Fig. 6: Prototype graft flow experiments.
From: Realtime monitoring of thrombus formation in vivo using a self-reporting vascular access graft

A Experimental setup of the flow experiment. Inflow-black, outflow-white, PCB-green, battery and switch in yellow box hatching. B Clot in situ inside of the graft (upper panel) and emboli flowing through inside of graft section (lower panel). C shows the impendence value of each pairing of electrodes. D shows all electrode parings grouped and tested using mean one-way ANOVA ± SEM (N = 3 independent biological repeats) F value = 129.8, degrees of freedom = 2 PBS vs. Blood, ****p < 0.0001, PBS vs. Emboli, ****p < 0.0001, Blood vs. Emboli ****p < 0.0001, PBS vs. Clot ****p < 0.0001, Blood vs. Clot ****p < 0.0001, Emboli vs. Clot ****p < 0.0001. Exact p-values are in the Supplementary Data, except for values where GraphPad Prism indicated that the p value was < 0.0001.