Table 1 Glucose concentrations in the blood serum samples (n = 3)

From: Rapid, Sensitive and Reusable Detection of Glucose by a Robust Radiofrequency Integrated Passive Device Biosensor Chip

Sample*

Found at hospital (mg/mL)

Determined by biosensor (mg/mL) ± RSD%

Added fructose (mg/mL)

Determined by biosensor (mg/mL) ± RSD%

Recovery (%)

1.

0.92

0.9352 ± 0.40

0.1

0.9352 ± 0.46

101.65

2.

1.05

1.0603 ± 0.25

0.1

1.0603 ± 0.54

100.98

3.

1.20

1.1853 ± 0.39

0.1

1.1853 ± 0.45

98.77

  1. *Each sample was measured four times.