Table 2 Summary of the mechanical design parameters and FEA simulation results.

From: Solid cryogen: a cooling system for future MgB2 MRI magnet

Mechanical design parameters

SN2 chamber

Radiation shield

Cryostat

Design pressure (MPa)

0.30 (internal)

0.30 (internal)

Allowable pressure (MPa)

0.30 (internal)

0.45 (internal), 1.07 (external)

Calculated cylinder thickness (mm)

1.70

Cylinder thickness (mm)

3.40

2.00

4.78

Calculated end flange thickness (mm)

7.84

14.62

End flange thickness (mm)

8.00

3.00

15.00

FEA simulated von Mises stress (MPa)

9.17

16.31

91.68

FEA simulated total displacement (mm)

0.004

0.28

0.20