Table 3 CPH removal data and CPH starting concentrations of the simulated treatment experiments using various filters and plasma as the test medium.
From: Cell-free plasma hemoglobin removal by dialyzers with various permeability profiles
Filter | CPH clearance [ml/min] | CPH start concentration [mg/L] | Removed CPH mass [mg] | A405reduction ratio [%] | CPH mass reduction ratio [%] |
---|---|---|---|---|---|
MCO1 | 5.8 ± 1.2 | 577 ± 21 | 151 ± 32 | 27 ± 5.8 | 26 ± 5.1 |
MCO2 | 7.3 ± 1.0 | 620 ± 13 | 185 ± 36 | 27 ± 9.0 | 30 ± 6.5 |
MCO3 | 8.8 ± 0.6 | 616 ± 7.7 | 234 ± 42 | 35 ± 0.9 | 38 ± 6.5 |
MCO4 | 12.7 ± 1.7 *13.7 ± 0.8 | 438 ± 8.0 | 260 ± 1.0 | 49 ± 3.4 | 59 ± 1.2 |
P170H | <0 | 629 ± 14 | not determined | <0 | not determined |
CorDiax | <0 | 515 ± 85 | not determined | <0 | not determined |