Table 1 Bactericidal activity of NTP-generating devices (Laboratory prototypes 1 and 2 and Plasmatico v1.0) against Pseudomonas aeruginosa PAO1 and methicillin resistant Staphylococcus aureus (MRSA) ATCC 43300 (clinical isolate).
NTP exposure time (min) | Laboratory prototype 1 | Laboratory prototype 2 | Plasmatico v1.0 |
---|---|---|---|
CFU/mL | CFU/mL | CFU/mL | |
Pseudomonas aeruginosa PAO1 | |||
0 | 5 × 107 | 5 × 107 | 5 × 107 |
15 | n.q. | n.q. | n.q. |
30 | 24 | n.q | n.q |
60 | 0 | 0 | 0 |
90 | 0 | 0 | 0 |
Methicillin-resistant Staphylococcus aureus ATCC 43300 (clinical isolate) | |||
0 | 2 × 107 | 2 × 107 | 2 × 107 |
15 | n.q. | n.q. | n.q. |
30 | 272 | n.q. | n.q. |
60 | 1 | 3 | 60 |
90 | 0 | 0 | 1 |