Table 4 CPH removal data and CPH starting concentrations of the simulated treatment experiments using various filters and whole blood as the test medium.
From: Cell-free plasma hemoglobin removal by dialyzers with various permeability profiles
Filter | CPH clearance [ml/min] | CPH start concentrations [mg/L] | Removed CPH mass [mg] | A405reduction ratio [%] | CPH mass reduction ratio [%] |
---|---|---|---|---|---|
MCO1 | 5.8 ± 1.2 | 908 ± 20 | 268 ± 18 | 38 ± 10 | 19 ± 2.4 |
MCO2 | 5.4 ± 0.6 | missing data | missing data | 36 ± 5.1 | missing data |
MCO3 | 7.5 ± 0.5 | 820 ± 6.9 | 363 ± 51 | 49 ± 4.3 | 39 ± 8.6 |
MCO4 | 11.3 ± 1.6 | missing data | missing data | 62 ± 10 | missing data |
P170H | <0 | 1186 ± 42 | not determined | <0 | not determined |
CorDiax | <0 | 970 ± 32 | not determined | <0 | not determined |