Table 2 Method performance and validation data.

From: Nanoparticle modified paper-based analytical sensor for calcium determination in human nasal secretions and its association with olfactory dysfunction

Parameter

Value

Linearity range, (M)

1.0 × 10–71.0 × 10–1

Slope, (mV/decade)

29.14 ± 0.3

Correlation coefficient, r2

0.9998

pH range

2–9

Response time, (s)

20

LOD, (M)

7.5 × 10–8

Shelf life, (d)

180

Accuracy, % R

96.85

Precision, %RSD

Repeatability

2.79

Intermediate precision

2.44

Selectivity coefficients,\({\text{k}}_{{\text{a},\text{b }}^{\text{pot}}}\)

Sodium chloride

0.022041

Potassium chloride

0.047267

Magnesium chloride

0.411651

Sodium phosphate

0.048825

Zinc chloride

0.000336

Sodium carbonate

0.001465