Figure 5
From: Effects of dose reduction on bone strength prediction using finite element analysis

Correlation between fracture load values for each individual dose, 80 mAs (A), 150 mAs (B) & 220 mAs (C) as a function of fracture load values obtained at 500 mAs, and Bland-Altman plots depicting the mean of each individual dose (80, 150, 200 mAs) and of 500 mAs versus the difference between the two doses (D–F). The horizontal lines indicate the mean value, and 95% confidence intervals at ±1.96 standard deviation (SD).