Table 2 A 2 × 2 contingency table showing the performance characteristics PIMA POC at the ≤ 200 CD4 cells/µl threshold for HIV using the results of the BD facscalibur as the reference.
From: Evaluation of point-of-care Alere Pima CD4 test for low CD4 cell counts test in Uganda
FACSCalibur (flowcytometry) | |||||
---|---|---|---|---|---|
≤ 200 cells/µL | > 200 cells/µL | ||||
PIMA POC | ≤ 200 cells/µL | 5 | 2 | 7 | PPV (71.43), LHR+(28.06) |
> 200 cells/µL | 4 | 99 | 103 | NPV (96.12), LHR-(0.45) | |
9 | 101 | 110 | |||
Sensitivity (55.56%) | Specificity (98.02%) |