Table 4 Comparative assessment of the proposed paper-based sensor and previously reported sensors.

From: Nanoparticle modified paper-based analytical sensor for calcium determination in human nasal secretions and its association with olfactory dysfunction

Sensor characters

Proposed method

Published method 27

Published method 28

Published method 29

Design state

Paper sensor modified with MWCNT and carbon dots

Coated carbon brush electrode modified with azomethine)

Solid-state electrode modified with carbon nanotube

Paper sensor modified with nano-optodes

Linearity range, (M)

1.0 × 10–71.0 × 10–1

1.0 × 10–61.0 × 10–1

1.0 × 10–53.2 × 10–3

5.0 × 10–55.0 × 10–3

LOD, (M)

7.5 × 10–8

1.90 × 10–6

6.3 × 10–7

5.0 × 10–5

Slope, (mV/decade)

29.14

15.09

29

Colorimetric readout

pH

2–9

4–8

Neutral sap

Neutra/tap water conditions

Response time, (s)

20

8

< 20

Immediate (naked-eye detection)

Special features

High sensitivity, wide linearity

Simple electrode but lower sensitivity

Miniaturized, stable, direct use in real sap

Low-cost, portable, equipment free