Table 1 Comparison of point scatter above and below the level of detection (LOD) for various analytes and tissues combinations.

From: The Statistical Value of Raw Fluorescence Signal in Luminex xMAP Based Multiplex Immunoassays

Cytokine

Tissue

LOD

N

Above

Below

CV-A

CV-B

Ratio

p.value

Sig.

IFN-g (21)

plasma

17.418

11

80

11

0.025

0.076

0.327

0.001

**

IFN-g (21)

serum

17.418

28

52

28

0.059

0.081

0.724

0.064

 

IL-10 (56)

plasma

34.966

29

29

62

0.161

0.129

1.248

0.472

 

IL-10 (56)

serum

34.966

32

48

32

0.116

0.133

0.873

0.319

 

IL-11 (39)

plasma

27.995

33

33

58

0.14

0.123

1.136

0.945

 

IL-11 (39)

serum

27.995

37

37

43

1.239

0.146

8.469

0

***

IL-22 (18)

plasma

27.463

12

79

12

0.016

0.092

0.173

0

***

IL-22 (18)

serum

27.463

24

56

24

0.082

0.108

0.759

0.188

 

IL-27 (13)

plasma

24.529

9

82

9

0.027

0.101

0.271

0.001

**

IL-27 (13)

serum

24.529

23

56

23

0.157

0.155

1.013

0.461

 

IL-29 (33)

serum

18.187

19

61

19

0.045

0.078

0.574

0.018

*

IL-32 (35)

plasma

58.531

37

37

54

0.262

0.204

1.282

0.867

 

IL-32 (35)

serum

58.531

39

41

39

0.322

0.284

1.131

0.902

 

IL-34 (15)

plasma

60.664

16

16

75

0.229

0.177

1.296

0.868

 

IL-34 (15)

serum

60.664

24

24

56

1.319

0.176

7.509

0

***

Light (51)

plasma

136.28

29

62

29

0.059

0.116

0.505

0.001

**

Light (51)

serum

136.28

8

72

8

0.049

0.106

0.465

0.056

 
  1. LOD represents the fluorescence level associated with the concentration determined as the lower level of detection, N the number of responses above and below the associated LOD used in the analysis. Above and Below give the actual number of observed responses above and below the LOD. The columns CV-A and CV-B give the coefficient of variation (CV) of the samples, each of size N, above and below the respective LOD. Ratio gives the value of CV-A/CV-B. P.value represents the uncorrected probability of an F test statistic. The number of *’s in the Sig. column indicates the significance of the associated p-values for each analyte and tissue combination.