Abstract
The new definition of the ampere calls for a quantum current standard able to deliver a flow of elementary charges, e, controlled with a relative uncertainty of 10−8. Despite many efforts, nanodevices handling electrons one by one have never demonstrated such an accuracy for a net flow. The alternative route based on applying Ohm’s law to the Josephson voltage and quantum Hall standards recently reached the target uncertainty in the milliampere range, but this was at the expense of the application of error corrections. Here, we present a new programmable quantum current generator, which combines both quantum standards and a superconducting cryogenic amplifier in a quantum electrical circuit enabling the current scaling without errors. Thanks to a full quantum instrumentation, we demonstrate the accuracy of the generated currents, in the microampere range, at quantized values, ±(n/p)efJ, with relative uncertainties less than 10−8, where n and p are integer control parameters and fJ is the Josephson frequency. This experiment sets the basis of a universal quantum realization of the electrical units, for example able of improving high-value resistance measurements and bridging the gap with other quantum current sources.
Similar content being viewed by others
Introduction
Since the last revision of the International System of Units (SI) in May 20, 2019, founded on seven fixed constants of nature1,2, any source generating an electric current which can be expressed in terms of ef, with e the elementary charge and f a frequency in Hz (s−1), provides a realisation of the ampere.
Single-electron current sources (SECS)3,4,5, which are mesoscopic devices6,7,8,9,10,11,12,13,14, able to handle electrons one by one at a rate fe, are often presented as the most obvious way to realize the definition. However, achieving currents above 100 pA using GaAs and Si-based tunable-barrier SECS accurate to within a relative uncertainty better than 10−7 remains a very challenging goal because of increasing error rates at high frequencies (~1 GHz)15,16. Very recently, as a consequence of the phase-charge quantum mechanical duality in Josephson junctions (JJ), dual Shapiro steps have been evidenced in superconducting nanowires and small JJ placed in high impedance environments under microwave radiation17,18,19. Here, the enhanced phase variance allows photon-assisted tunneling of fluxons ϕ0 = h/2e (h is the Planck constant) and a synchronized transfer of Cooper pairs. Sharp current steps appearing at integer multiples of 2efe in the DC current-voltage characteristics could be promising candidates as quantum sources in the nanoampere range, although their flatness is still in debate20. More generally, for all mesoscopic current sources, the control of charge fluctuations, which are dependent on the device coupling with the electromagnetic environment, remains a crucial issue.
Concurrently, another route to the SI realization consists in applying Ohm’s law to the Josephson voltage and quantum Hall resistance standards, since the Josephson effect21 and the quantum Hall effect22 now provide direct and universal realizations of the volt and the ohm from h/2e and h/e 2 constants, respectively2, with a 10−9 measurement uncertainty. The high accuracy of the Josephson voltage standards, which are series arrays of JJ, relies on the phase rigidity of macroscopic superconductors. Under application of dc current bias and a microwave radiation fJ, the transfer through each JJ of one fluxon per period of the microwave tone is ensured and results in a quantized voltage V = nJϕ0 fJ23, where nJ is the number of JJ. If this quantized voltage can be accurately applied to a quantum Hall resistance standard (QHRS) in the ν=2 Landau level filling factor, taking advantage of the charge rigidity of the quantum Hall edge states, nJe charges are transferred at a rate fJ through the QHRS of h/2e 2 resistance. Hence, a current nJefJ, easily reaching microamperes, can be generated. Recently, a calculable current of 1 μA generated from the series connection of both quantum standards has been measured with a relative uncertainty of 1.3 × 10−7 24,25, but the accuracy was reached owing to the very low-resistance of the ammeter (~0.1 mΩ). The main issue is therefore to implement the accurate series connection of the two quantum standards while realizing a true current source. The programmable quantum current generator (PQCG)26,27,28 has addressed this issue by locking an electron flow to the current circulating in the loop formed by the quantum standards, with the help of a superconducting amplifier, allowing simultaneously the scaling over a wider range of current values. Its accuracy was demonstrated in the milliampere range with a relative measurement uncertainty of 10−8. This result was however obtained at the expense of corrections of the order of a few parts in 107, originating from the non-ideal connection of the quantum standards, and determined after time consuming additional measurements. This impaired the final uncertainty, the simplicity and hence the full potential of the new quantum current standard.
Here, we report on a next-generation PQCG able of delivering lower-noise currents, at the theoretical quantized values driven by the Josephson frequency, without any classical correction. This performance results from the implementation of a three-terminal connection of the QHRS, which allows a highly-accurate application of the Josephson voltage to the quantized resistance. We demonstrate the realization of the ampere with relative uncertainties below 10−8 for different current levels, filling the gap between the milliampere range and the microampere range. This is achieved using a full quantum instrumentation made of five quantum devices, which opens the way to the realization of several electrical units in a single experiment.
Results
Next-generation PQCG
Figure 1 a shows the implementation (see also Quantum devices section in “Methods”) of two programmable Josephson voltage standards (PJVS), two QHRS and a cryogenic current comparator (CCC). The two PJVS are binary divided 1 V Nb/NbxSi1−x/Nb series arrays29, both having a total of 8192 JJ and working around 70 GHz. The voltage of the two PJVS are given by ± n1,2ϕ0 f1,2 with n1,2 the number of JJ biased on the ± 1 Shapiro steps. The two QHRS are both GaAs/AlGaAs heterostructures30 of quantized resistance R1 = R2 = h/2e2. The CCC is a dc current transformer31, made of several superconducting windings of different number of turns, able to compare currents with a great accuracy (below one part in 109) and sensitivity (80 pA ⋅ turns/Hz1/2) owing to Ampère’s law and Meissner effect.
a The PQCG is composed of PJVS1 and QHRS1 connected through a triple connection (red lines 1, 2, 3) ensured via the windings of N1 turns of a new CCC, which are used to measure and amplify the quantized current I1. The shields at ground potential of the high and low potential cables are not shown on the scheme. The SQUID detector of the CCC feedbacks on an external current source (VCCS) which maintains the current IPQCG into the series connection of QHRS2 and of the winding of N2 turns. To reduce the finite gain error of the SQUID, a voltage source of the programmable Josephson bias source (PJBS) drives the VCCS and preadjusts the output current, such that the SQUID feedbacks only on a small fraction of IPQCG (see Uncertainties section in Methods). The voltage drop at QHRS2 is measured in a quantum voltmeter configuration made of PJVS2 and a null detector, ND (EM Electronics Model N11). The frequency of the Josephson microwave signals are referenced to a 10 MHz frequency standard linked to an atomic clock time base. Both PJVS are controlled and synchronized using the same PJBS. For measurements performed with n1 = 4096, a decoupling stage ensures the electrical isolation between the two PJVS so that the PQCG and the quantum voltmeter can each be grounded. For n1 = 1920, no decoupling stage is used implying that only PJVS1 is grounded. Four different cryostats are used to cool down quantum devices (represented by purple, blue gray, cyan and light blue colors). b ND recordings, ΔV, for frequencies f2 + Δ f (red circle) and f2 − Δf (blue circle) as a function of time τ while the current IPQCG is periodically switching on and off (I+ measurement protocol).
The new version of the PQCG is composed of PJVS129 connected to QHRS1 with a triple connection (see Quantum devices section in “Methods”) ensured through three identical windings of N1 turns of a new specially designed CCC (see CCC section in “Methods”). This connection technique32 reduces the impact of the series resistances to an insignificant effect. More precisely, one current contact and two voltage contacts of the same equipotential of QHRS1 are connected all together at each superconducting pads of PJVS1. Because of the topological properties of Hall edge-states, namely their chirality, their h/e2 two-wire resistance and their immunity against backscattering33, the current flowing through the third contact is only a fraction \({(r/{R}_{1})}^{2}\) of the current circulating in the first one, where r is the typical resistance of the connections. The resistance seen by PJVS1 is close to h/2e 2 within a typical small correction of order \({(r/{R}_{1})}^{3}\). It results that the total current circulating I1, is close to ± n1ef1 within 1.5 parts in 1010 for series resistance values lower than 5 ohms (see Multiple series connection section in “Methods”). Compared to ref. 27, where the double connection required the application of a relative correction to the current of a few 10−7, the operation of the PQCG is simplified since no correction is necessary here. The quantized current I1, divided in the three connections, is measured by the three identical windings of N1 turns. A DC SQUID is used to detect the unbalance ampere ⋅ turns in the different windings of the new CCC. It feedbacks on the new battery-powered voltage controlled current source (VCCS), which supplies a winding of N2 turns with current IPQCG in order to maintain the ampere ⋅ turns balance N1I1 − N2IPQCG = 0. It results that the PQCG is able to output a current equal to:
to within an estimated Type B relative uncertainty of 2 parts in 109 (see Table 1 and Uncertainties section in “Methods”). In practice, the CCC gain G = N1/N2 can span two orders of magnitude on either side of the unity, allowing the generation of currents with values from nanoampere to milliampere.
Accuracy test principle
No matter how accurately the Type B uncertainty may be estimated, a metrological requirement is to check that the experimental current IPQCG is given by the relationship (1) using another quantum measurement method through an accuracy test. The output current is determined by feeding QHRS2, and by measuring the voltage drop, V2, at its Hall terminals using a quantum voltmeter (Fig. 1 and Supplementary Fig. 1). The latter is made of PJVS2 and an analog null detector (ND), which measures the voltage difference ΔV. From Kirchhoff’s voltage law, IPQCG is determined according to the expression:
with ΔV = 0 ideally at the equilibrium frequency \({f}_{2}^{{\rm{eq}}}\). Using a quantized resistance R2 = h/2e 2 (~ 12.9 kΩ), about 129 times higher than in ref. 27, allows increasing the signal-to-noise ratio while eliminating an extra resistance calibration. The relative deviation of the measured current to the theoretical one, \(\Delta I/I=({I}_{{{{\rm{PQCG}}}}}/{I}_{{{{\rm{PQCG}}}}}^{{{{\rm{th}}}}})-1\), is given by \(\Delta I/I=\frac{{n}_{2}\, {f}_{2}^{{\rm{eq}}}}{G{n}_{1}\, {f}_{1}}-1\). In practice, the CCC gain and the numbers of JJ are chosen so that Gn1 = n2. Thus, the nominal relative deviation can be expressed as:
This relationship (3) clearly emphasizes that the accuracy test resumes to the determination of the equilibrium frequency \({f}_{2}^{{{{\rm{eq}}}}}\) which should be equal to f1, if the current is accurately quantized, but can correspond to another value in case of a systematic error. However, noise and offset drifts can make it difficult to find the frequency corresponding precisely to ΔV = 0. Here, instead of performing a single measurement at frequency f2 close to \({f}_{2}^{{{{\rm{eq}}}}} \sim {f}_{1}\), and measuring a voltage signal hidden by the noise, two successive voltage mean values, \(\overline{{\Delta }{V}_{{f}_{2}^{+}}}\) and \(\overline{{\Delta }{V}_{{f}_{2}^{-}}}\), are measured at two detuned frequencies \({f}_{2}^{+}={f}_{2}+\Delta f\) and \({f}_{2}^{-}={f}_{2}-\Delta f\), respectively, where Δf is set to 40 kHz or 80 kHz in our experiments (Fig. 1b and Supplementary Fig. 2a). The measurement of significant voltage signals is a straightforward way to evaluate the signal to noise ratio and quickly detect any anomalous event such as trapped magnetic flux either in the SQUID or in the JJ. In order to mitigate the effect of offsets, drifts and 1/f noise, each voltage mean value is obtained from a measurement series consisting in periodically either switching on and off the current with IPQCG > 0 (I+) or < 0 (I−), or completely reversing the current (I±). The equilibrium frequency is then determined from:
which implies voltage ratios only. The two successive voltage measurements allow an in-situ and short-time calibration of the nanovoltmeter gain from the Josephson frequency, overcoming its lack of stability over time. Finally, the determination of ΔI/I does not require any calibration, resulting in a reduced Type B standard uncertainty, uB, of 2.1 × 10−9 (see Table 1 and Uncertainties section in Methods). The Type A standard uncertainty of ΔI/I, \({u}^{{{{\rm{A}}}}}=\frac{{u}^{{{{\rm{A}}}}}({f}_{2}^{{{{\rm{eq}}}}})}{{f}_{1}}\) is determined from the standard deviations of the mean of the voltage series \(\Delta {V}_{{f}_{2}^{+}}\) and \(\Delta {V}_{{f}_{2}^{-}}\) (see Uncertainties section in Methods). This is justified by the calculation of the relative Allan deviation27 showing a dominant white noise contribution (Supplementary Fig. 2b).
Quantized current accuracy
Measurements of ΔI/I were performed at four different values of current 5.74 μA, 11.48 μA, 45.94 μA and 57.42 μA, using a primary current I1 of 45.94 μA obtained with n1 = 4096. This large current improves the operational margins of the PQCG compared to ref. 27 and increases the signal-to-noise ratio while ensuring a perfect quantization of the Hall resistance of the QHRS1 device. Each measurement series was typically carried out over one day using I+, I−, and I± measurement protocols to reveal any systematic effect related to the current direction. Note that implementing complete current reversals I± required the reduction of the noise in the circuit28 (see CCC section in Methods). Measurements were performed with N1 = 160 or also with 465 to test the effect of the number of ampere ⋅ turns. However, the downside of the latter configuration is the higher instability of the feedback loop encountered during the current reversals which prevented the use of the I± measurement protocol. The different output currents were obtained by changing N2 from 80 to 1860. Other PQCG parameters are reported in Table 2 in “Methods”.
At the lower current values of 5.74 μA (Fig. 2a) and 11.48 μA (Fig. 2c), discrepancies of ΔI/I are covered by Type A uncertainties, uA, ranging from 1 to 3 × 10−8. Weighted means, ΔI/IWM, for each measurement protocol I+, I− and I±, reported in Fig. 2b, d, show that there is no significant deviation of the current from its theoretical value within combined measurement uncertainties of about 10−8. Besides, the mean value of ΔI/IWM(I+) and ΔI/IWM(I−) is clearly in agreement with ΔI/IWM(I±) at 5.74 μA, which confirms the equivalence of averaging measurements carried out using the I+ and I− protocols with the measurement obtained using the I± protocol. Combining the different results (see Weighted mean values section in “Methods”), one obtains the relative deviations, for the equivalent protocol I±, \(\overline{\Delta I/I}=(2\pm 4.3) \times 1{0}^{-9}\) and \(\overline{\Delta I/I} = (-7.9 \pm 8.6)\times 1{0}^{-9}\) at current levels of 5.74 μA and 11.48 μA, respectively.
Series of relative deviations ΔI/I with n1 = 4096 measured at 5.742201056 μA using N1 = 160 (a), at 11.48440211 μA using N1 = 465 (c), at 45.93760845 μA using N1 = 160 (e) and at 45.93760845 μA using N1 = 465 (g). Measurements were carried out over one or two days, using I+, I− or I± measurement protocols (circle). Some measurements are performed with Ibias shifted by +0.1 mA and –0.1 mA (a and e), with a frequency f1 = 70.02 GHz (a), using the double connection (applying the corresponding α2 = (2.344 ± 0.037) × 10−7 correction) and using the triple connection with 50 Ω inserted into the first connection (applying the corresponding α3 = (1.4 ± 0.02) × 10−9 correction). All uncertainties correspond to one standard deviation, i.e., are given with a coverage factor k=1. Error bars correspond to type A standard uncertainties, uA. Weighted mean values ΔI/IWM of the series measured at 5.74 μA (b), at 11.48 μA (d), at 45.94 μA using N1 = 160 (f) and at 45.94 μA using N1 = 465 (h). In (b, d), weighted mean values for I+, I−, I± (filled diamond) are calculated from the uA, mean value of I+ and I− results (open diamond). In (f, h), weighted mean values for I+, I−, I± (filled triangle) are calculated from \(\sqrt{{({u}^{{{{\rm{A}}}}})}^{2}+{({u}_{{{{\rm{it-noise}}}}}^{{{{\rm{A}}}}})}^{2}}\), where \({u}_{{{{\rm{it-noise}}}}}^{{{{\rm{A}}}}}\) accounts for the observed residual intra-day noise, mean value of I+ and I− results (open triangle). Error bars correspond to combined standard uncertainties \({u}_{{{{\rm{WM}}}}}^{{{{\rm{c}}}}}\) (see Uncertainties section in Methods). Other parameter: N2 is fixed such as G = N1/N2 = n2/n1.
At the higher current value of 45.94 μA, ΔI/I measurements, reported in Fig. 2e, g, are characterized by smaller Type A standard uncertainties, uA, in agreement with the larger voltage drop at the terminals of QHRS2 (about 0.59 V). As expected, even smaller uncertainties are observed for N1 = 465 than for N1 = 160 due to an enhanced ampere ⋅ turns value N1I1. The lower uncertainties permit the scrutiny of small but significant deviations revealing intra-day noise at the 10−8 level, the origin of which has not been clearly identified yet. To account for this, the standard uncertainty of each ΔI/I measurement is increased of an additional Type A uncertainty component \({u}_{{{{\rm{id}}}}-{{{\rm{noise}}}}}^{{{{\rm{A}}}}}=1{0}^{-8}\) chosen so that the χ2 criterion is fulfilled (see Uncertainties section in “Methods”). The resulting weighted mean values, ΔI/IWM, shown in Fig. 2f, h, do not reveal any significant deviation from zero with regards to the combined measurement uncertainties of only a few 10−9. Combining results obtained using the different measurement protocols, one obtains \(\overline{{\Delta }I/I}=(5.4\pm 3.1)\times 1{0}^{-9}\) for N1 = 160 and \(\overline{{\Delta }I/I}=(-1.1\pm 3.6)\times 1{0}^{-9}\) for N1 = 465. Hence, on average over a day, the current delivered by the PQCG is quantized and the deviation from zero is covered by an uncertainty of about 4 × 10−9. At shorter terms, the uncertainty due to the intra-day noise does not average out and the combined uncertainty is ≃ 10−8. Generally, one might guess a small discrepancy between measurements performed using either I+ or I− protocols, in Fig. 2b–d, which could come from a small Peltier type effect. However using the protocol I± cancels this potential effect.
Similar results are obtained for \(\overline{{\Delta }I/I}\) at current value of 57.42 μA (N1 = 160), giving (1.9 ± 2.6) × 10−9 (Fig. 3a). The margins over which the current values remain quantized at the same level of uncertainty have been tested in different situations. No significant deviation of the generated current is measured when shifting by ±0.1 mA the Josephson bias current of PJVS1, Ibias (shown in Fig. 2a at 5.74 μA and in Fig. 2e at 45.94 μA) or by varying the PJVS1 frequency from 70 to 70.02 GHz (Fig. 2a). The efficiency of the triple connection against large cable resistance value and the accuracy of the cable corrections, when applied, have been demonstrated by inserting a large resistance (50 Ω) into the first connection of the triple connection scheme, and by application of the cable correction in the double connection scheme (Figs. 2e, 3b, respectively). Finally, at 43.07 μA, another connection scheme (see Quantum devices section in “Methods”), including JJ into the triple connection of PJVS1 with n1 = 1920, although less reliable with respect to magnetic flux trapping, confirms an accuracy at a level of a few parts in 108 (Fig. 3c).
a Relative deviations, ΔI/I, at 57.42201056 μA measured with n1 = 4096 for I+, I− and I± protocols, mean value of I+ and I− results (open circle). Errors bars corresponds to combined uncertainties (see Uncertainties section in Methods). b Relative current shift from a reference value Iref, in back and forth measurements testing the effectiveness of the multiple connection (data extracted from Fig. 2e (see Combined results section in “Methods”) at 45.94 μA): measurements using the double connection with the 2nd order correction (α2) applied, using the triple connection with 50 Ω inserted in the first connection with third order correction α3 applied. The results are consistent within the Type A standard uncertainty. This confirms the accuracy of the cable corrections when applied and the efficiency of the triple connection against large cable resistance value. c Series of ΔI/I measurements carried out at 43.06650792 μA for I±, I+ and I− measurement protocols using a second PJVS1 configuration with n1 = 1920 and G = N1/N2 = 160/80. Errors bars corresponds to uA uncertainties. Discrepancies are within less than two parts in 108. This demonstrates that the PQCG can be accurate within this uncertainty even though the biasing scheme is less robust with respect to trapping magnetic flux (see Quantum devices section in Methods).
Evaluation of short-term noise sources
The analysis of the uA uncertainties provide a further insight into the understanding of the experiment. Figure 4a, b shows averages, \({u}_{\exp }^{{{{\rm{A}}}}}({\tau }_{m})\), of the uA uncertainties measured in the different accuracy tests, after normalization to the same measurement time τm = 16τ0 (with τ0 = 66 s, τm ~ 18 min) and to the same measurement protocol I±, as a function of 1/V2 and 1/N1I1, respectively. For comparison, they also report the theoretical standard uncertainties, \({u}_{{{{\rm{calc}}}}}^{{{{\rm{A}}}}}({\tau }_{m})=\frac{\sqrt{3}}{16}\sqrt{\frac{{S}_{\Delta I/I}}{{\tau }_{A}}}\), calculated using τA = 12 s from the ΔI/I noise density (see Experimental and Calculated standard uncertainties section in “Methods”):
considering a magnetic flux noise detected by the SQUID of \(\sqrt{{S}_{{{\upphi }}}}=62\,{{\upmu}} {{{{\upphi }}}}_{0}/{{{{\rm{Hz}}}}}^{1/2}\), a voltage noise of \(\sqrt{{S}_{V}}=28\,{{{\rm{nV}}}}/{{{{\rm{Hz}}}}}^{1/2}\) and using the CCC sensitivity γCCC = 8 μA ⋅ turn/ϕ0. These values give a very satisfactory agreement with the experimental data. The \(\sqrt{{S}_{{{\upphi}}}}\) value is compatible with the noise spectrum of the CCC (see CCC section in “Methods”). The \(\sqrt{{S}_{V}}\) value is higher than the nominal voltage noise expected from the ND (EM electronics N11) for a 13 kΩ resistor load. As it was previously observed34, the manifestation of large transient voltages at the input of the null detector during current switchings is a possible explanation for larger scattering of voltage measurements, but we cannot rule out the capture of external noise by the primary loop28. In both cases, there is room for improvement in the future. Figure 4a confirms that, at low V2, the main noise contribution comes from the voltage noise of the quantum voltmeter, as emphasized by the 1/V2 dependence which is reproduced by \({u}_{{{{\rm{calc}}}}}^{{{{\rm{A}}}}}({\tau }_{m})\) calculated for Sϕ = 0. On the other hand, at higher V2 (Fig. 4b), the number of ampere ⋅ turns becomes the dominant parameter and the experimental uncertainties follow the 1/N1I1 dependence of \({u}_{{{{\rm{calc}}}}}^{{{{\rm{A}}}}}({\tau }_{m})\) calculated for SV = 0. One can therefore deduce the Type A uncertainty contributions of the PQCG and the quantum voltmeter, which amount to 1.6 × 10−11/(N1I1) and 8.8 × 10−10/V2, respectively. The low value of 7.3 × 10−10 calculated for N1 = 465 and I1 = 45.94 μA for the PQCG itself allows considering the generation of even smaller currents. However, the demonstration of their accuracy would require the increase of R2, by using a series quantum Hall array of 1.29 MΩ resistance35 to increase V2.
Experimental uncertainties, \({u}_{\exp }^{{{{\rm{A}}}}}\), determined from all data for a normalized measurement time (τm = 16τ0 ~ 18 min) and the I± measurement protocol, as a function of 1/V2 (a) and 1/N1I1 (b). Calculated uncertainties, \({u}_{{{{\rm{calc}}}}}^{{{{\rm{A}}}}}\) (cross). Functions 8.8 × 10−10/V2 (dash-dotted line in a) and 1.6 × 10−11/(N1I1) (dash-line in b) represent \({u}_{{{{\rm{calc}}}}}^{{{{\rm{A}}}}}({\tau }_{m})\) calculated for Sϕ = 0 and SV = 0, respectively.
Application to an ammeter calibration
Figure 5a shows the relative deviations, ΔI/I DA, between the currents measured by the DA and the quantized currents generated by the PQCG in the 100 μA range using the second configuration of PJVS1 with n1 = 1920 (Supplementary Table. 1). The coarse adjustment of the quantized currents, about ± 107.7 μA and ± 62.6 μA, is done by using G = 465/93 (or 160/32) and G = 465/160, respectively. Using either I+ or I− measurement protocols (Supplementary Fig. 3a), Allan deviation shows that the Type A relative uncertainty for the τm = 144 s measurement time amounts to about 2 × 10−7 at ± 107.7 μA (Supplementary Fig. 3b). Data demonstrate that the DA is reproducible over the current range within about 5 parts in 107, similar to results obtained in the milliampere range27. Finally, Fig. 5b illustrates the possibility of a fine tuning of the current by varying fJ from 69.98 to 70.02 GHz, which represents a relative shift of the quantized current of ± 3 × 10−4 around 107.7 μA.
a Relative deviation, ΔI/I DA, as a function of current value in 100 μA range for different PQCG configurations using n1 = 1920 : N1 = 465, N2 = 93, 160 (black diamond), double connection with N1 = 465, N2 = 93 and application the correction α2 (open black diamond), N1 = 160, N2 = 32 (red circle). Calibrations performed at ±107.7 μA using either N1=465 or N1=160, or using the double connection scheme and applying the cable correction α2 are all in agreement. b ΔI/I DA as a function of the Josephson frequency f1, or relative deviation of current around I0 = 107.666272 μA (black circle).
Discussion
Quantum current standard: state-of-the-art
We have demonstrated the accuracy of the flow rate of electrons generated by the new generation PQCG, at current values bridging the gap between the microampere range and the milliampere range with relative uncertainties ≤ 10−8, as summarized in Fig. 6 (see inset). The results support our estimation of the Type B uncertainty budget. Moreover, Type A uncertainties of only a few 10−9 have been measured. These progress stem from eliminating the need of any classical correction, improving the signal-to-noise ratio, extending the operating margins and applying new measurement protocols based on tuning Josephson frequencies. These results pave the way for a quantum current standard that is as accurate as voltage and resistance standards, and similarly requires only verification of quantization criteria27,28,36 when used for current calibration.
Comparison between the relative uncertainty achieved in the generation of currents traceable to the ampere using the PQCG (this work and ref. 27) and others quantum current sources: metallic (8,54), GaAs (9,11,14) and Si (12,13) SECS, series connection of QHRS array and PJVS for generating (24) current. CMCs (black line)55. Inset: \(\overline{{\Delta }I/I}\) measurements performed in this work (orange diamond) as a function of current value I. The value at 45.94 μA corresponds to the weighted mean of \(\overline{{\Delta }I/I}\) values obtained using N1 = 160 and N1 = 465.
Figure 6 shows the state-of-the-art of the uncertainties of the accuracy tests of quantum current sources based on different quantum technologies, along with the best calibration measurement capabilities (CMCs) achieved in national metrology institutes (NMIs) for comparison. It recalls the uncertainties achieved in the present work along with those reported in our previous work27. This illustrates the wide range of current covered by our quantum current standard. At much lower currents, around 100 pA, uncertainties at the level of 10−7 are achieved by the best SECS. This uncertainty level is also reached for currents around 1 μA by one experiment based on the series connection of quantum Hall resistance array and PJVS24. Let us remark that the uncertainties achieved depend not only on the current source itself but also on the method used to measure the generated current. To this respect, the best known measurement techniques reach relative uncertainties of about 10−7: in ref. 37,38 (from 100 pA to 1 μA), in ref. 39 (around 1 μA), in ref. 40 (around 10 mA). On the other hand, the uncertainties <10−8 demonstrated with the PQCG comes not only from its own accuracy and stability but also from the measurement with the quantum voltmeter. Providing such an accurate primary quantum current standard in the current range of the best CMCs, which are limited by uncertainties two orders of magnitude larger, is essential both to improve the transfer of the ampere towards end-users and to foster the development of more accurate instruments, as emphasized by the calibration of a digital ammeter with uncertainties limited by the instrument itself.
Figure 6 also emphasizes the importance of exploring PQCG capabilities towards even smaller currents, in order to bridge the gap in the current delivered by SECS and devices exhibiting dual Shapiro steps. This would open the way to a new metrological triangle experiment4,41. More precisely, considering the variant of the PQCG proposed in ref. 26 and the noise level estimated in this work, we could expect generating and measuring a 10 nA current with a relative uncertainty of 5 × 10−8 after 12 h measurement (N1 = 160, I1 = 0.33 μA, N2 = 5400).
Future prospects
Another important result is the demonstration of the PQCG accuracy using a 129 times larger resistance (QHRS2) than in ref. 27, which shows its robustness against the load resistance, as required for a true current source. Moreover, the PQCG accuracy being now established, our experiments can be interpreted as calibrations of resistors of 13 kΩ and 100 Ω values with a 10−8 measurement uncertainty. More generally, combining Eqs. (1), (2) leads to:
where the equilibrium can be coarsely set by choosing G and n1,2 and finely tuned by adjusting f1 and \({f}_{2}^{{\rm{eq}}}\). This new method combining the PQCG and the quantum voltmeter (see Supplementary Fig. 1) paves the way for a paradigm shift for the resistance calibration. It allows to simplify the calibration of a large resistance from \(\frac{h}{2{e}^{2}}\) to a single step, suppressing the intermediate steps needed using a conventional resistance comparison bridge42.
Furthermore, the full quantum instrumentation developed gives foundation to a DC quantum calibrator-multimeter able to provide the primary references of voltage, resistance and current, which are needed in NMIs. This development opens the way to a flexible traceability of electrical measurements to the SI units based on a set of quantum devices gathered in the same experiment. In this perspective, a simplified implementation and operation of the quantum instrumentation is required. The demonstration of the PQCG using a PJVS cooled down in a cryostat equipped with a pulse-tube cryocooler constitutes a first simplification. By gathering the two PJVS and the CCC in the same cryostat, and the two QHRS in a new dry cryostat, our quantum instrumentation will be more practical. Besides, graphene-based single Hall bars43,44 or arrays45,46 replacing GaAs devices could provide noise reduction and further simplification owing to an operation under relaxed experimental conditions. In the longer term, QHRS based on the quantum anomalous Hall effect47,48,49,50 operating at zero magnetic field like PJVS, or heterostructures-based JJ comprising stacked cuprates51, could lead to even more compact and easy instrumentation.
Methods
Quantum devices
Implementation
The two quantum Hall resistance standards, QHRS1 and QHRS2, are both cooled down in the same cryostat at 1.3 K under of magnetic field of 10.8 T. PJVS1 is cooled down in a small amount of liquid helium maintained at 4.2 K in a recondensing cryostat based on a pulse-tube refrigerator while PJVS2 is cooled down in a 100 l liquid He Dewar at 4.2 K. The quantization state of PJVS and QHRS is periodically checked following technical guidelines36,52. In case of occasional trapped flux in the PJVS, a quick heating of the array allows to fully restore the quantized voltage steps. Finally, the CCC is placed in another liquid He Dewar. Connecting five quantum devices, while ensuring quantized operation of PJVS devices and minimizing noise, is very challenging. Many efforts have been spent in optimizing the wiring, the positions of the grounding points and the bias configuration of the PJVS.
Shielding
It is essential to cancel the leakage current that could alter the accurate equality of the total currents circulating through the QHRS and the windings. This is achieved by placing high- and low-potential cables (high-insulation RL > 1 TΩ resistance) connected to the PJVS, to the QHRS and to the CCC inside two separated shields, which are then twisted together and connected to ground. In this way, direct leakage currents short circuiting the QHRS, the most troublesome, are canceled. Other leakage currents are redirected to ground.
PJVS cabling
The two PJVS are binary divided 1 V Nb/NbxSi1−x/Nb series arrays29, both having a total of 8192 JJ and working around 70 GHz. The sequence of the segments that can be biased in the Josephson arrays is the following : 4096/2048/1024/512/256/128/1/31/32/64. Three bondings wires have been added at both ends of the arrays, on the same superconducting pad, in order to implement the triple connection as illustrated in Fig. 1. A 50 Ω heater is placed close to the Josephson array chip allowing to get rid of trapped flux within few minutes. We have used a prototype of the hermetic cryoprobe developed for the recondensing cryostat, which had only 8 available wires, reducing the possible wiring configurations for PJVS1. Figure 7a, b shows the two wiring configurations used with n1 = 4096 and n1 = 1920, respectively. They differ essentially on the way the triple connection is implemented at the low potential side of the bias source (PJBS). In Fig. 7a, the triple connection is done on the same superconducting pad, while in Fig. 7b, JJ are present between the bias wire and the wires connecting QHRS1, but also between the second and third connection of the triple connection. If the current circulating in the JJ is less than half the amplitude of the n = 0 Shapiro step (<500 μA), both configurations are equivalent. However, the second one turned out to be less reliable than the first one when connecting the rest of the circuit. Because of the ground loop including the JJ, it was very sensitive to trapping magnetic flux. Nonetheless, the results of Fig. 3c show that quantized currents can be generated using this configuration. It was used for the calibration of the DA. The best accuracy tests reported in Figs. 2, 3a were done with the first configuration.
The PJBS is represented by a voltage source VDAC and a resistance R = 1 kΩ. The PJBS is connected to ground at the low potential side with of a capacitance C = 1 μF. Black lines : superconducting wires, gray lines : bias lines and connection to ground, blue lines : triple connection. a Configuration with n1 = 4096. b Configuration with n1 = 1920 (CCC and QHRS1 are not represented).
CCC for correction-free PQCG
Design
The new CCC (Fig. 8a, b) is made of 20 windings of 1, 1, 1, 2, 2, 16, 16, 16, 32, 64, 128, 128, 160, 160, 465, 465, 1600, 1600, 2065 and 2065 turns, with a total number of 8789 turns. They are embedded in a superconducting toroidal shield made of 150 μm thick Pb foils, forming three electrically isolated turns to prevent non-ideal behavior at the ends of the shield. The architecture is inspired from the design of a CCC used in a quantum Hall resistance bridge42 (enabling ratios close to 1.29), but with 5 additional windings. The triple connection is possible for the windings of 1, 2, 16, 128, 160, 465 and 1600 turns. The dimensions have been chosen to be mounted on a cryogenic probe designed to be compatible with a 70 mm diameter neck of a liquid He Dewar. The inner and outer diameter of the toroidal shield are 19 mm and 47 mm, respectively. The chimney is about 125 mm high. The CCC is enclosed in two successive 0.5 mm thick Pb superconducting cylindrical screens and in a Cryoperm shield surrounding the whole, corresponding to an expected overall magnetic attenuation of about 200 dB. It is equipped with a Quantum Design Inc. DC SQUID, placed in a separate superconducting Nb shield, and coupled to the CCC via a superconducting flux transformer composed of a wire wound sensing coil placed as close as possible to the inner surface of the CCC. The CCC sensitivity γCCC = 8 μA ⋅ turn/ϕ0 has been maximized with a sensing coil of 9 turns compatible with the geometrical constraints. The 20 windings are connected by 40 copper alloy wires (AWG 34) placed in a stainless steel shield.
a Photo of the CCC. b Photo of the CCC mounted in the cryogenic probe. (c) Magnetic flux noise measured at the output of the SQUID alone (black line) and coupled to the bare CCC with damping circuit in this work (blue line) and in best conditions (green line). Blue and green dash dotted lines present \(1/\sqrt{f}\) slopes.
Noise
Figure 8c shows the noise spectrum at the output of the SQUID. The base noise level of the CCC (green line) amounts to 10 \({{\upmu}} {{{\upphi }}}_{0}/\sqrt{{{{\rm{Hz}}}}}\) at 1 Hz in the best noise conditions, slightly higher than the white noise level of 3 \({{\upmu}} {{{\upphi }}}_{0}/\sqrt{{{{\rm{Hz}}}}}\) of the SQUID alone (black line). Below 1 Hz, one can observe an increased noise compatible with a 1/f noise contribution. Resonance peaks present at frequencies below 10 Hz are certainly due to an imperfect decoupling of the Dewar from the ground vibrations42.
Damping circuit
A damped resonance at 1.6 kHz is due to the use of a damping circuit to improve the stability of the feedback loop. The damping circuit is made of a CD = 100 nF capacitance at room temperature in series with a RD = 1 kΩ resistor and a ND = 2065 turns CCC winding at TD = 4.2 K. It strongly damps the CCC resonances (around 10 kHz) which are excited by the external noise captured. The counterpart is an increase of the magnetic flux noise detected by the SQUID around 1.6 kHz caused by the Johnson-Nyquist noise emitted by the resistor. However, placing the resistor at low temperature reduced the noise magnitude by a factor of ten compared to the previous experiment27, with a maximum flux noise of \(\frac{{N}_{{{{\rm{D}}}}}}{{\gamma }_{{{{\rm{CCC}}}}}}\sqrt{\frac{4{k}_{{{{\rm{B}}}}}{T}_{{{{\rm{D}}}}}}{{R}_{{{{\rm{D}}}}}}}\simeq\) 124 \({{\upmu}} {{{\upphi }}}_{0}/\sqrt{{{{\rm{Hz}}}}}\) at 1.6 kHz28. In the experimental conditions of this paper, the noise was measured slightly higher, as described by blue curve in Fig. 8c. The noise level at 1 Hz rises to about 20 μϕ0/Hz1/2.
Current sensitivity
It is related to the flux generated by the screening current circulating on the shield and can be estimated from the CCC noise spectrum and γCCC, it corresponds to 80 pA ⋅ turns/\(\sqrt{{{{\rm{Hz}}}}}\) at 1 Hz in the best conditions.
Accuracy
The CCC accuracy can be altered by magnetic flux leakage detected by the pickup coil. It can be tested by series opposition measurements of windings of identical number of turns. In the best noise conditions, using a measurement current of 30–100 mA, the ratio error on the number of turns δN/N was measured < 10−9 for N numbers considered in this paper. However, larger errors were measured at the time of this work leading to consider a CCC contribution to the PQCG Type B uncertainty of 2 × 10−9 (see Type B uncertainty section).
Multiple series connection
In the multiple series connection (see Figs. 1, 7a), the series resistances of the connections result in an effective resistance, which adds to the quantized Hall resistance26,32. This leads to a lower value of the quantized current IPQCG = Gn1ef1(1 − αn), where αn is positive and exponentially decreasing with the number of connections n. The series resistances r1, r2, r3, \({r}_{1}^{{\prime} }\), \({r}_{2}^{{\prime} }\) and \({r}_{3}^{{\prime} }\), as indicated in Fig. 7a, one calculates, using a Ricketts and Kemeny model53 of the Hall bar, \({\alpha }_{2}=\frac{{r}_{1}{r}_{2}}{{R}_{{{{\rm{H}}}}}^{2}}+\frac{{r}_{1}^{{\prime} }{r}_{2}^{{\prime} }}{{R}_{{{{\rm{H}}}}}^{2}}\) and \({\alpha }_{3}=\frac{{r}_{1}{r}_{2}{r}_{3}}{{R}_{{{{\rm{H}}}}}^{3}}+\frac{{r}_{1}^{{\prime} }{r}_{2}^{{\prime} }{r}_{3}^{{\prime} }}{{R}_{{{{\rm{H}}}}}^{3}}\) for the double series connection and the triple series connection, respectively. The series resistances to be considered are those of the QHRS contacts and cables, those of the long cables linking the quantum devices, and those of the different combinations of CCC windings necessary to obtain the desired number of turns N1. For all the experiments based on the triple connection technique, α3 is calculated below 1.5 × 10−10, except for the measurement performed with a 50 Ω resistor added in series with the first CCC winding (see Figs. 2e, 3b), which results in α3 = (1.44 ± 0.023) × 10−9. For the measurement reported in Fig. 2e and in Fig. 3b using the double connection scheme, One calculates α2 = (2.344 ± 0.037) × 10−7.
Uncertainties, weighted mean values, combined results and errors bars
Two types of measurement uncertainties are considered: the Type A uncertainties which are evaluated by statistical methods, the Type B uncertainties evaluated by others methods.
Type A uncertainty
The type A uncertainty for one measurement of ΔI/I, uA, is given by:
where
with \({w}_{+}^{{{{\rm{A}}}}}\) and \({w}_{-}^{{{{\rm{A}}}}}\) the standard uncertainties (coverage factor k = 1) of the mean of voltage series \(\Delta {V}_{{f}_{2}^{+}}\) and \(\Delta {V}_{{f}_{2}^{-}}\), respectively. Their calculation is legitimated by the time dependence of the Allan deviation which demonstrates a dominant white noise (Supplementary Fig. 2b).
To account for the intra-day noise observed in measurements reported in Fig. 2e, g, a Type A uncertainty, \({u}_{{{{\rm{id}}}}-{{{\rm{noise}}}}}^{{{{\rm{A}}}}}=1{0}^{-8}\), is added to each data points. Its value was determined so that the \({\chi }^{2}=\frac{1}{N}{\Sigma }_{1}^{N}\frac{{({X}_{i}-\overline{X})}^{2}}{u{({X}_{i})}^{2}} < 1\) criterion is fulfilled, where here N is the number of values, Xi the value, \(\overline{X}\) the weighted mean of all Xi and u(Xi) the standard uncertainty of the Xi. As suggested by an investigation of the quality of the power line of our laboratory, the observed intra-day noise could be caused by a recent increase of the electrical noise pollution in the main line (back-up power supply) supplying the laboratory. We have recently revealed an increase of the voltage noise between the ground and the neutral, probably caused by an increase power consumption by non-linear equipments. This random noise increase could lead to extra noise in SQUID measurements. This investigation is in course.
Type B uncertainty
Table 1 shows the basic Type B standard uncertainty budget of the accuracy test, which includes contributions of both the PQCG and the quantum voltmeter. The implementation of the triple series connection and the use of new measurement protocols based on the adjustment of the current using only the Josephson parameters have canceled (current divider) or strongly reduced (cable correction) the most important contributions of the previous experiment27. We would expect a total uncertainty below 1 × 10−9. However, it turns out that we measured, momentarily during the measurement campaign, CCC ratio errors slightly higher than the typical values. The ratio errors ranged from 2 × 10−9 to 0.1 × 10−9 for windings of number of turns from 128 to 1600, respectively, which were used either for N1 or N2 in the experiment reported here. Using a conservative approach, we have therefore considered here a Type B uncertainty of 2 × 10−9 for the CCC, which dominates the uncertainty budget. Other components were detailed in ref. 27. The SQUID electronic feedback is based on the same SQUID type and same pre-amplifier from Quantum Design. It is set in the same way using a 4.2V/ϕ0 close-loop gain whatever the number of turns N2 used. The VCCS current source preadjusts the output current, such that the SQUID feedbacks only on a small fraction lower than 2 × 10−5 × IPQCG (see ref. 27). Owing to the cable shields, leakage to ground are redirected to ground, i.e. parallel to CCC winding26. The current leakage error amount to r1/RL, which is below 7 × 10−12 in our experiments. It results a relative Type B uncertainty of 2 × 10−9 for the PQCG and a relative Type B uncertainty of uB = 2.1 × 10−9 for the accuracy test.
Weighted mean values
Weighted mean values, ΔI/IWM and their uncertainties, \({u}_{{{{\rm{WM}}}}}^{{{{\rm{A}}}}}\), are calculated from the ΔI/Ij series values and their Type A uncertainties.
In Fig. 2b, d, they are given by:
In Fig. 2f, h, they are given by:
where \({u}_{{{{\rm{id}}}}-{{{\rm{noise}}}}}^{{{{\rm{A}}}}}=1{0}^{-8}\) is the additional Type A component added to each data point to take account of the intra-day noise.
Combined results \(\overline{\Delta I/I}\)
For accuracy tests performed at 11.48 μA using N1 = 465 and at 45.94 μA using N1 = 465, the combined result \(\overline{\Delta I/I}\) is the mean value of the ΔI/IWM values obtained using the measurement protocol I+ and I−: \(\overline{\Delta I/I}=(\Delta I/{I}_{{{{\rm{WM}}}}}({{{{\rm{I}}}}}_{+})+\Delta I/{I}_{{{{\rm{WM}}}}}({{{{\rm{I}}}}}_{-}))/2\).
For accuracy tests performed at 5.74 μA using N1 = 160, at 45.94 μA using N1 = 160, at 57.42 μA using N1 = 160 and at 43.07 µA using n1 = 1920, the combined result \(\overline{{\Delta }I/I}\) is the weighted mean value calculated from the values (ΔI/IWM(I+) + ΔI/IWM(I−))/2 and ΔI/IWM(I±) and their respective Type A uncertainties.
Combined uncertainties
The combined uncertainty, \({u}_{{{{\rm{WM}}}}}^{{{{\rm{c}}}}}\), is given by \({u}_{{{{\rm{WM}}}}}^{{{{\rm{c}}}}}=\sqrt{{({u}_{{{{\rm{WM}}}}}^{{{{\rm{A}}}}})}^{2}+{({u}^{{{{\rm{B}}}}})}^{2}}\). The combined uncertainty of \(\overline{\Delta I/I}\) is given by \({u}^{{{{\rm{c}}}}}= \sqrt{{\left({u}_{\overline{\Delta I/I}}^{{{{\rm{A}}}}}\right)}^{2}+{({u}^{{{{\rm{B}}}}})}^{2}}\)
Error bars
Error bars in the different figures represent measurement uncertainties corresponding to one standard deviation (i.e. k = 1). This means an interval of confidence of 68% if a gaussian distribution law is assumed. These measurement uncertainties are either Type A uncertainties or combined uncertainties.
Figure 2 | In a, c, e and g, error bars correspond to only Type A uncertainties, uA. In b, d, f and g, errors bars correspond to combined uncertainties, \({u}_{{{\rm{WM}}}}^{{{\rm{c}}}}\) |
Figure 3 | In a, error bars correspond to combined standard uncertainties \(\sqrt{{({u}^{{{\rm{A}}}})}^{2}+{({u}^{{{\rm{B}}}})}^{2}}\) In b, error bars correspond to the combination of Type A standard uncertainties according to \(\sqrt{{({u}^{{{\rm{A}}}})}^{2}+{u}^{{{\rm{A}}}}{(\Delta {I}_{{{\rm{ref}}}}/{I}_{{{\rm{ref}}}})}^{2}}\) In c, error bars correspond to uA. |
Figure 4 | In a and b, errors bars correspond to uncertainties \(u({u}_{\exp }^{{{\rm{A}}}})\), which are standard deviations of uA values, and not standard deviations of the means, in order to reflect the ranges over which the noise levels vary. |
Figure 5 | In a and b, error bars correspond to Type A standard uncertainties. |
Figure 6 | In inset of Fig. 6a, error bars correspond to combined standard uncertainties, uc. |
Experimental settings of the PQCG used during the accuracy tests
All accuracy tests are performed using the PQCG settings reported in Table 2, except one measurement at 5.74 μA using frequencies f1 = f2 = 70.02 GHz (see Fig. 2a) and one measurement at 45.94 μA which uses different frequencies f1 = 70 GHz and f2 = 69.999976 GHz, respectively, to accommodate for the deviation of a few parts in 107 of the PQCG current from equation (1) caused by the implementation of the double connection only (see Fig. 2e). Measurements are carried out with Δf=40 or 80 kHz.
Experimental, \({u}_{\exp }^{{{{\rm{A}}}}}\) and calculated, \({u}_{{{{\rm{calc}}}}}^{{{{\rm{A}}}}}\), standard uncertainties
The uncertainties \({u}_{\exp }^{{{{\rm{A}}}}}({\tau }_{m})\) are calculated by averaging the uncertainty values, uA, of each series, after normalization to the same measurement time τm = Nsτ0 where Ns = 16 is the number of sequences, and to the same measurement protocol I±. The standard deviation, \(u({u}_{\exp }^{{{{\rm{A}}}}})\), is calculated from the different values of a series. \({u}_{{{{\rm{calc}}}}}^{{{{\rm{A}}}}}({\tau }_{m})\) is calculated using the relationship \({u}_{{{{\rm{calc}}}}}^{{{{\rm{A}}}}}({\tau }_{m}={N}_{s}{\tau }_{0})=\frac{1}{\sqrt{{N}_{s}}}\sqrt{\frac{3}{8}}\sqrt{\frac{{S}_{\Delta I/I}}{2{\tau }_{A}}}\), where τA = 12 s is the acquisition time for one single voltage measurement (Supplementary Fig. 2a) and \(\sqrt{{S}_{\Delta I/I}}\) is the noise density of ΔI/I. More precisely, \(\sqrt{\frac{{S}_{\Delta I/I}}{2{\tau }_{A}}}\) is the relative standard deviation corresponding to the acquisition time τA assuming an effective white noise density. The pre-factor \(\sqrt{\frac{3}{8}}\) comes from the combination of the standard deviations corresponding to the measurement protocol I±, where the voltage of each sequence is obtained by [(ΔV(+I)1 + ΔV(+I)3)/2 − ΔV(−I)2], with ΔV(+I)1, ΔV(−I)2 and ΔV(+I)3 three successive voltage acquisitions, performed with positive current, then negative current, and then positive current (Supplementary Fig. 2a). Finally, the factor \(\frac{1}{\sqrt{{N}_{s}}}\) comes from the white noise hypothesis justified by Supplementary Fig. 2b. The noise density \(\sqrt{{S}_{\Delta I/I}}\) is given by:
where kB is the Boltzmann constant and T=1.3 K. Three noise contributions are considered: the voltage noise power spectral density, SV, of the quantum voltmeter, which includes the noise of the null detector and some external voltage noise captured, the magnetic flux noise power spectral density, Sϕ, detected by the SQUID, which includes the SQUID noise and some external magnetic flux noise captured and the Johnson-Nyquist noise power spectral density emitted by the resistor R1 in the primary loop. The Johnson-Nyquist noise of the resistor R2 is included in SV. The third term contributes to \(\sqrt{{S}_{\Delta I/I}}\) by about 1.6 × 10−9/Hz1/2 for I1 = 45.94 μA, leading to a negligible uncertainty contribution of 5 × 10−11 for a measurement time τm = 16τ0. This third term is therefore not considered in our calculations. Values reported in Fig. 2e and f are calculated using a voltage noise of \(\sqrt{{S}_{V}}=28\,{{{\rm{nV}}}}/{{{{\rm{Hz}}}}}^{1/2}\) and a magnetic flux noise detected by the SQUID of \(\sqrt{{S}_{{{\upphi }}}}=6.2\times 1{0}^{-5}{{{\upphi }}}_{0}/{{{{\rm{Hz}}}}}^{1/2}\).
Measurement protocol for the ammeter calibration
Calibrations of the ammeter HP3458A are performed using settings of the PQCG reported in Supplementary Table. 1. In these experiments, output current values are changed by varying both the gain G and the frequency f1. Using the PQCG to perform ammeter calibration consists in replacing QHRS2 by the device under test and removing the quantum voltmeter. The connection is done through a low-pass filter (highly insulated PTFE 100 nF on the differential input). A common mode torus has also been introduced to minimize the noise. Supplementary Fig. 3a shows recordings by the ammeter (HP3458A) as a function of time for several alternations of IPQCG at 107.666272 μA using the measurement protocol I+. The acquisition time, the waiting time are of 10 s and 2 s, respectively. The measured current is determined from the average of the values obtained for several measurement groups. The time dependence of the Allan deviation, reported in Supplementary Fig. 3b, shows that the standard deviation of the mean is a relevant estimate of the Type A relative uncertainty at τm = 144 s. An uncertainty of 2 × 10−7 is typically achieved after a total measurement time of 144 s.
References
BIPM, The International System of units (SI) (http://www.bipm.org/en/si/, Sèvres, 9th edition, 2019).
Poirier, W., Djordjevic, S., Schopfer, F. & Thévenot, O. The ampere and the electrical units in the quantum era. C. R. Phys. 20, 92 (2019).
Pothier, H., Lafarge, P., Urbina, C., Estève, D. & Devoret, M. H. Single-electron pump based on charging effects. Eur. Phys. Lett. 17, 249 (1992).
Pekola, J. P. et al. Single-electron current sources: Toward a refined definition of the ampere. Rev. Mod. Phys. 85, 1421–1472 (2013).
Scherer, H. J. & Schumacher, H. W. Single-electron pumps and quantum current metrology in the revised SI. Ann. der Phys. 531, 1800371 (2019).
Keller, M. W., Eichenberger, A. L., Martinis, J. M. & Zimmermann, N. M. A capacitance standard based on counting electrons. Science 285, 1706 (1999).
Pekola, J. P. et al. Hybrid single-electron transistor as a source of quantized electric current. Nat. Phys. 4, 120 (2008).
Camarota, B. et al. Electron counting capacitance standard with an improved five-junction R-pump. Metrologia 49, 8–14 (2012).
Giblin, S. P. et al. Towards a quantum representation of the ampere using single electron pumps. Nat. Commun. 3, 930 (2012).
Stein, F. et al. Validation of a quantized-current source with 0.2 ppm uncertainty. Appl. Phys. Lett. 107, 103501 (2015).
Stein, F. et al. Robustness of single-electron pumps at sub-ppm current accuracy level. Metrologia 54, S1 (2017).
Yamahata, G., Giblin, S. P., Kataoka, M., Karasawa, T. & Fujiwara, A. Gigahertz single-electron pumping in silicon with an accuracy better than 9.2 parts in 107. Appl. Phys. Lett. 109, 013101 (2016).
Zhao, R. et al. Thermal-error regime in high-accuracy gigahertz single-electron pumping. Phys. Rev. Appl. 8, 044021 (2017).
Bae, M.-H. et al. Precision measurement of single-electron current with quantized Hall array resistance and Josephson voltage. Metrologia 57, 065025 (2020).
Kataoka, M. et al. Tunable nonadiabatic excitation in a single-electron quantum dot. Phys. Rev. Lett. 106, 126801 (2011).
Ahn, Y.-H. et al. Upper frequency limit depending on potential shape in a QD-based single electron pump. J. Appl. Phys. 122, 194502 (2017).
Shaikhaidarov, R.-S. et al. Quantized current steps due to the a.c. coherent quantum phase-slip effect. Nature 608, 45 (2022).
Crescini, N. et al. Evidence of dual shapiro steps in a Josephson junction array. Nat. Phys. 19, 851 (2023).
Kaap, F., Kissling, C., Gaydamachenko, V., Grünhaupt, L. & Lotkhov, S. Demonstration of dual Shapiro steps in small Josephson junctions. Nat. commun. 15, 8726 (2024).
Kurilovich, V. D., Remez, B. & Glazman, L. I. Quantum theory of bloch oscillations in a resistively shunted transmon. Preprint at https://doi.org/10.48550/arXiv.2403.04624 (2024).
Josephson, B. D. Possible new effects in superconductive tunnelling. Phys. Lett. 1, 251–253 (1962).
von Klitzing, K., Dorda, G. & Pepper, M. New method for high-accuracy determination of the fine structure constant based on quantized Hall resistance. Phys. Rev. Lett. 45, 494–497 (1980).
Shapiro, S. Josephson currents in superconducting tunneling: the effect of microwaves and another observations. Phys. Rev. Lett. 11, 80–82 (1963).
Chae, D.-H., Kim, M.-S., Oe, T. & Kaneko, N.-H. Series connection of quantum Hall resistance array and programmable Josephson voltage standard for current generation at one microampere. Metrologia 59, 065011 (2022).
Kaneko, N.-H., Tanaka, T. & Okazaki, Y. Perspectives of the generation and measurement of small electric currents. Meas. Sci. Technol. 35, 011001 (2024).
Poirier, W., Lafont, F., Djordjevic, S., Schopfer, F. & Devoille, L. A programmable quantum current standard from the Josephson and the quantum Hall effects. J. Appl. Phys. 115, 044509 (2014).
Brun-Picard, J., Djordjevic, S., Leprat, D., Schopfer, F. & Poirier, W. Practical quantum realization of the ampere from the elementary charge. Phys. Rev. X 6, 041051 (2016).
Djordjevic, S., Behr, R., Drung, D., Götz, M. & Poirier, W. Improvements of the programmable quantum current generator for better traceability of electrical current measurements. Metrologia 58, 045005 (2021).
Behr, R., Kieler, O., Kohlmann, J., Müller, F. & Palafox, L. Development and metrological applications of Josephson arrays at ptb. Meas. Sci. Technol. 23, 124002 (2012).
Piquemal, F. et al. Report on a joint BIPM-EUROMET project for the fabrication of QHE samples by the LEPs. IEEE Trans. Instrum. Meas. 42, 264 (1993).
Harvey, I. K. A precise low temperature dc ratio transformer. Rev. Sci. Instrum. 43, 1626–1629 (1972).
Delahaye, F. Series and parallel connection of multiterminal quantum Hall-effect devices. J. Appl. Phys. 73, 7914 (1993).
Buttiker, M. Absence of backscattering in the quantum Hall effect in multiprobe conductors. Phys. Rev. B 38, 9375 (1988).
Drung, D. & Storm, J.-H. Ultralow-noise chopper amplifier with low input charge injection. IEEE Trans. Instrum. Meas. 60, 2347 (2011).
Poirier, W., Bounouh, A., Piquemal, F. & André, J. P. A new generation of QHARS: discussion about the technical criteria for quantization. Metrologia 41, 285 (2004).
Delahaye, F. & Jeckelmann, B. Revised technical guidelines for reliable dc measurements of the quantized Hall resistance. Metrologia 40, 217 (2003).
Giblin, S. P. Re-evaluation of uncertainty for calibration of 100 MΩ and 1 GΩ resistors at NPL. Metrologia 56, 015014 (2019).
Scherer, H., Drung, D., Krause, C., Götz, M. & Becker, U. Electrometer calibration with sub-part-per-million uncertainty. IEEE Trans. Instrum. Meas. 68, 1887 (2019).
Chae, D.-H., Kim, M.-S., Kim, W.-S., Oe, T. & Kaneko, N.-H. Quantum mechanical current-to-voltage conversion with quantum Hall resistance array. Metrologia 57, 025004 (2020).
Lee, J. et al. From ac quantum voltmeter to quantum calibrator, in https://doi.org/10.1109/CPEM.2016.7540470CPEM 2016 pp. 1–2 (2016).
Likharev, K. K. & Zorin, A. B. Theory of the bloch-wave oscillations in small Josephson junctions. J. Low. Temp. Phys. 59, 347 (1985).
Poirier, W., Leprat, D. & Schopfer, F. A resistance bridge based on a cryogenic current comparator achieving sub-10−9 measurement uncertainties. IEEE Trans. Instrum. Meas. 70, 1–14 (2021).
Lafont, F. et al. Quantum Hall resistance standards from graphene grown by chemical vapour deposition on silicon carbide. Nat. Commun. 6, 6806 (2015).
Ribeiro-Palau, R. et al. Quantum Hall resistance standard in graphene devices under relaxed experimental conditions. Nat. Nano. 10, 965–971 (2015).
Panna, A. R. et al. Graphene quantum Hall effect parallel resistance arrays. Phys. Rev. B 103, 075408 (2021).
He, H. et al. Accurate graphene quantum Hall arrays for the new international system of units. Nat. Commun. 13, 6933 (2023).
Fox, E. J. et al. Part-per-million quantization and current-induced breakdown of the quantum anomalous Hall effect. Phys. Rev. B 98, 075145 (2018).
Götz, M. et al. Precision measurement of the quantized anomalous Hall resistance at zero magnetic field. Appl. Phys. Lett. 112, 072102 (2018).
Okazaki, Y. et al. Quantum anomalous Hall effect with a permanent magnet defines a quantum resistance standard. Nat. Phys. 18, 25–29 (2022).
Patel, D. K. et al. A zero external magnetic field quantum standard of resistance at the 10−9 level. Nat. Electron. 7, 1111–1116 (2024).
Martini, M. et al. Twisted cuprate van der waals heterostructures with controlled Josephson coupling. Mater. Today 67, 106 (2024).
Rüfenacht, A., Flowers-Jacobs, N. E. & Benz, S. P. Impact of the latest generation of Josephson voltage standards in ac and dc electric metrology. Metrologia 55, S152 (2018).
Ricketts, B. W. & Kemeny, P. C. Quantum Hall effect devices as circuit elements. J. Phys. D. 21, 483 (1988).
Keller, M. W., Zimmermann, N. M. & Eichenberger, A. L. Uncertainty budget for the NIST electron counting capacitance standard, ECCS-1. Metrologia 44, 505 (2007).
The BIPM key comparison database (KCDB), calibration and measurement capabilities - CMCs (appendix c), dc current database. https://kcdb.bipm.org (2024).
Acknowledgements
We wish to acknowledge Mohammed Mghalfi for its technical support. We thank Daniel Estève (CEA/SPEC, France), Christian Glattli (CEA/SPEC, France), Yannick De Wilde (ESPCI, France) and Almazbek Imanaliev (LNE, France) for their critical reading and comments. Part of this work was supported by the project 23FUN05 AQuanTEC. The project 23FUN05 AQuanTEC has received funding from the European Partnership on Metrology, co-financed from the European Union’s Horizon Europe Research and Innovation Program and by the Participating States.
Author information
Authors and Affiliations
Contributions
S. D. and W. P. planned the experiments. R. B performed the cabling and the characterizations of the programmable Josephson standards fabricated by PTB. S. D. and W. P. developed the instrumentation, conducted the electrical metrological measurements, analyzed the data and wrote the paper. All authors contributed to the final version.
Corresponding author
Ethics declarations
Competing interests
The authors declare no competing interests.
Peer review
Peer review information
Nature Communications thanks Nobu-Hisa Kaneko, and the other, anonymous, reviewers for their contribution to the peer review of this work. A peer review file is available.
Additional information
Publisher’s note Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Supplementary information
Source data
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International License, which permits any non-commercial use, sharing, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if you modified the licensed material. You do not have permission under this licence to share adapted material derived from this article or parts of it. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by-nc-nd/4.0/.
About this article
Cite this article
Djordjevic, S., Behr, R. & Poirier, W. A primary quantum current standard based on the Josephson and the quantum Hall effects. Nat Commun 16, 1447 (2025). https://doi.org/10.1038/s41467-025-56413-9
Received:
Accepted:
Published:
DOI: https://doi.org/10.1038/s41467-025-56413-9