Abstract
Aims
To compare the intraocular pressure (IOP) measurements obtained using the Pascal dynamic contour tonometer (PDCT) with the standard Goldmann applanation tonometer (GAT) and to correlate these with central corneal thickness (CCT) in patients with normal corneas.
Methods
A prospective, masked, comparative case series of 116 eyes from patients attending a glaucoma clinic. IOP was measured with PDCT by one examiner and with GAT by a masked, independent examiner. A mean of six CCT readings was used for analysis.
Results
IOP measured by the two instruments correlated significantly (r=0.77; P<0.0001). IOP measured by GAT correlated strongly with CCT (r=0.37, P=0.0001) whereas the relationship between IOP measured by PDCT and CCT approached significance (r=0.17, P=0.073). The differences between GAT and PDCT measured IOP also correlated strongly with CCT (r=0.37, P<0.0001). The 95% limits of agreement between GAT and PDCT were ±4.2 mmHg. Dividing the eyes into three groups on the basis of CCT, demonstrated those in the thickest tertile showed a poorer agreement between instruments and the GAT measured significantly higher IOP in this group (P=0.003) while the PDCT showed no significant differences with different CCTs (P=0.37).
Conclusion
Demonstration of the relative independence of PDCT IOP measurements from CCT supports a potential clinical role for this instrument, particularly for subjects with CCT outside the normal range.
Similar content being viewed by others
Log in or create a free account to read this content
Gain free access to this article, as well as selected content from this journal and more on nature.com
or
References
Whitacre MM, Stein R . Sources of error with use of Goldmann-type tonometers. Surv Ophthalmol 1993; 38(1): 1–30.
Goldmann V, Schmidt T . Uber Applanationstonometrie. Ophthalmologica 1957; 134: 221–242.
Hansen FK . A clinical study of the normal human central corneal thickness. Acta Ophthalmologica 1971; 49(1): 82–99.
Herman DC, Hodge DO, Bourne WM . Increased corneal thickness in patients with ocular hypertension. Arch Ophthalmol 2001; 119(3): 334–336.
Wolfs RC, Klaver CC, Vingerling JR, Grobbee DE, Hofman A, de Jong PT . Distribution of central corneal thickness and its association with intraocular pressure: the Rotterdam Study. Am J Ophthalmol 1997; 123(6): 767–772.
Copt RP, Thomas R, Mermoud A . Corneal thickness in ocular hypertension, primary open-angle glaucoma, and normal tension glaucoma. Arch Ophthalmol 1999; 117(1): 14–16.
Doughty MJ, Zaman ML . Human corneal thickness and its impact on intraocular pressure measures: a review and meta-analysis approach. Surv Ophthalmol 2000; 44(5): 367–408.
Ehlers N, Bramsen T, Sperling S . Applanation tonometry and central corneal thickness. Acta Ophthalmologica 1975; 53(1): 34–43.
Whitacre MM, Stein RA, Hassanein K . The effect of corneal thickness on applanation tonometry. Am J Ophthalmol 1993; 115(5): 592–596.
Bron AM, Creuzot-Garcher C, Goudeau-Boutillon S, d'Athis P . Falsely elevated intraocular pressure due to increased central corneal thickness. Graefes Arch Clin Exp Ophthalmol 1999; 237(3): 220–224.
Herndon LW, Choudhri SA, Cox T, Damji KF, Shields MB, Allingham RR . Central corneal thickness in normal, glaucomatous, and ocular hypertensive eyes. Arch Ophthalmol 1997; 115(9): 1137–1141.
Gordon MO, Beiser JA, Brandt JD, Heuer DK, Higginbotham EJ, Johnson CA et al. The Ocular Hypertension Treatment Study: baseline factors that predict the onset of primary open-angle glaucoma. Arch Ophthalmol 2002; 120(6): 714–720.
Kass MA, Heuer DK, Higginbotham EJ, Johnson CA, Keltner JL, Miller JP et al. The Ocular Hypertension Treatment Study: a randomized trial determines that topical ocular hypotensive medication delays or prevents the onset of primary open-angle glaucoma. Arch Ophthalmol 2002; 120(6): 701–713.
Phelan PS, McGhee CN, Bryce IG . Excimer laser PRK and corticosteroid induced IOP elevation: the tip of an emerging iceberg? Br J Ophthalmol 1994; 78(10): 802–803.
Zadok D, Tran DB, Twa M, Carpenter M, Schanzlin DJ . Pneumotonometry versus Goldmann tonometry after laser in situ keratomileusis for myopia. J Cataract Refract Surg 1999; 25(10): 1344–1348.
Mardelli PG, Piebenga LW, Whitacre MM, Siegmund KD . The effect of excimer laser photorefractive keratectomy on intraocular pressure measurements using the Goldmann applanation tonometer. Ophthalmol 1997; 104(6): 945–948.
Chatterjee A, Shah S, Bessant DA, Naroo SA, Doyle SJ . Reduction in intraocular pressure after excimer laser photorefractive keratectomy. Correlation with pretreatment myopia. Ophthalmology 1997; 104(3): 355–359.
Shaikh NM, Shaikh S, Singh K, Manche E . Progression to end-stage glaucoma after laser in situ keratomileusis. J Cataract Refract Surg 2002; 28(2): 356–359.
Hamilton DR, Manche EE, Rich LF, Maloney RK . Steroid-induced glaucoma after laser in situ keratomileusis associated with interface fluid. Ophthalmology 2002; 109(4): 659–665.
Mitchell P, Hourihan F, Sandbach J, Wang JJ . The relationship between glaucoma and myopia: the Blue Mountains Eye Study. Ophthalmol 1999; 106(10): 2010–2015.
Kaufmann C, Bachmann LM, Thiel MA . Intraocular pressure measurements using dynamic contour tonometry after laser in situ keratomileusis. Invest Ophthalmol Vis Sci 2003; 44(9): 3790–3794.
SMT Swiss Microtechnology AG. Pascal Dynamic Contour Tonometer Operating Manual, version 1.4, Swiss Microtechnology: Switzerland, 2004, pp 30–32.
Bland J, Altman D . Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1986; 1: 307–310.
Siganos DS, Papastergiou GI, Moedas C . Assessment of the Pascal dynamic contour tonometer in monitoring intraocular pressure in unoperated eyes and eyes after LASIK. J Cataract Refract Surg 2004; 30(4): 746–751.
Duba I, Wirthlin AC . Dynamic contour tonometry for post-LASIK intraocular pressure measurements. Klin Monatsbl fur Augenheilkd 2004; 221(5): 347–350.
Feltgen N, Leifert D, Funk J . Correlation between central corneal thickness, applanation tonometry, and direct intracameral IOP readings. Br J Ophthalmol 2001; 85(1): 85–87.
Leske MC, Heijl A, Hussein M, Bengtsson B, Hyman L, Komaroff E et al. Factors for glaucoma progression and the effect of treatment. Arch Ophthalmol 2003; 121: 48–56.
Parrish RK, Palmberg P, Sheu WP, XLT Study Group. A comparison of latanoprost, bimatoprost, and travoprost in patients with elevated intraocular pressure: a 12 week, randomized, masked-evaluator multicenter study. Am J Ophthalmol 2003; 135: 688–703.
Noecker RS, Dirks MS, Choplin NT, Bernstein P, Batoosingh AL, Whitcup SM et al. A six-month randomized clinical trial comparing the intraocular pressure-lowering efficacy of bimatoprost and latanoprost in patients with ocular hypertension or glaucoma. Am J Ophthalmol 2003; 135: 55–63.
Shih CY, Graff ZJS, Trokel SL, Tsai JC . Clinical significance of central corneal thickness in the management of glaucoma. Arch Ophthalmol 2004; 122(9): 1270–1275.
Foster PJ, Wong JS, Wong E, Chen FG, Machin D, Chew PT . Accuracy of clinical estimates of intraocular pressure in Chinese eyes. Ophthalmology 2000; 107(10): 1816–1821.
Korey M, Gieser D, Kass MA, Waltman SR, Gordon M, Becker B . Central corneal endothelial cell density and central corneal thickness in ocular hypertension and primary open-angle glaucoma. Am J Ophthalmol 1982; 94(5): 610–616.
Wang TH, Huang JY, Hung PT, Shieh JW, Chen YF . Ocular hypotensive effect and safety of brinzolamide ophthalmic solution in open angle glaucoma patients. J Formos Med Assoc 2004; 103(5): 369–373.
Lass JH, Eriksson GL, Osterling L, Simpson CV, Latanoprost Corneal Effects Study Group. Comparison of the corneal effects of latanoprost, fixed combination latanoprost-timolol, and timolol: a double-masked, randomized, one-year study. Ophthalmology 2001; 108(2): 264–271.
Schwenn O, Troost R, Vogel A, Grus F, Beck S, Pfeiffer N . Ocular pulse amplitude in patients with open angle glaucoma, normal tension glaucoma, and ocular hypertension. Br J Ophthalmol 2002; 86(9): 981–984.
Kaufmann C, Thiel MA, Valavanis A, Landau K, Duba I, Wirthlin AC . Normalisation of ocular pulse amplitude after embolisation of dural cavernous sinus arteriovenous fistula. Klin Monatsbl fur Augenheilkd 2004; 221(5): 431–434.
Kerr J, Nelson P, O'Brien C . A comparison of ocular blood flow in untreated primary open-angle glaucoma and ocular hypertension. Am J Ophthalmol 1998; 126(1): 42–51.
Georgopoulos GT, Diestelhorst M, Fisher R, Ruokonen P, Krieglstein GK . The short-term effect of latanoprost on intraocular pressure and pulsatile ocular blood flow. Acta Ophthalmol Scand 2002; 80(1): 54–58.
Acknowledgements
The study was partially supported by unrestricted research funding from: the Alcon glaucoma and neuro-ophthalmology fellowship fund, and the Sir William and Lady Stevenson Trust. We thank Designs for Vision for the unrestricted loan of the Pascal dynamic contour tonometer used in this study.
Author information
Authors and Affiliations
Corresponding author
Additional information
Statement
The authors have no proprietary or commercial interest in the equipment or studies presented in this paper.
Ethics approval
This study was approved by the Auckland Ethics Committee, Auckland, New Zealand.
Rights and permissions
About this article
Cite this article
Ku, J., Danesh-Meyer, H., Craig, J. et al. Comparison of intraocular pressure measured by Pascal dynamic contour tonometry and Goldmann applanation tonometry. Eye 20, 191–198 (2006). https://doi.org/10.1038/sj.eye.6701849
Received:
Accepted:
Published:
Issue date:
DOI: https://doi.org/10.1038/sj.eye.6701849
Keywords
This article is cited by
-
Evaluation von Korrekturformeln für die Tonometrie
Der Ophthalmologe (2017)
-
Evaluation of lamina cribrosa thickness and depth in ocular hypertension
Japanese Journal of Ophthalmology (2016)
-
Comparison of Goldmann applanation and dynamic contour tonometry in a population of Mexican open-angle glaucoma patients
International Ophthalmology (2013)
-
Comparison of dynamic contour tonometry and Goldmann applanation tonometry in relation to central corneal thickness in primary congenital glaucoma
Graefe's Archive for Clinical and Experimental Ophthalmology (2013)
-
The influence of soft contact lenses on the intraocular pressure measurement
Eye (2012)