Figure 7 | Scientific Reports

Figure 7

From: Detector development for spin-echo SANS techniques using ZnS:Ag/6LiF and 6Li glass scintillators

Figure 7

Measured rate at 1Å  as a function of the expected rate on one strip of the multi-strip GS20 detector (in red) and with the WLSF scintillator detector (in blue). The maximum peak rate for one strip of the GS20 detector is approximately 300 kcps with a 2\(\%\) deviation from the expected rate. The linearity condition is satisfied up to 60 kcps for the WLSF scintillator detector.

Back to article page