Table 6 PTV margins for Varian and Elekta linear accelerators that could cover the gross tumor volume with 95% probability at 95% of the institutions, determined assuming that the distance between the imaging and radiation isocenters is zero owing to rigorous QA implementation and that the intra-fraction error caused by treatment couch rotation is zero owing to implementation of a positional monitoring system such as surface image guidance.
Manufacturer | Direction | Reduced PTV margin size [mm] | |
---|---|---|---|
Ong | Zhang | ||
Varian | X | 1.0 | 1.5 |
Y | 1.0 | 1.1 | |
Z | 1.0 | 2.0 | |
Elekta | X | 1.5 | 1.9 |
Y | 1.5 | 1.6 | |
Z | 1.5 | 2.3 |