

3

3

2

No

17

3.08

6.84

4.25

5 × 2.5

3

2

1

Yes

18

3.90

10.23

3.99

6 × 3

3

4

3

No

19

3.88

7.18

3.01

6 × 3

3

2

2

Yes

20

3.34

7.50

4.38

6 × 3

3

1

1

Yes

21

2.95

5.71

4.44

5 × 2.5

3

4

3

No

  1. NFD, novel flow disruptor; DSA, digital subtraction angiography.