Abstract
Purpose
To measure optic shift (OS) of a single piece monofocal intraocular lens (IOL) with varying relationships between anterior capsule cover and IOL optic.
Methods
This is a prospective randomized masked study of 150 eyes undergoing phacoemulsification. Eyes received either option: 360o total capsule cover (group I); partial cover (group II); or no cover (group III). OS was calculated as difference in anterior chamber depth after administration of cyclopentolate 1% and pilocarpine 2% on IOLMaster® at separate visits at 6 months follow-up. Subsequently, using retro-illumination photographs, percentage area of capsule cover was calculated. OS within and between groups I and II was analyzed. The impact of one quartile change in area of capsule cover on percentage change in OS was measured for both groups. Unpaired t-test, correlation, and regression were applied.
Results
In groups I, II, and III, mean age of patients was 56.68±6.38, 57.09±7.34, 59.15±6.35 years, respectively; mean OS (mm) was 1.25±0.28, 1.20±0.24; 0.95±0.26 (P=0.013), respectively; and percentage area of capsule cover (%) was 47.35±10.48, 33.83±10.11, 0.16±0.13 (P=0.001), respectively. Mean percentage area of capsule cover in group I vsgroup II was significant (P=0.001). OS was 1.22±0.26 mm in groups I and II (combined) vs0.95±0.26 mm in group III (P=0.004, (0.06, 0.33)). OS in group I vsgroup II was not significant (P=0.46). Correlation coefficient was r=0.38 (P<0.001). With every increment on quartile (a quartile is any of the three values which divide the sorted data set into four equal parts, so that each part represents one-fourth of the sample or population) class of area of capsule cover the OS increased by 0.12 mm.
Conclusion
The OS differed significantly between total and partial cover groups combined vsno cover group.
Similar content being viewed by others
Log in or create a free account to read this content
Gain free access to this article, as well as selected content from this journal and more on nature.com
or
References
Langenbucher A, Seitz B, Huber S, Nguyen NX, Kuchle M . Theoretical & measured pseudophakic accommodation after implantation of a new accommodative posterior chamber intraocular lens. Arch Ophthalmol 2003; 121: 1722–1727.
Langenbucher A, Huber S, Nguyen NX, Seitz B, Gusek-Schneider GC, Kuchle M . Measurement of accommodation after implantation of an accommodation posterior chamber intraocular lens. J Cataract Refract Surg 2003; 29: 677–685.
Wirtitsch MG, Findl O, Menapace R, Kriechbaum K, Koeppl C, Buehl W et al. Effect of haptic design on change in axial lens position after cataract surgery. J Cataract Refract Surg 2004; 30: 45–51.
Koeppl C, Findl O, Menapace R, Kriechbaum K, Wirtitsch M, Buehl W et al. Pilocarpine-induced shift of an accommodating intraocular lens: AT-45 Crystalens. J Cataract Refract Surg 2005; 31: 1290–1297.
Gonzalez F, Capeans C, Santos L, Suarez J, Cadarso L . Anteroposterior shift in rigid and soft implants supported by the intraocular capsular bag. Graefes Arch Exp Ophthalmol 1992; 230: 237–239.
Findl O, Drexler W, Menapace R, Bobr B, Bittermann S, Vass C et al. Accurate determination of effective lens position and lens-capsule distance with 4 intraocular lenses. J Cataract Refract Surg 1998; 24: 1094–1098.
Lesiewska-Junk H, Kahuzny J . Intraocular lens movement and accommodation in eyes of young patients. J Cataract Refract Surg 2000; 26: 562–565.
Findl O, Kiss B, Petternel V, Menapace R, Georgopoulos M, Rainer G et al. Intraocular lens movement caused by ciliary muscle contraction. J Cataract Refract Surg 2003; 29: 669–676.
Muftuoglu O, Hosal BM, Karel F, Zilelioglu G . Drug-induced intraocular lens movement and near visual acuity after AcrySof intraocular lens implantation. J Cataract Refract Surg 2005; 31: 1298–1305.
Vargas LG, Auffarth GU, Becker KA, Rabsilber TM, Holzer MP . Performance of the ICU accommodating intraocular lens in relation to capsulorhexis size. J Cataract Refract Surg 2005; 31: 363–368.
Cekic O, Batman C . The relationship between capsulorhexis size and anterior chamber depth relation. Ophthalmic Surg Lasers 1999; 30: 185–190.
Nawa Y, Ueda T, Nakatsuka M, Tsuji H, Marutani H, Hara Y et al. Accommodation obtained per 1.0 mm forward movement of a posterior chamber intraocular lens. J Cataract Refract Surg 2003; 29: 2069–2072.
Arshinoff SA . Dispersive-cohesive viscoelastic soft shell technique. J Cataract Refract Surg 1999; 25: 167–173.
Fine IH . Cortical cleaving hydrodissection. J Cataract Refract Surg. 2000; 26: 943–944.
Vasavada AR, Raj SM, Johar K, Nanavaty MA . Effect of hydrodissection alone and hydrodissection combined with rotation on lens epithelial cells: surgical approach for the prevention of posterior capsule opacification. J Cataract Refract Surg. 2006; 32: 145–150.
Vasavada AR, Singh R . Step-by-step, chop in situ and separation of very dense cataracts. J Cataract Refract Surg 1998; 24: 156–159.
Vasavada AR, Desai JP . Stop, chop, chop, and stuff. J Cataract Refract Surg 1996; 22: 526–529.
Vasavada AR, Raj S . Step-down technique. J Cataract Refract Surg 2003; 29: 1077–1079.
Vasavada AR, Desai J, Singh R . Enlarging the capsulorhexis. J Cataract Refract Surg 1997; 23: 329–331.
Vasavada AR, Shastri L . Initial and definitive capsulorhexes: an extended application. J Cataract Refract Surg 2000; 26: 634.
Vasavada AR, Raj SM . Anterior capsule relationship of the AcrySof intraocular lens optic and posterior capsule opacification: a prospective randomized clinical trial. Ophthalmology. 2004; 111: 886–894.
Pande MV, Ursell PG, Spalton DJ, Heath G, Kundaiker S . High-resolution digital imaging of the posterior lens capsule after cataract surgery. J Cataract Refract Surg 1997; 23: 1521–1527.
Ravalico G, Baccara F . Apparent accommodation in Pseudophakic eyes. Acta Ophthalmol (Copenh) 1990; 68: 604–606.
Nanavaty MA, Vasavada AR, Patel AS, Raj SM, Desai TH . Analysis of patients with good uncorrected distance and near vision after monofocal intraocular lens implantation. J Cataract Refract Surg 2006; 32: 1091–1097.
Legeais JM, Werner L, Werner L, Abenhaim A, Renard G . Pseudoaccommodation: BioConFold versus a foldable silicone intraocular lens. J Cataract Refract Surg 1999; 25: 262–267.
Holladay JT, Rubin ML . Avoiding refractive problems in cataract surgery. Surv Opthalmol 1988; 357–360.
Holladay JT . Standardizing constants for ultrasonic biometry, keratometry, and intraocular lens power calculations. J Cataract Refract Surg 1997; 23: 1356–1370.
Noda H, Nishida S, Ogino N, Ariki G, Tsunekawa M, Hayashi H et al. Objective evaluation of apparent accommodation of pseudophakic eyes. [Article in Japanese]. Nippon Ganka Gakkai Zasshi 1994; 98: 187–191.
Nakaizumi H, Sasaki K, Sakamoto Y . In vivo observation of axial movement of intraocular lenses through an anterior eye segment analysis system. Ophthalmic Res 1992; 24 (Suppl 1): 21–25.
Hardman Lea SJ, Rubinstein MP, Snead MP, Haworth SM . Pseudophakic accommodation? A study of stability of capsular bag supported, one piece, rigid tripod, or soft flexible implants. Br J Ophthalmol 1990; 74: 22–25.
Nakazawa M, Ohtsuki K . Apparent accommodation in pseudophakic eyes after implantation of posterior chamber intraocular lenses. Am J Ophthalmol 1983; 96: 435–438.
Huber C . Myopic astigmatism as a substitute for accommodation in pseudophakia. Dev Ophthalmol 1981; 5: 17–26.
Kuchle M, Nguyen NX, Langenbucher A, Gusek-Schneider GC, Seitz B, Hanna KD . Implantation of a new accommodative posterior chamber intraocular lens. J Refract Surg 2002; 18: 208–216.
Auffarth GU, Tetz MR, Biazid Y, Volker HE . Measuring anterior chamber depth with Orbscan Topography System. J Cataract Refract Surg 1997; 23: 1351–1353.
Fukuyama M, Oshika T, Amano S, Yoshitomi F . Relationship between apparent accommodation and corneal multifocality in pseudophakic eyes. Ophthalmology 1999; 106: 1178–1181.
Kriechbaum K, Findl O, Koeppl C, Menapace R, Drexler W . Stimulus-driven versus pilocarpine-induced biometric changes in pseudophakic eyes. Ophthalmology 2005; 112: 453–459.
Vetrugno M, Cardascia N, Cardia L . Anterior chamber depth measured by two methods in myopic and hyperopic phakic IOL implant. Br J Ophthalmol 2000; 84: 1113–1116.
Naeser K, Naeser A, Boberg-Ans J, Bargum R . Axial length following implantation of posterior chamber lenses. J cataract Refract Surg 1989; 15: 673–675.
Kriechbaum K, Findl O, Kiss B, Sacu S, Petternel V, Drexler W . Comparison of anterior chamber depth measurement methods in phakic and pseudophakic eyes. J Cataract Refract Surg 2003; 29: 89–94.
Koeppl C, Findl O, Kriechbaum K, Sacu S, Drexler W . Change in IOL position and capsular bag size with an angulated intraocular lens early after cataract surgery. J Cataract Refract Surg 2005; 31: 348–353.
Kriechbaum K, Findl O, Preussner PR, Koppl C, Wahl J, Drexler W . Determining postoperative anterior chamber depth. J Cataract Refract Surg. 200; 29: 2122–2126.
Findl O, Drexler W, Menapace R, Hitzenberger CK, Fercher AF . High precision biometry of pseudophakic eyes using partial coherence interferometry. J Cataract Refract Surg 1998; 24: 1087–1093.
Findl O, Drexler W, Menapace R, Heinzl H, Hitzenberger CK, Fercher AF . Improved prediction of intraocular lens power using partial coherence interferometry. J Cataract Refract Surg 2001; 27: 861–867.
Findl O, Drexler W, Menapace R, Bobr B, Bittermann S, Vass C et al. Accurate determination of effective lens position and lens-capsule distance with 4 intraocular lenses. J Cataract Refract Surg 1998; 24: 1094–1098.
Drexler W, Baumgartner A, Findl O, Hitzenberger CK, Sattmann H, Fercher AF . Submicrometer precision biometry of the anterior segment of the human eye. Invest Ophthalmol Vis Sci 1997; 38: 1304–1313.
Urbak SF . Ultrasound biomicroscopy III. Accuracy and agreement of measurements. Acta Ophthalmol Scand 1999; 77: 293–297.
Davidson JA . Capsule contraction syndrome. J Cataract Refract Surg 1993; 19: 582–589.
Ursell PG, Spalton DJ, Pande MV . Anterior capsule stability in eyes with intraocular lenses made of poly-(methyl methacrylate), silicone, and AcrySof. J Cataract Refract Surg 1997; 23: 1532–1538.
Young DA, Orlin SE . Capsulorhexis contraction in phacoemulsification surgery. Ophthalmic Surg 1994; 25: 477–478.
Hansen SO, Crandall AS, Olson RJ . Progressive constriction of the anterior capsular opening following intact capsulorhexis. J Cataract Refract Surg 1993; 19: 77–82.
Spang KM, Rohrbach JM, Weidle EG . Complete occlusion of the anterior capsular opening after intact capsulorhexis: clinicopathologic correlation. Am J Ophthalmol 1999; 127: 343–345.
Schmidbauer JM, Vargas LG, Apple DJ, Escobar-Gomez M, Izak A, Arthur SN et al. Evaluation of neodymiumyttrium-aluminum-garnet capsulotomies in eyes implanted with AcrySof intraocular lenses. Ophthalmology 2002; 109: 1421–1426.
Author information
Authors and Affiliations
Corresponding author
Additional information
This paper was presented in part as a poster at European Society of Cataract and Refractive Surgery 2005, Lisbon. The authors do not have any commercial connection and/or interest in marketing any product, instrument, or piece of equipment discussed in this paper
Rights and permissions
About this article
Cite this article
Nanavaty, M., Raj, S., Vasavada, V. et al. Anterior capsule cover and axial movement of intraocular lens. Eye 22, 1015–1023 (2008). https://doi.org/10.1038/sj.eye.6702817
Received:
Accepted:
Published:
Issue date:
DOI: https://doi.org/10.1038/sj.eye.6702817
Keywords
This article is cited by
-
Femtosecond laser-assisted cataract surgery in shallow anterior chamber cases
International Ophthalmology (2021)
-
Pseudophakic anterior chamber depth measurement
Eye (2009)
-
Reply: Pseudophakic anterior chamber depth measurement
Eye (2009)