Table 4 Exam card flash slab
From: Multi-modal ultra-high resolution structural 7-Tesla MRI data repository
General | |
|---|---|
TA | 17:18 |
PAT | Off |
Voxel size | 0.5×0.5×0.5 mm |
Rel. SNR | 1.0 |
Properties | |
Prio Recon | Off |
Load to viewer | On |
Inline movie | Off |
Auto store images | On |
Load to stamp segments | Off |
Load images to graphic segments | Off |
Auto open inline display | Off |
Start measurement without further preparation | On |
Wait for user to start | Off |
Start measurements | single |
Routine | |
Slab group 1 | |
Slabs | 1 |
Dist. factor | 20% |
Position | R2.4 A29.1 H23.0 |
Orientation | T>C-23.0 |
Phase enc. dir. | R>>L |
Rotation | 90.00 deg |
Phase oversampling | 0% |
Slice oversampling | 12.5% |
Slices per slab | 128 |
FoV read | 192 mm |
FoV phase | 81.3% |
Slice thickness | 0.50 mm |
TR | 41 ms |
TE 1 | 11.22 ms |
TE 2 | 20.39 ms |
TE 3 | 29.57 ms |
Averages | 1 |
Concatenations | 1 |
Filter | None |
Coil elements | A24 |
Contrast | |
MTC | Off |
Magn. preperation | None |
Flip angle | 14 deg |
Fat suppr. | None |
Water suppr. | None |
SWI | Off |
Averaging mode | Short term |
Reconstruction | Magn./Phase |
Measurements | 1 |
Multiple series | Each measurement |
Resolution | |
Base resolution | 384 |
Phase resolution | 100% |
Slice resolution | 100% |
Phase partial Fourier | 6/8 |
Slice partial Fourier | 6/8 |
Interpolation | Off |
PAT mode | None |
Image Filter | Off |
Distortion Corr. | Off |
Prescan Normalize | Off |
Normalize | Off |
B1 filter | Off |
Raw filter | Off |
Elliptical filter | Off |
Geometry | |
Multi-slice mode | Interleaved |
Series | Interleaved |
Saturation mode | Standard |
Special sat. | None |
Table position | H |
Table position | 0 mm |
Inline Composing | Off |
Tim CT mode | Off |
System | |
V24 | Off |
A24 | On |
Positioning mode | REF |
MSMA | S-C-T |
Sagittal | R >>L |
Coronal | A>>P |
Transversal | F>>H |
Save uncombined | Off |
Coil Combine Mode | Adaptive Combine |
AutoAlign | --- |
Auto Coil Select | Off |
Shim mode | Standard |
Adjust with body coil | Off |
Confirm freq. adjustment | Off |
Assume Silicone | Off |
? Ref. amplitude 1H | 0.000 V |
Adjustment Tolerance | Auto |
Adjust volume | |
Position | R2.4 A29.1 H23.0 |
Orientation | T>C-23.0 |
Rotation | 90.00 deg |
A>>P | 192 mm |
R>>L | 156 mm |
F>>H | 64 mm |
Physio | |
1st Signal/Mode | None |
Segments | 1 |
Tagging | None |
Dark blood | Off |
Resp. control | Off |
Inline | |
Subtract | Off |
Liver registration | Off |
Std-Dev-Sag | Off |
Std-Dev-Cor | Off |
Std-Dev-Tra | Off |
Std-Dev-Time | Off |
MIP-Sag | Off |
MIP-Cor | Off |
MIP-Tra | Off |
MIP-Time | Off |
Save original images | On |
Wash – In | Off |
Wash – Out | Off |
TTP | Off |
PEI | Off |
MIP – time | Off |
MapIt | None |
Contrasts | 3 |
Sequence | |
Introduction | On |
Dimension | 3D |
Elliptical scanning | Off |
Phase stabilization | On |
Asymmetric echo | Off |
Bandwidth 1 | 160 Hz/Px |
Bandwidth 2 | 160 Hz/Px |
Bandwidth 3 | 160 Hz/Px |
Flow comp. 1 | Yes |
Flow comp. 2 | No |
Flow comp. 3 | No |
Readout mode | Monopolar |
RF pulse type | Normal |
Gradieet mode | Whisper |
Excitation | Slab-sel. |
RF spoiling | On |
length exc pulse | 3,000 us |
Ernst Angle? | On |
T1 | 1,300 ms |
FFT scale factor | 2.5 |