Aim To determine the ability of frequency doubling technology (FDT) perimetry to detect visual field defects of neurological origin.
Methods A total of 15 eyes of nine patients who all had complete hemianopias or quadrantanopias underwent the FDT 20-5 screening mode test and Humphrey 24-2 SITA Fast visual field test (HFA). The FDT results were scored according to the number of abnormal test locations (out of a maximum of 4) in each affected quadrant. FDT locations showing a defect of P<2% were considered abnormal.
Results Of the 15 eyes, six showed complete superior quadrantanopic and nine complete hemianopic field defects on HFA. Of 96 FDT test locations in these quadrants or hemifields only 38 were abnormal on FDT testing (40%). For the quadrantanopic field defects, five out of 24 locations were abnormal (21%). For the hemianopic field defects, 33 out of 72 locations were abnormal (49%). In three eyes (two with quadrantanopias and one with complete hemianopia), FDT perimetry failed to demonstrate any corresponding abnormality.
Conclusions The FDT screening test can fail to demonstrate complete hemianopic and quadrantanopic field defects. Users should be aware of this deficiency when using FDT to screen for field defects.
Similar content being viewed by others
Log in or create a free account to read this content
Gain free access to this article, as well as selected content from this journal and more on nature.com
or
References
Kelly DH . Frequency doubling in visual responses. J Opt Soc Am A. 1966; 56: 1628–1633.
Johnson CA, Samuels SJ . Screening for glaucomatous visual field loss with frequency-doubling perimetry. Invest Ophthalmol Vis Sci. 1997; 38(2): 413–425.
Alward WLM . Frequency doubling technology perimetry for the detection of glaucomatous visual field loss. Am J Ophthalmol 2000; 129(3): 376–377.
Quigley HA . Identification of glaucoma-related visual field abnormality with the screening protocol of frequency doubling technology. Am J Ophthalmol 1998; 125: 819–828.
Woodward K, Brito CF, Wall M . Variability of FDT perimetry: effect of a 3 degree offset along the vertical in patients with hemianopias and normals. Invest Ophthalmol Vis Sci 2001; 42(4): S150. Abstract 803.
Wall M, Newman S, Johnson C, Fingeret M, Maxner C, Woodward K . Improvement of sensitivity of FDT perimetry to detect hemianopias with a 3 degree stimulus offset along from the vertical. Invest Ophthalmol Vis Sci 2001; 42(4): S153 Abstract 820.
Zeki S . A Vision of the Brain. Blackwell Scientific Publications: Oxford, 1993, pp 347–350.
Morland AB, Jones SR, Finlay AL, Deyzac E, Le S, Kemp S . Visual perception of motion, luminance and colour in a human hemianope. Brain 1999; 122: 1183–1198.
Celesia GG, Bushnell D, Toleikis SC, Brigell MG . Cortical blindness and residual vision: is the ‘second’ visual system in humans capable of more than rudimentary visual perception? Neurology 1991; 41: 862–869.
Weiskrantz L . In: Mackintosh NJ, Shallice T, Schacter D, Treisman A, Weiskrantz L (eds). Blindsight: A Case Study and Implications. Oxford University Press: Oxford, 1986.
Nicholas JA, Wall M, Newman S, Johnson C, Fingeret M, Maxner C et al. Sensitivity of FDT perimetry screening tests compared with full threshold testing in patients with homonymous hemianopias. Invest Ophthalmol Vis Sci 2001; 42(4): S150. Abstract 800.
Acknowledgements
The authors are grateful to Mr Nigel Davies and Dr Tony Morland for their helpful comments on the paper.
Author information
Authors and Affiliations
Corresponding author
Additional information
Presented as a poster at the Oxford Ophthalmological Congress 2001
Rights and permissions
About this article
Cite this article
Fong, K., Byles, D. & Constable, P. Does frequency doubling technology perimetry reliably detect neurological visual field defects?. Eye 17, 330–333 (2003). https://doi.org/10.1038/sj.eye.6700376
Received:
Published:
Issue date:
DOI: https://doi.org/10.1038/sj.eye.6700376
Keywords
This article is cited by
-
Using frequency-doubling perimetry to detect optic neuropathy in patients with Graves’ orbitopathy
Japanese Journal of Ophthalmology (2008)
-
Clinical assessment of optic nerve disorders
Eye (2004)