Fig. 6: State-of-the-art of the accuracy of quantum current sources and current generation. | Nature Communications

Fig. 6: State-of-the-art of the accuracy of quantum current sources and current generation.

From: A primary quantum current standard based on the Josephson and the quantum Hall effects

Fig. 6

Comparison between the relative uncertainty achieved in the generation of currents traceable to the ampere using the PQCG (this work and ref. 27) and others quantum current sources: metallic (8,54), GaAs (9,11,14) and Si (12,13) SECS, series connection of QHRS array and PJVS for generating (24) current. CMCs (black line)55. Inset: \(\overline{{\Delta }I/I}\) measurements performed in this work (orange diamond) as a function of current value I. The value at 45.94 μA corresponds to the weighted mean of \(\overline{{\Delta }I/I}\) values obtained using N1 = 160 and N1 = 465.