Table 2 Confusion matrix showing results for all analysed 90-s windows based on the CW2’s PPG algorithm in comparison to a Holter reference across 24 h

From: Ambulatory atrial fibrillation detection and quantification by wristworn AI device compared to standard holter monitoring

 

Holter: AF

Holter: non-AF/NSR

Total

CW2-PPG: AF

3898

934

4832

CW2-PPG: NSR

65

51884

51949

Total

3963

52818

56781