Fig. 8: Impedance measurements from preclinical studies. | Communications Medicine

Fig. 8: Impedance measurements from preclinical studies.

From: Realtime monitoring of thrombus formation in vivo using a self-reporting vascular access graft

Fig. 8

A Impedance measurements across the sensors. B Shows the grouping of the repeats of blood and emboli (N = 18), Mean ± SEM, t-test, degrees of freedom (n = 17), ****p = <0.0001. C shows the difference from blood at T0 to clot at T5 (N = 12) Mean ± SEM, t-test, degrees of freedom = 11 ***p < 0.001. D Shows the graft clotting (clot) over time versus a heparinised graft (hep) (N = 4) mean ±SD. @ 4 min **p = <0.01, then at 5 min, p < 0.05. Exact p-values are in the Supplementary Data. 1x female cadaver for developing surgical approach, 2x female swine for non-recovery surgery sacrifice.