Table 3 Lin’s concordance correlation coefficient for repeatability and the total Gage R&R for all examined eyes, healthy samples and diseased samples.

From: Manual measurement of SD-OCT images of hyperreflective retinal layers in acute retinal vessel occlusion is reliable, repeatable and reproducible

Layer

All samples

Healthy samples

Diseased samples

Lin’s CCC

Gage R&R

Lin’s CCC

Gage R&R

Lin’s CCC

Gage R&R

as %

VB

99.12

8.91

94.13

21.85

99.32

7.42

IMRL

99.31

7.86

98.12

12.33

89.90

28.81

MRL

96.90

16.74

84.40

36.07

96.41

17.06

IRL

98.43

11.90

97.18

15.10

90.43

28.09

ORL

97.11

16.14

75.39

45.41

97.80

13.31