Table 1 Sensitivity, specificity, and accuracy of the PVA-EBPD hydrogel sensor

From: Instantaneous self-recovery and ultra-low detection limit hydrogel electronic sensor for temporomandibular disorders intelligent diagnosis

  

Clinical diagnosis

Total

  

TMJ disc displacement with reduction

Normal TMJ

 

PVA-EBPD hydrogel sensor diagnosis

TMJ abnormality

62 (TP)

12 (FN)

74

TMJ normality

0 (FP)

46 (TN)

46

Total

 

62

58

120

Sensitivity

100.00% (94.22%–100.00%)

Specificity

79.31% (66.65%–88.83%)

Accuracy

90.00%

Youden Index

0.79