Table 8 Diagnostic efficiency of eight morphological parameters of the central anterior chamber section for NACA using the Dataset-3

From: Intelligent screening of narrow anterior chamber angle based on portable slit lamp

Light band

Total

Success

Parameter

Unit

Accuracy

Sensitivity

Specificity

F1

Horizontal

and

upper

349

(eyes)

309

(88.54%)

I

Pixel

0.74

0.78

0.71

0.70

II

Pixel

0.75

0.79

0.72

0.71

III

Ratio

0.67

0.86

0.54

0.67

IV

Pixel

0.71

0.77

0.68

0.68

V

Pixel

0.74

0.76

0.72

0.69

VI

Degree

0.74

0.75

0.73

0.69

VII

Ratio

0.65

0.64

0.66

0.59

VIII

Pixel

0.64

0.62

0.65

0.57

Horizontal

and

lower

349

(eyes)

312

(89.40%)

I

Pixel

0.74

0.89

0.65

0.73

II

Pixel

0.74

0.87

0.66

0.73

III

Ratio

0.7

0.82

0.62

0.68

IV

Pixel

0.68

0.74

0.63

0.64

V

Pixel

0.74

0.81

0.69

0.71

VI

Degree

0.76

0.88

0.69

0.74

VII

Ratio

0.62

0.65

0.6

0.58

VIII

Pixel

0.61

0.72

0.53

0.59

  1. Parameter I: closed area; Parameter II: central anterior chamber section area; Parameter III: ratio of central anterior chamber section area to limbus area; Parameter IV: anterior chamber angle area; Parameter V: central area of the central anterior chamber section; Parameter VI: anterior chamber angle; Parameter VII: ratio of lengths of the corneal band to iris band; Parameter VIII: difference in lengths between the corneal band and iris band.