Abstract
The aim of this study was to correlate three commercially available copeptin assays and their diagnostic accuracy in the differential diagnosis of the polyuria-polydipsia syndrome. Analyzed data include repeated copeptin measures of 8 healthy volunteers and 40 patients with polyuria-polydipsia syndrome undergoing osmotic stimulation and of 40 patients hospitalized with pneumonia. Copeptin was measured using the automated Brahms KRYPTOR, the manual Brahms LIA and the manual Cloud Clone ELISA assay. Primary outcome was the interrater correlation coefficient (ICC) and diagnostic accuracy in the polyuria-polydipsia syndrome of the three assays. In healthy volunteers, there was a moderate correlation for the KRYPTOR and LIA (ICC 0.74; 95% CI 0.07 to 0.91), and a poor correlation for the KRYPTOR and ELISA (ICC 0.07; 95% CI − 0.06 to 0.29), as for the LIA and ELISA (ICC 0.04; 95% CI − 0.04 to 0.17). The KRYPTOR had the highest diagnostic accuracy (98% (95% CI 83 to100)), comparable to the LIA (88% (95% CI 74 to 100)), while the ELISA had a poor diagnostic accuracy (55% (95% CI 34 to 68)) in the differential diagnosis of the polyuria-polydipsia syndrome. The KRYPTOR and LIA yield comparable copeptin concentrations and high diagnostic accuracy, while the ELISA correlates poorly with the other two assays and shows a poor diagnostic accuracy for polyuria-polydipsia patients. The current copeptin cut-off is valid for the KRYPTOR and LIA assay. Our results indicate that interpretation with other assays should be performed with caution and separate validation studies are required before their use in differentiating patients with polyuria-polydipsia syndrome.
Trial registration: NCT02647736 January 6, 2016/NCT01940614 September 12, 2013/NCT00973154 September 9, 2009.
Similar content being viewed by others
Introduction
Copeptin (also known as CT-proAVP) is the c-terminal portion of pre-pro-vasopressin, the precursor of the pituitary hormone vasopressin, which is the main regulator hormone of the sodium-water homeostasis1,2,3,4. In comparison to the instable and difficult to measure hormone vasopressin, copeptin is easily handled, is stable at room temperature and provides a reliable measure of the bioactive hormone1. As copeptin mirrors the physiological and pathological features of the pituitary hormone vasopressin, despite its different half-time, it is considered a true surrogate marker for vasopressin and has been increasingly used in clinical research and routine instead5,6. There are two main clinical implications of copeptin: first, as copeptin and vasopressin correlate closely over a wide range of plasma osmolality, copeptin can be used as a marker for vasopressin activity in the sodium-water homeostasis5,6. In this regard, copeptin has been investigated in the differential diagnosis of the polyuria-polydipsia syndrome7,8,9. It has been shown that copeptin is able to differentiate between patients with central diabetes insipidus, i.e., lack of vasopressin, and primary polydipsia following the hypertonic saline or arginine infusion test with a higher diagnostic accuracy than the long-time used indirect water deprivation test8,9. Second, copeptin is elevated in critically ill patients, such as pneumonia, stroke or myocardial infarction10,11 and has been shown to predict outcome of these diseases, with higher copeptin levels correlating with an unfavorable outcome11,12.
As for many other hormones and biomarkers, different types of immunoassays exist to measure copeptin. While some assays show a high inter-assay correlation13, others may yield quite different results14. To the best of our knowledge, there is no study investigating the inter-assay correlation of different copeptin assays. This is especially important with regards to the copeptin cut-off used in the differential diagnosis of the polyuria-polydipsia syndrome.
Therefore, the aim of this study was to compare three commercially available copeptin assays in different settings: first, over a wide range of serum osmolality in healthy volunteers, second, in the differential diagnosis of the polyuria-polydipsia syndrome, and third, in critically ill patients with community acquired pneumonia.
Methods
Patients and controls
For this analysis, data was combined from three different studies: first, healthy volunteers undergoing a test protocol of osmotic stimulation with hypertonic saline infusion5; second, patients with polyuria-polydipsia syndrome undergoing a hypertonic saline infusion as differential diagnosis between patients with central diabetes insipidus and primary polydipsia8; third, severely ill patients admitted to the hospital with community acquired pneumonia15. Full details of the studies rationales, designs and statistical analyses have been published elsewhere5,8,15. All three studies were registered on ClinicalTrials.gov (NCT02647736/NCT01940614/NCT00973154), conducted in accordance to the declaration of Helsinki and Good Clinical Practice Guidelines, and approved by the local ethical committees of all participating sites (Switzerland: Ethics Committee Northwest and Central Switzerland (EKNZ), Ethics Committee of Berne (KEK); Germany: the Ethical Committee of the University of Leipzig, Ethics Committee of the Medical Faculty of the Julius-Maximilians-University Würzburg; Brazil: Ethics Committees of the Federal University of Minas Gerais). Written informed consent was obtained from each participant after full explanation of the purpose and nature of all procedures used.
Study procedure and expected copeptin concentrations in the original studies
Healthy volunteers
Participants were recruited at two tertiary care hospitals in Switzerland and Germany from September 2012 to July 20165. Participants received hypertonic saline infusion (3% NaCl) while plasma sodium and copeptin were measured every 30 min until plasma sodium increased to > 147 mmol/l5. Expected copeptin concentrations cover the range of normal plasma copeptin, i.e., 1.23 pmol/l to 40 pmol/l.
Polyuria-polydipsia syndrome
Patients with polyuria-polydipsia syndrome, i.e., patients with complete central diabetes insipidus or primary polydipsia, were recruited at 11 tertiary medical centers in Switzerland, Germany, and Brazil from July 2013 to June 20178. Patients underwent a test protocol with hypertonic saline infusion (3% NaCl). Blood for the analysis of copeptin was sampled at baseline and once plasma sodium was > 149 mmol/l. According to the results of the indirect classical water deprivation test, patients’ history and therapy response, patients were diagnosed by experienced and board certified endocrinologists. Expected copeptin concentrations cover the lower range of plasma copeptin, 1.23 pmol/l to 6 pmol/l, while stimulated plasma copeptin allows for the differential diagnosis of the polyuria-polydipsia syndrome.
Severely ill patients
Patients hospitalized with community acquired pneumonia were recruited in seven tertiary care hospitals in Switzerland from December 2009 to May 201415. Community acquired pneumonia was defined by a new infiltrate on chest radiograph and the presence of at least one of the following acute respiratory signs and symptoms: cough, sputum production, dyspnea, core body temperature of 38.0 °C or higher, auscultatory findings of abnormal breathing sounds or rales, leucocyte count > 10,000 cells per μl or < 4000 cells per μl. Patients had blood sampling for copeptin concentrations on admission to the medical emergency department15. Expected copeptin concentrations cover the range of high plasma copeptin, i.e., above 40 pmol/l.
Laboratory measurements and copeptin assays
In all three studies, blood for copeptin analysis was sampled in EDTA plasma tubes, immediately centrifuged at 4 °C for 10 min at 4000 rpm and stored at − 80 °C until analysis. Storage time for severely ill patients was longest (5–10 years), moderate in healthy volunteers (4–8 years) and shortest in patients with polyuria-polydipsia syndrome (3–7 years). Analysis for this study was performed for each assay in one batch with the same freeze–thaw cycles. Several “research use only” (RUO) and few CE marked (in vitro diagnostic, IVD) copeptin assays are commercially available. Copeptin was measured in duplicates with the Brahms Copeptin proAVP KRYPTOR assay (short KRYPTOR), the Brahms CT-proAVP LIA assay (short LIA) and the Cloud Clone CT-proAVP ELISA assay (short ELISA). The first two assays are CE marked while the Cloud Clone is an example of a RUO, which are mainly based on ELISA technology.
The KRYPTOR
The KRYPTOR is a CE marked automated immunofluorescent assay for the quantitative determination of copeptin. The measurement principle is based on Time-Resolved Amplified Cryptate Emission Technology (TRACE), which measures the signal that is emitted from an immunocomplex with a time delay. The fluorescent signal is emitted once the donor and acceptor antibodies form an immunocomplex with the antigen. The signal is proportional to the copeptin concentration.
According to the manufacturer’s instructions for use, the limit of detection was assessed as being 0.69 pmol/l, the intra-assay coefficient of variation as ranging from < 15% to < 4% for copeptin concentrations of 2 pmol/l to > 50 pmol/l and the inter-assay precision as ranging from < 18% to < 5% for copeptin concentrations from 2 pmol/l to > 50 pmol/l.
The LIA
The LIA is a CE marked immunoluminometric assay for quantitative determination of copeptin. Two antigen-specific antibodies bind copeptin at two different binding sites of which one is a labelled (chemiluminescent) tracer and the other is fixed to the inner walls of the reaction tubes (capture antibody; coated tube system). During incubation, both antibodies react with copeptin to form a sandwich complex. The tracer is quantified by measuring the luminescent signal and the signal intensity is proportional to the copeptin concentrations bound to the capture antibody.
According to the manufacturer’s instruction for use, the limit of detection was assessed as being 0.4 pmol/l, the intra-assay coefficient of variation as being < 5% for copeptin concentrations of > 2 pmol/l and the inter-assay coefficient of variation as being < 10% for copeptin concentrations > 2.5 pmol/l.
The ELISA
The ELISA is a “research use only” sandwich enzyme immunoassay for the in vitro quantitative measurement of copeptin. The microplate is pre-coated with an antibody specific to copeptin. During the incubation, the specific antibodies react with copeptin which leads to a color change. The color change is proportional to the copeptin concentrations and measured with a spectrophotometer at a wavelength of 450 nm. The copeptin concentration is determined by comparing the results to a standard curve.
According to the manufacturer’s manual, the limit of detection was assessed as being 1.4 pmol/l, the intra-assay coefficient of variation as being < 10% and the inter-assay coefficient of variation as being < 12% for the whole copeptin measuring range.
Statistical analysis
The full analysis set included the data where copeptin was measured with all three assays (n = 150). Results are shown as median and interquartile range (IQR) and number (n) with percentage (%), appropriately. Three-group comparison for paired samples was done using the Friedman test. Two-group posthoc comparison was done using the Wilcoxon signed-rank test for paired samples with p-value adjustment according to Bonferroni.
Assay agreement was assessed using the intraclass correlation coefficient (ICC), a correlation analysis used to assess the consistency of two observers measuring the same quantity, herein copeptin concentrations measured with the KRYPTOR, the LIA and the ELISA. An ICC < 0.5 was considered a poor correlation, an ICC between 0.50 and 0.75 was considered a moderate correlation, an ICC between 0.75 and 0.90 was considered a good correlation and an ICC > 0.90 was considered an excellent correlation16. Graphical representation of how well the three copeptin assays correlate and mean copeptin difference between the assays including the lower and upper confidence limits was performed using the Bland–Altman Plot and the Bland–Altman Statistics.
Differential diagnosis of patients with polyuria-polydipsia syndrome was conducted using the area under the receiver operating characteristic curve (AUC) to assess the sensitivity and specificity of each copeptin assay. The “diagnostic gold standard” was used with the final diagnosis of the previous publication using the validated copeptin cut off of 4.9 pmol/l8.
Statistical and graphical analyses were performed using the statistic program R Statistical Software17. Hypothesis testing was two-sided and p-values < 0.05 were considered statistically significant.
Results
Patients’ characteristics
We included 8 healthy volunteers with a total of 30 plasma copeptin measurements (8 at baseline and 22 with rising plasma osmolality until serum sodium > 147 pmol/l), 40 patients with polyuria-polydipsia syndrome, 20 patients with primary polydipsia and 20 patients with complete central diabetes insipidus, with a total of 80 copeptin measurements (40 basal and 40 osmotically stimulated copeptin measurements), and 40 severely ill patients with community acquired pneumonia, with a total of 40 copeptin measurements (all on admission to the medical emergency department). Baseline characteristics of the three patient cohorts are displayed in Table 1. The median plasma copeptin concentrations for all three assays are displayed in Table 2.
Different copeptin assays in healthy volunteers
Median plasma copeptin concentrations
There was a significant difference between the three assays (p-value < 0.001), with the LIA reporting the lowest and the ELISA reporting the highest copeptin concentrations (posthoc analysis: KRYPTOR vs LIA p-value < 0.001; KRYPTOR vs ELISA p-value < 0.001; LIA vs ELISA p-value < 0.001).
Intra-assay comparison
There was an excellent correlation between the duplicated copeptin measurements using the KRYPTOR (ICC 0.99 (95% CI 0.99 to 1), mean copeptin difference − 0.08 pmol/l, (95% limits − 0.75 to 0.48)); an excellent correlation using the LIA (ICC 0.90 (95% CI 0.81 to 0.95), mean copeptin difference − 0.55 pmol/l (95% limits − 6.69 to 5.59)); and a good correlation using the ELISA (ICC 0.81 (95% CI 0.63 to 0.90), mean copeptin difference 0.92 pmol/l (95% limits − 17.32 to 19.16)) (Fig. 1a–c).
Intra-assay difference in copeptin concentrations in healthy volunteers. Bland–Altman graph of intra-assay difference in copeptin concentrations in healthy volunteers for (a) the KRYPTOR, (b) the LIA and (c) the ELISA. The dotted middle line represents the mean difference in copeptin concentrations between the two different assays (a): − 0.08 pmol/l (95% limits − 0.75 to 0.48), (b): − 0.55 pmol/l (95% limits − 6.69 to 5.59), (c): 0.92 pmol/l (95% limits − 17.32 to 19.16)). The outer dotted lines represent the 95% confident interval limits of agreement. (a) KRYPTOR 1 vs KRYPTOR 2, (b) LIA 1 vs LIA 2, (c) ELISA 1 vs ELISA 2.
Inter-assay comparison
There was a moderate correlation between the mean copeptin concentrations using the KRYPTOR and the LIA (ICC 0.74 (95% CI 0.07–0.91), mean copeptin difference 5.51 pmol/l (95% limits − 4.16 to 15.18)); a poor correlation using the KRYPTOR and the ELISA (ICC 0.07 (95% CI − 0.06 to 0.29), mean copeptin difference − 30.8 pmol/l (95% limits − 60.63 to − 1.03)); and a poor correlation using the LIA and the ELISA (ICC 0.04 (95% CI − 0.04 to 0.17), mean copeptin difference − 36.34 pmol/l (95% limits − 63.37 to − 9.32)) (Fig. 2a–c).
Inter-assay difference in copeptin concentrations in healthy volunteers. Bland–Altman graph of inter-assay difference in copeptin concentrations in healthy volunteers for (a) the KRYPTOR vs LIA, (b) the KRYPTOR vs ELISA and (c) the LIA vs ELISA. The dotted middle line represents the mean difference in copeptin concentrations between the two different assays ((a): 5.51 pmol/l (95% limits − 4.16 to 15.18), (b): − 30.8 pmol/l (95% limits − 60.63 to − 1.03), (c): − 36.34 pmol/l (95% limits − 63.37 to − 9.32)). The outer dotted lines represent the 95% confident interval limits of agreement. (a) KRYPTOR vs LIA, (b) KRYPTOR vs ELISA, (c) LIA vs ELISA.
Different copeptin assays in the differential diagnosis of polyuria-polydipsia syndrome
Median plasma copeptin concentrations
There was a significant difference between the three assays (p-value < 0.001). The posthoc analysis indicated that there was no significant difference between the KRYPOTR and the LIA (p-value = 0.70) but that there was a significant difference between the KRYPOTR and the ELISA (p-value < 0.001) as well as the LIA and the ELISA (p-value < 0.001).
Central diabetes insipidus
There was no significant difference in basal nor stimulated plasma copeptin concentrations in patients with central diabetes insipidus between the KRYPTOR or the LIA (basal p-value = 0.52, stimulated p-value = 1), but the ELISA reported significantly higher copeptin concentrations compared to both assays (basal and stimulated p-value < 0.001).
Primary polydipsia
There was no significant difference in basal plasma copeptin concentrations in patients with primary polydipsia between the KRYPTOR and the LIA (p-value = 1), but the ELISA reported significantly higher copeptin concentrations compared to the other two assays (p-value < 0.001). There was no significant difference between any of the assays for stimulated copeptin concentrations in patients with primary polydipsia (p-value = 0.1) (Table 3).
Diagnostic accuracy
The overall diagnostic accuracy of the hypertonic saline infusion test with the previously established copeptin cut-off of 4.9 pmol/l7,8 in the differential diagnosis of the polyuria-polydipsia syndrome was poor using the ELISA (55% (95% CI 34–68) (Table 4, Supplementary Fig. S1). The sensitivity (correctly identifying patients as having central diabetes insipidus) was 10% (95% CI 1, 32) and the specificity (correctly identifying patients as having primary polydipsia was 100% (95% CI 83, 100) using the ELISA. The diagnostic accuracy, sensitivity and specificity of the KRYPTOR and LIA are comparable to what has been shown in previous studies by Timper et al. and Fenske et al.7,8.
Different copeptin assays in critically ill patients with community acquired pneumonia
Median plasma copeptin concentrations
There was a significant difference between the three assays (p-value = 0.025). The posthoc analysis indicated that there was a significant difference between the KRYPOTR and the LIA (p-value = 0.023) but no significant difference between the KRYPOTR and the ELISA (p-value = 1) nor the LIA and the ELISA (p-value = 0.71).
Intra-assay comparison
There was an excellent correlation between the duplicated copeptin measurements using the KRYPTOR (ICC 0.99 (95% CI 0.99 to 1), mean copeptin difference − 0.12 pmol/l (95% limits − 5.4 to 5.7)); an excellent correlation using the LIA (ICC 0.99 (95% CI 0.98 to 0.99), mean copeptin difference 3.43 pmol/l (95% limits − 14.42 to 21.80); and a good correlation using the ELISA (ICC 0.84 (95% CI 0.71 to 0.91), mean copeptin difference − 1.49 pmol/l (95% limits − 41.34 to 38.36)) (Supplementary Fig. S2a–c).
Inter-assay comparison
There was an excellent correlation between the mean copeptin measurements using the KRYPTOR and the LIA (ICC 0.94 (95% CI 0.89 to 0.97), mean copeptin difference 8.38 pmol/l (95% limits − 32.48 to 49.24)); a poor correlation between the KRYPTOR and the ELISA (ICC 0.44 (95% CI 0.16 to 0.66), mean copeptin difference 5.5 pmol/l (95% limits − 104.65 to 115.64)); and a poor correlation between the LIA and the ELISA (ICC 0.44 (95% CI 0.14 to 0.67), mean copeptin difference − 2.65 pmol/l (95% limits − 108.25 to 102.96)) (Supplementary Fig. S3a–c).
Discussion
This study has four main findings: first, copeptin concentrations measured by the KRYPTOR and the LIA assays showed a good correlation over a wide range of copeptin concentrations. Second, copeptin concentrations measured with the ELISA only correlate poorly with both the KRYPTOR and the LIA, especially in the lower detection range of copeptin concentrations. Third, the KRYPTOR and the LIA measured copeptin concentrations have a high diagnostic accuracy in the differential diagnosis of the polyuria-polydipsia syndrome using the existing copeptin cut-off of 4.9 pmol/l. Fourth, the ELISA measured copeptin concentrations have a poor diagnostic accuracy in this differential diagnosis with an especially low sensitivity in correctly diagnosing patients with central diabetes insipidus.
The first copeptin assay was a chemiluminescence sandwich immunoassay with coated tubes, developed in 2006 by Brahms GmbH, Hennigsdorf/Berlin, Germany1 and validated using healthy volunteers and severely ill patients with sepsis1. Over the past years, additional copeptin assays have been developed such as the automated KRYPTOR18,19 and various ELISA20,21,22 tests. Consequently, there are now several commercial assays increasingly being used but a systematic correlation of these different assays has not yet been performed. Here, we compared three commercially available assays. First, a Copeptin proAVP KRYPTOR test, an assay that has been used, amongst others, in two recently published studies in the differential diagnosis of the polyuria-polydipsia syndrome8,9,23,24. This test is most often used in clinical studies. Second, a LIA, an assay that has been used, amongst others, in the differential diagnosis of the polyuria-polydipsia syndrome7 and as outcome predictor of stroke and sepsis11,12,25. This test was one of the first copeptin assays and used in early clinical studies on copeptin. Third, an ELISA, an assay that has been used in several research studies such as a predictor of treatment response in children with nocturnal enuresis, in children exposed to maltreatment and antihypertensive treatment response22,26,27,28. Several ELISA assays have been developed by companies for research use only. The chosen ELISA is one of the most commonly used copeptin ELISA assays.
Our results indicate that the KRYPTOR and the LIA have a high inter-assay agreement, which means that they yield comparable copeptin concentrations when measuring the same sample. Even though the KRYPTOR compared to the LIA measures on average higher copeptin concentrations, the difference is not clinically significant in the lower and mid-copeptin range, as evident by basal copeptin concentrations in patients with central diabetes insipidus and primary polydipsia. With higher copeptin concentrations, the discrepancy between the KRYPTOR and the LIA becomes more significant. The discrepancy in the lower range could be caused by a different minimal detection limit, for both cases, extreme hormonal concentrations in general are more difficult to measure29,30,31. An alternative explanation for the discrepancy between the KRYPOTR and the LIA might be that there is currently no international standardization to calibrate assays measuring copeptin concentrations, as it exists for other assays. Furthermore, different technologies (TRACE vs chemiluminescence) are used for the KRYPTOR and LIA. These different technologies may lead to different results with higher variability over the range of copeptin concentrations. Despite these discrepancies in copeptin concentrations between the two assays, their diagnostic accuracy in the polyuria-polydipsia syndrome was excellent and there was no significant difference in patients with central diabetes insipidus or primary polydipsia. The copeptin cut-off of 4.9 pmol/l was established in a study using the LIA (sensitivity of 94.0%, specificity of 94.4%)7 and validated in a recently published multicenter study using the KRYPTOR8 (sensitivity of 93.2%, specificity of 100%). In the latter study, a posthoc analysis indicated that a copeptin cut-off of 6.5 pmol/l had an even higher accuracy (sensitivity of 94.9%, specificity of 100%), supporting our finding that the KRYPOTR measures slightly higher copeptin concentrations but that both assays can be used interchangeably.
The ELISA on the other hand correlates poorly over the wide range of copeptin concentrations and specifically in the differential diagnosis of the polyuria-polydipsia syndrome. This is especially true for the lower detection range (copeptin concentrations < 10 pmol/l). In the differential diagnosis of the polyuria-polydipsia syndrome this leads to falsely diagnosing patients with central diabetes insipidus as having primary polydipsia. A misclassification in this case may lead to false treatment and a fatal outcome as fluid restriction, the primary treatment option for patients with primary polydipsia, is contraindicated in patients with central diabetes insipidus. When this assay was used to assess copeptin concentrations in other indications than the polyuria-polydipsia syndrome, the copeptin results were generally in the range between 10 and 20 pmol/l, indicating a higher range of copeptin concentrations. The higher copeptin concentrations also in other studies may indicate that the ELISA in general measures higher copeptin concentrations22,26,27,28. The higher results could be related to a higher rate of copeptin fragments or other substances that are measured with the ELISA, which has been described for other assays14 or the used antibodies detect other epitopes on copeptin than the KRYPTOR or the LIA.
Since the copeptin cut-off of 4.9 pmol/l for the polyuria-polydipsia syndrome was evaluated using the LIA and validated using the KRYPTOR, it is obvious that these assays have the highest diagnostic reliability. The ELISA as an easy-to-handle assay can be done in most laboratories, in contrast to the automated Copeptin proAVP KRYPTOR test which requires the KRYPTOR instrument. Importantly, however, based on our data, the copeptin cut-off of 4.9 pmol/l which is now integrated into different diagnostic algorithms of the polyuria-polydipsia syndrome32 cannot be used with the tested ELISA. Separate validation studies are required before this assay can be applied in the differential diagnosis of central diabetes insipidus.
A few limitations should be mentioned: we investigated three commercially available copeptin assays but cannot extrapolate our results to other copeptin assays. We aimed to use the most commonly used copeptin assays to best represent different measurement techniques. Second, we did not include a study where copeptin concentrations was used as outcome predictor. However, the results from severely ill patients indicate that in the higher copeptin range, correlation between the ELISA and the other two assays is better than at the lower end. This is also in line with findings from other immunoassay comparisons, where detecting higher hormone concentrations seems more reliable than lower hormone concentrations. Third, we only included patients with complete, but not partial central diabetes insipidus. The principal aim of this study was to compare three commercially available assays in the diagnosis of central diabetes insipidus. To get the best possible answer, we have, in a first step, included only the clear cases which are best discriminated, i.e., complete central diabetes insipidus and primary polydipsia.
In summary, the existing cut-off for the differential diagnosis of central diabetes insipidus following hypertonic saline infusion are validated for the LIA and the KRYPTOR measured copeptin concentrations. Other copeptin assays may not yield the same results and hence results must be interpreted with caution. Use of other copeptin assays requires separate validation studies.
References
Morgenthaler, N. G., Struck, J., Alonso, C. & Bergmann, A. Assay for the measurement of copeptin, a stable peptide derived from the precursor of vasopressin. Clin. Chem. 52, 112–119 (2006).
Barat, C., Simpson, L. & Breslow, E. Properties of human vasopressin precursor constructs: Inefficient monomer folding in the absence of copeptin as a potential contributor to diabetes insipidus. Biochemistry 43, 8191–8203 (2004).
Augustine, V., Lee, S. & Oka, Y. Neural control and modulation of thirst, sodium appetite, and hunger. Cell 180, 25–32 (2020).
Pool, A.-H. et al. The cellular basis of distinct thirst modalities. Nature https://doi.org/10.1038/s41586-020-2821-8 (2020).
Fenske, W. K. et al. Release and decay kinetics of copeptin vs AVP in response to osmotic alterations in healthy volunteers. J. Clin. Endocrinol. Metab. 103, 505–513 (2018).
Balanescu, S. et al. Correlation of plasma copeptin and vasopressin concentrations in hypo-, iso-, and hyperosmolar States. J. Clin. Endocrinol. Metab. 96, 1046–1052 (2011).
Timper, K. et al. Diagnostic accuracy of copeptin in the differential diagnosis of the polyuria-polydipsia syndrome: A prospective multicenter study. J. Clin. Endocrinol. Metab. 100, 2268–2274 (2015).
Fenske, W. et al. A copeptin-based approach in the diagnosis of diabetes insipidus. N. Engl. J. Med. 379, 428–439 (2018).
Winzeler, B. et al. Arginine-stimulated copeptin measurements in the differential diagnosis of diabetes insipidus: A prospective diagnostic study. Lancet 394, 587–595 (2019).
Katan, M. et al. Copeptin, a stable peptide derived from the vasopressin precursor, correlates with the individual stress level. Neuro Endocrinol. Lett. 29, 341–346 (2008).
Struck, J., Morgenthaler, N. G. & Bergmann, A. Copeptin, a stable peptide derived from the vasopressin precursor, is elevated in serum of sepsis patients. Peptides 26, 2500–2504 (2005).
Urwyler, S. A. et al. Prognostic value of copeptin: One-year outcome in patients with acute stroke. Stroke 41, 1564–1567 (2010).
Boeddinghaus, J. et al. High-sensitivity cardiac troponin i assay for early diagnosis of acute myocardial infarction. Clin. Chem. 65, 893–904 (2019).
Leng, G. & Sabatier, N. Measuring oxytocin and vasopressin: Bioassays, immunoassays and random numbers. J. Neuroendocrinol. https://doi.org/10.1111/jne.12413 (2016).
Blum, C. A. et al. Adjunct prednisone therapy for patients with community-acquired pneumonia: A multicentre, double-blind, randomised, placebo-controlled trial. Lancet 385, 1511–1518 (2015).
Koo, T. K. & Li, M. Y. A guideline of selecting and reporting intraclass correlation coefficients for reliability research. J. Chiropr. Med. 15, 155–163 (2016).
R Core, T. R: A Language and Environment for Statistical Computing (2018).
Dupuy, A. M. et al. Analytical performances of the newly developed, fully automated Kryptor® Copeptin assay: Which impact factor for myocardial infarction rules out in the emergency department?. Clin. Lab. 58, 635–644 (2012).
Schlapbach, L. J. et al. Copeptin concentration in cord blood in infants with early-onset sepsis, chorioamnionitis and perinatal asphyxia. BMC Pediatr. https://doi.org/10.1186/1471-2431-11-38 (2011).
Abdel-Fattah, M. et al. Research letters. Indian Pediatr. 52, 807–809 (2015).
Dong, X. Q., Huang, M., Yang, S., Bin, Y. W. H. & Zhang, Z. Y. Copeptin is associated with mortality in patients with traumatic brain injury. J. Trauma Inj. Infect. Crit. Care 71, 1194–1198 (2011).
Hara, T. et al. Evaluation of urinary aquaporin 2 and plasma copeptin as biomarkers of effectiveness of desmopressin acetate for the treatment of monosymptomatic nocturnal enuresis. J. Urol. 198, 921–927 (2017).
Enhörning, S. et al. Water supplementation reduces copeptin and plasma glucose in adults with high copeptin: The H2O metabolism pilot study. J. Clin. Endocrinol. Metab. 104, 1917–1925 (2019).
Wannamethee, S. G. et al. Copeptin, insulin resistance, and risk of incident diabetes in older men. J. Clin. Endocrinol. Metab. 100, 3332–3339 (2015).
Stolz, D. et al. Copeptin, C-reactive protein, and procalcitonin as prognostic biomarkers in acute exacerbation of COPD. Chest 131, 1058–1067 (2007).
Coelho, R. et al. Serum copeptin in children exposed to maltreatment. Psychiatry Clin. Neurosci. 70, 434–441 (2016).
Wyskida, K. et al. N-terminal prohormone of brain natriuretic peptide but not C-terminal pre-pro vasopressin (Copeptin) level is associated with the response to antihypertensive therapy in haemodialysis patients. Kidney Blood Press. Res. 42, 1013–1022 (2017).
Pais, R. et al. Role of enteroendocrine L-cells in arginine vasopressin-mediated inhibition of colonic anion secretion. J. Physiol. 594, 4865–4878 (2016).
Mooney, M. H., Bergwerff, A. A., van Meeuwen, J. A., Luppa, P. B. & Elliott, C. T. Biosensor-based detection of reduced sex hormone-binding globulin binding capacities in response to growth-promoter administrations. Anal. Chim. Acta 637, 235–240 (2009).
Wang, R. et al. Highly sensitive detection of hormone estradiol E2 using surface-enhanced Raman scattering based immunoassays for the clinical diagnosis of precocious puberty. ACS Appl. Mater. Interfaces 8, 10665–10672 (2016).
Lin, Y., Liu, K., Wang, C., Li, L. & Liu, Y. Electrochemical immunosensor for detection of epidermal growth factor reaching lower detection limit: Toward oxidized glutathione as a more efficient blocking reagent for the antibody functionalized silver nanoparticles and antigen interaction. Anal. Chem. 87, 8047–8051 (2015).
Christ-Crain, M. et al. Diabetes insipidus. Nat. Rev. Dis. Primers. https://doi.org/10.1038/s41572-019-0103-2 (2019).
Acknowledgements
We thank all participants for their valuable contribution to our study. We thank the clinical and laboratory staff, especially Cemile Bathelt, Nina Hutter, Stefanie Hahner, Anna Schrader, Christian Haas and Mirjam Jäggy for their valuable help during the study conduction. We thank all participating sites and the personnel at University Hospital Basel, Switzerland; University Hospital Leipzig, Germany; University Hospital Würzburg, Germany; Faculdade de Medicina da Universidade Federal de Minas Gerais, Brazil; Kantonsspital St. Gallen, Switzerland; Inselspital Bern, Switzerland; Luzerner Kantonsspital, Switzerland; Kantonsspital Aarau, Switzerland; University Hospital Munich, Germany; University Hospital Hamburg- Eppendorf, Germany; and University Hospital Lübeck, Germany.
Funding
The study was investigator-initiated and was supported by a grant from the Swiss National Science Foundation to M. Christ-Crain (SNF-162608) and the University Hospital Basel, Switzerland. W. Fenske was supported by the Federal Ministry of Education and Research (BMBF) Germany (FKZ: 01EO1501) and Deutsche Forschungsgemeinschaft (DFG) Germany (AOBJ: 624808). C. O. Sailer was supported by a grant of the Swiss Academy of Medical Sciences and of the G. & J. Bangerter-Rhyner Foundation.
Author information
Authors and Affiliations
Contributions
C.O.S. designed the current study, analyzed the data and wrote the manuscript. J.R. planned and coordinated the study in patients with central diabetes insipidus and revised the manuscript. C.A.B. planned and coordinated the study in patients with pneumonia and revised the manuscript. I.S. planned and coordinated the study in healthy volunteers and revised the manuscript. J.A.M.T. analyzed data and revised the manuscript. W.F. planned and coordinated the study in health volunteers and central diabetes insipidus and revised the manuscript. M.C.C. planned, organized all three studies, designed the current study and revised the manuscript.
Corresponding author
Ethics declarations
Competing interests
MCC received speaking honoraria from Thermo Fisher Scientific, the manufacturer of Brahms Copeptin proAVP KRYPTOR and Brahms CT-proAVP LIA assay. The other authors declare no competing interests.
Additional information
Publisher's note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Supplementary Information
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Sailer, C.O., Refardt, J., Blum, C.A. et al. Validity of different copeptin assays in the differential diagnosis of the polyuria-polydipsia syndrome. Sci Rep 11, 10104 (2021). https://doi.org/10.1038/s41598-021-89505-9
Received:
Accepted:
Published:
DOI: https://doi.org/10.1038/s41598-021-89505-9
This article is cited by
-
Copeptin: a biochemical predictor of metabolic syndrome in children and adolescents with obesity
Egyptian Pediatric Association Gazette (2025)
-
Copeptin Association with Theory of Mind, Cognitive Functions and Symptom Severity in Children with Autism Spectrum Disorder
Bratislava Medical Journal (2025)
-
Copeptin in the diagnosis and management of renal tubular disorders
Pediatric Nephrology (2025)
-
Copeptin as an inflammatory marker in diagnosis and prognosis of neonatal sepsis
Beni-Suef University Journal of Basic and Applied Sciences (2024)
-
Arginine vasopressin deficiency: diagnosis, management and the relevance of oxytocin deficiency
Nature Reviews Endocrinology (2024)