Fig. 9
From: A novel AI-coupled flow chamber method quantifying erythrocyte osmotic fragility

Parametric Bland–Altman plot comparing MCF₅₀ values obtained from the flow chamber and classical osmotic fragility methods for AQP inhibited and LPS incubated RBCs (a) and (b), respectively (single subject, n = 6 repeated, paired measurements). The y-axis represents the difference in MCF₅₀ values (Flow Chamber-Classical OF), while the x-axis shows the average of the two methods. The solid red line indicates the median bias (− 0.030 for AQP inhibited, − 0.027 for LPS incubated RBCs), and the blue lines represent the upper and lower LoA. The majority of points fall within this range, indicating an encouraging agreement between the two methods.