Table 1 Several parameters in temperature calibration.
From: Real-time micro-scale temperature imaging at low cost based on fluorescent intensity ratio
T (°C) | The average value of FIR | Standard deviation | Precision | Resolution of T (°C) |
|---|---|---|---|---|
120 | 0.0436 | 0.0021 | 0.049 | 1.6 |
130 | 0.0618 | 0.0028 | 0.045 | 1.5 |
140 | 0.0811 | 0.0034 | 0.042 | 1.5 |
150 | 0.1024 | 0.0042 | 0.041 | 1.5 |
160 | 0.1210 | 0.0045 | 0.037 | 1.4 |
170 | 0.1690 | 0.0059 | 0.035 | 1.4 |
180 | 0.2110 | 0.0073 | 0.035 | 1.4 |
190 | 0.269 | 0.010 | 0.037 | 1.6 |
200 | 0.343 | 0.013 | 0.038 | 1.7 |
210 | 0.417 | 0.015 | 0.035 | 1.6 |
220 | 0.515 | 0.018 | 0.035 | 1.7 |
230 | 0.621 | 0.023 | 0.037 | 1.9 |
240 | 0.765 | 0.029 | 0.038 | 2.0 |
250 | 0.932 | 0.035 | 0.038 | 2.1 |
260 | 1.111 | 0.042 | 0.038 | 2.1 |
270 | 1.334 | 0.051 | 0.039 | 2.3 |
280 | 1.537 | 0.058 | 0.038 | 2.3 |