Extended Data Fig. 3: Detailed analysis of electrographic seizure dynamics controlled by tFUS neurostimulation with the default protocol. | Nature Electronics

Extended Data Fig. 3: Detailed analysis of electrographic seizure dynamics controlled by tFUS neurostimulation with the default protocol.

From: A shape-morphing cortex-adhesive sensor for closed-loop transcranial ultrasound neurostimulation

Extended Data Fig. 3

a, b, tFUS with the default protocol (40 Hz, 5% DC, 0.5 W cm−2 Ispta, 1 W cm−2 Isppa) suppressed seizure spike waves. In the closed-loop operation determined by neural activity information recorded from the SMCA device, tFUS stimulation automatically turned on () by HFO detection and turned off () after the seizure SW was suppressed. Representative timetrace of all channels (a) and timetrace and spectrogram of a single channel (Ch.9) (b) of the closed-loop seizure control episodes. c, Two-dimensional topographic frames of electrographic seizures illustrating the spatiotemporal pattern within the duration of interest from seizure SW initiation (*1 in phase II) and SW suppressing (*2 in phase III) events.

Source data

Back to article page