Table 2 Experimental settings of the PQCG for the accuracy tests

From: A primary quantum current standard based on the Josephson and the quantum Hall effects

Current

IPQCG

n1

f1 = f2

I1

G=N1/N2

n2

V2

value

(μA)

 

(GHz)

(μA)

  

(V)

@ 5.74

5.742201056

4096

70

45.94

160/1280

512

0.074

@11.48

11.48440211

4096

70

45.94

465/1860

1024

0.148

@43.07

43.06650792

1920

70

21.53

160/80

3840

0.556

@45.94

45.93760845

4096

70

45.94

160/160

4096

0.593

@45.94

45.93760845

4096

70

45.94

465/465

4096

0.593

@57.42

57.42201056

4096

70

45.94

160/128

5120

0.741