Table 1 The Information on Data Acquisition.

From: Diff5T: Benchmarking human brain diffusion MRI with an extensive 5.0 Tesla k-space and spatial dataset

Parameters

dMRI

T1-weigted

T2-weigted

Sequence

EPI

GRE-FSP

FSE-MATRIX3D

Acquisition orientation

Axial

Sagittal

Sagittal

TR / TE (ms)

8277 / 67.9

10.1/3.4

3000/421.68

FA (degree)

90

9

Max = 160, Min = 23

Echo spacing (ms)

0.8

/

5.02

ETL

66

165

180

Band width (Hz/Pixel)

1850

160

450

In-plane resolution (RO × PE mm2)

1.2 × 1.2

0.5 × 0.5

0.5 × 0.5

Acquisition matrix (RO × PE)

176 × 176

512 × 512

512 × 424

Slice thickness (mm)

1.2

0.5

0.5

Number of slices

114

300

300

Multi-band factor

2

/

/

In-plane acceleration factor

2

/

/

Partial Fourier

6/8

/

/

Acceleration method

GRAPPA

uCS

uCS

Combined Acceleration factor

4

3

3

Channel combine mode

Adaptive combine

Adaptive combine

Adaptive combine

Fat suppression

Fat Sat

Off

Off

Intensity uniformity correction

Off

Combined field

Combined field

Distortion correction

Off

2D

2D

Image Filtering

Off

Smoothing: 5

Enhancement: 1

Edge Smoothing: 4

Smoothing: 5

Enhancement: 3

Edge Smoothing: 3

k-space filtering (mode/intensity)

Standard / High

Default / Medium

Optimized / Medium

Other

b = 0, 21 volumes (6 AP, 15 PA)

b = 1000 s/mm2, 90 directions (PA)

b = 2000 s/mm2, 90 directions (PA)

b = 3000 s/mm2, 90 directions (PA)

TI = 1000 ms

 

Scanning time

45.2 min

13.6 min

13.7 min

  1. Notes. EPI = echo planar imaging; GRE-FSP = fast spoiled gradient echo; FSE-MATRIX3D = 3D fast spin echo using modulated flip angle technique in refocused imaging with extended echo train; TR = repetition time; TE = echo time; TI = inversion time; FA = flip angle; ETL = echo train length; RO = read out; PE = phase encoding; uCS = United Imaging compressed sensing.