Table 1 In vivo functionality of PC-BU aortic valves.
1 month N = 2 | 3 months N = 3 | 6 months N = 3 | 9 months N = 2 | 12 months N = 4 | |
---|---|---|---|---|---|
Valve Insufficiency Grade | Mild: 1 Moderate: 1 | None: 1 Mild: 2 | Mild: 2 Severe: 1 | None: 1 Moderate: 1 | Mild: 2 Severe: 2 |
Valve Stenosis Grade | None: 1 Mild: 1 | None: 3 | None: 3 | None: 1 Moderate: 1 | None: 2 Mild: 1 Severe: 1 |
PLVOT systole, mmHg (mean, SD) | X | 120 ± 18 | 86 ± 14 | 81 ± 5 | 95 ± 9 |
PLVOT diastole, mmHg (mean, SD) | X | 107 ± 17 | 68 ± 18 | 64 ± 12 | 68 ± 10 |
PLV systole, mmHg (mean, SD) | X | 150 ± 20 | 117 ± 13 | 117 ± 5 | 128 ± 5 |
Transvalvular ΔP peak-to-peak gradient, mmHg (mean, SD) | X | 31 ± 4 | 31 ± 4 | 36 ± 14 | 33 ± 14 |