Abstract
The effect of the skin–capsular distance (SCD) on the controlled attenuation parameter (CAP) for diagnosis of liver steatosis in patients with nonalcoholic fatty liver disease (NAFLD) remains unclear. The SCD was measured using B-mode ultrasound, and the CAP was measured using the M probe of FibroScan®. According to the indications of the M probe, 113 patients with an SCD of ≤ 25 mm were included in the present study. The association between the SCD and CAP was investigated, and the diagnostic performance of the SCD-adjusted CAP was tested. The SCD showed the most significant positive correlation with the CAP (ρ = 0.329, p < 0.001). In the multiple regression analysis, the SCD and serum albumin concentration were associated with the CAP, independent of pathological liver steatosis. According to the multivariate analysis, two different formulas were developed to obtain the adjusted CAP using the SCD and serum albumin concentration as follows: adjusted CAP (dB/m) = CAP − (5.26 × SCD) and adjusted CAP (dB/m) = CAP − (5.35 × SCD) − (25.77 × serum albumin concentration). The area under the receiver operating characteristic curve for diagnosis of a steatosis score ≥ 2 of adjusted CAP was 0.678 and 0.684 respectively, which were significantly greater than the original CAP (0.621: p = 0.030 and p = 0.024). The SCD is associated with the CAP independent of liver steatosis. Adjustment of the CAP using the SCD improves the diagnostic performance of the CAP in NAFLD.
Similar content being viewed by others
Introduction
Nonalcoholic fatty liver disease (NAFLD) is a chronic liver disease caused by obesity and metabolic syndrome1. The presence of liver steatosis is the principal criterion for the diagnosis of NAFLD. Pathologically, a ≥ 5% area of steatosis in the liver parenchyma is required for the diagnosis of NAFLD2. Ultrasound is commonly used to detect liver steatosis, especially in the primary care setting. Liver steatosis is diagnosed by the finding of a “bright liver,” which is characterized by increased hepatic echogenicity, poor penetration of the posterior segment of the right lobe, and poor or no visualization of the hepatic vessels and diaphragm3,4. Hepatorenal echo contrast is a well-validated ultrasonographic finding of fatty liver; a ≥ 20% area of steatosis in a liver specimen has a sensitivity of 96.4% and specificity of 97.8% for the diagnosis of fatty liver5. However, subjective diagnosis of liver steatosis based on B-mode imaging shows wide interobserver and intraobserver variability6. Because of the global high prevalence of NAFLD and limitations of liver biopsy, including high invasiveness, high cost, and diagnostic discordance among pathologists7,8,9, a simple, reliable, and quantitative screening procedure to identify liver steatosis is warranted.
The FibroScan® (Echosens, Paris, France) is an ultrasound-based transient elastography device that enables noninvasive evaluation of liver fibrosis through assessment of the liver stiffness measurement (LSM) and steatosis through assessment of the controlled attenuation parameter (CAP)10,11. The CAP represents the attenuation of the ultrasound beam in the liver. More severe liver steatosis produces a more attenuated ultrasound pulse passing through the liver tissue. The CAP value ranges from 100 to 400 dB/m based on this property12. CAP measurement is useful for diagnosing pathological liver steatosis in patients with NAFLD11,13,14,15. Moreover, Caussy et al.15 reported that the CAP showed representable diagnostic accuracy when the proton density fat fraction measured by magnetic resonance imaging (MRI-PDFF) was used as a gold standard.
In terms of the clinical implications, it is important to know that the LSM and CAP are affected by several factors in addition to fibrosis and steatosis, respectively16. The LSM is increased by inflammation, venous pressure, cholestasis, and amyloid deposition in the liver16,17,18,19,20,21,22. Additionally, the body mass index (BMI) is reportedly a significant confounding factor that increases the CAP11,16,23,24,25,26,27,28. A greater distance between the skin surface and the liver capsule (Supplementary Figure 1), termed the skin–capsular distance (SCD), increases the LSM because of multiple ultrasonic echo reflections. In one study, the LSM increased in patients with an SCD of ≥ 20 mm29. However, whether the SCD affects the CAP remains unclear. The SCD includes the thickness of the subcutaneous fat tissue. Therefore, we hypothesized that the SCD affects the CAP of obese patients with NALFD and that the diagnostic performance of CAP measurement improves by adjustment of the SCD. This study was performed to clarify whether the SCD affects the CAP and to generate a formula for calculation of the adjusted CAP.
Methods
Patients
We retrospectively enrolled 150 consecutive patients who visited Saga University Hospital from May 2017 to August 2019. All patients were diagnosed with NAFLD by liver biopsy, and underwent CAP measurement with the M probe or XL probe of the FibroScan®. According to the indications of the M probe (SCD of ≤ 25 mm), 37 patients with an SCD of > 25 mm were excluded and the data of 113 patients were finally analyzed in the current study.
Physical examination and serum biochemical measurements
Patients were diagnosed with diabetes mellitus if their fasting plasma glucose (FPG) concentration was > 126 mg/dL and/or they were undergoing treatment with antidiabetic drugs. Patients were diagnosed with hypertension if their systolic blood pressure was > 140 mmHg and/or diastolic blood pressure was > 90 mmHg and/or they were undergoing treatment with antihypertensive drugs. The patients’ body mass and height were measured, and the BMI was calculated as body mass (kg) divided by height squared (m2). Venous blood samples were obtained after overnight fasting and used to determine the following parameters by conventional laboratory techniques: blood cell counts, platelet count, FPG, and serum levels of aspartate aminotransferase, alanine transaminase, alkaline phosphatase, γ-glutamyl transpeptidase, total bilirubin, total protein, albumin, total cholesterol, triglycerides, high-density lipoprotein cholesterol, low-density lipoprotein cholesterol, ferritin, and type IV collagen 7 s.
Liver biopsy and histological assessment
Ultrasonography-guided liver biopsy was performed using a 16-gauge biopsy needle. All liver biopsy specimens were approximately ≥ 20 mm in length. Liver biopsy slides stained with hematoxylin–eosin and Azan stain were evaluated by a single experienced pathologist (S.A.) specializing in liver pathology. The pathologist was blinded to the clinical data. NAFLD was pathologically diagnosed if the steatosis area was > 5%. Hepatic steatosis, lobular inflammation, and hepatocyte ballooning were evaluated using the NAFLD activity score30. Liver fibrosis was classified according to Kleiner et al.30 and Brunt et al.31. Steatosis was assigned a score of 0, 1, 2, or 3 (score of 0, < 5%; score of 1, 5–33%; score of 2, 34–66%; and score of 3, > 66% of biopsy specimen affected), and fibrosis was assigned a score of 0, 1, 2, 3, or 4 (stage 0, no fibrosis; stage 1, perisinusoidal or periportal fibrosis; stage 2, perisinusoidal and portal/periportal fibrosis; stage 3, bridging fibrosis; and stage 4, cirrhosis).
LSM and CAP
Experienced operators who had performed at least 500 examinations assessed the CAP in the right liver lobe using the FibroScan® 502. Patients were examined after an overnight fast using the M probe or XL probe within 6 months before and after liver biopsy. First, we observed the optimal site for FibroScan® examination without blood vessels or any space-occupying lesions using B-mode ultrasound, and we measured the SCD. CAP measurements were performed using the FibroScan® M probe for patients with an SCD of ≤ 25 mm and the XL probe for patients with an SCD of > 25 mm and ≤ 35 mm until 10 valid measurements were obtained for each patient, and the median values were used to quantify liver steatosis. CAP measurement was expressed in dB/m. Based on previous reports, we defined measurement failure as examinations in which 10 valid CAP measurement were not obtained after 10 or more acquisitions32.
Statistical analysis
Correlations were tested using Spearman’s rank correlation coefficient. Single regression analysis and multiple regression analysis were performed to identify the factors associated with the CAP. Explanatory variables for the multiple regression analysis were chosen as follows: Model 1, pathological steatosis and CAP; Model 2, variables of Model 1 + BMI; Model 3, variables of Model 2 + significant variables in the single regression analysis. Using the partial regression coefficient (B) obtained in the multiple regression analysis (Models 1 and 3), the adjusted CAP was calculated as follows: Model 1-adjusted CAP = CAP − (B × SCD); Model 3-adjusted CAP (dB/m) = CAP − (B × SCD) − (B × serum albumin concentration). The diagnostic performance of the LSM and CAP measurements was determined using receiver operating characteristic (ROC) curves. The optimal cut-off values were chosen to maximize the sum of the sensitivity and specificity in Youden’s index33. Comparisons of the area under the ROC curve (AUROC) between the original CAP and adjusted CAP were performed using the DeLong test34. A p value of < 0.05 was considered statistically significant. All statistical analyses were conducted under the supervision of a statistical expert (A.K.), and were performed using JMP ver. 14.2.0 (SAS Institute Japan, Tokyo, Japan).
Ethics
The study protocol was approved by the Clinical Research Ethics Review Committee in Saga University Hospital, and informed consent was obtained from all the subjects. This study was performed in accordance with the principles of the 1975 Declaration of Helsinki, revised in 2013.
Results
Patients’ characteristics
The patients’ characteristics are summarized in Table 1. The median age was 63 years and 64 (57%) patients were female. Diabetes and hypertension were diagnosed in 75 (66%) and 63 (56%) patients, respectively. The prevalence of mild, moderate, and severe pathological liver steatosis was 64.6%, 25.7%, and 9.7%, respectively. The median SCD was 19.9 mm.
Correlation between CAP and various parameters
The BMI did not show a significant correlation with the CAP; however, the albumin concentration showed a significant positive correlation with the CAP (ρ = 0.245, p = 0.009) (Table 2). Among all parameters, the SCD showed the most significant positive correlation with the CAP (ρ = 0.329, p < 0.001) (Table 2 and Fig. 1). Correlations between the CAP and pathological findings were also tested (Supplementary Table 1). The CAP increased as the steatosis score increased. The CAP decreased as the ballooning score increased, but there was no significant difference between the scores. The CAP significantly decreased in the patients with cirrhosis.
Factors associated with CAP
Single and multiple regression analyses were performed to identify the factors associated with the CAP. Single regression analysis showed that sex (p = 0.027), BMI (p = 0.040), albumin (p = 0.002), pathological steatosis (p = 0.006), pathological ballooning (p = 0.011), and SCD (p < 0.001) were significantly associated with the CAP (Table 3). Multiple regression analysis demonstrated that the SCD and pathological steatosis were independently associated with the CAP in Model 1 (steatosis: β = 0.318, p < 0.001; SCD: β = 0.408, p < 0.001), Model 2 (steatosis: β = 0.334, p < 0.001; SCD: β = 0.474, p < 0.001), and Model 3 (steatosis: β = 0.306, p < 0.001; SCD: β = 0.415, p < 0.001) (Table 4). The albumin concentration was associated with the CAP in Model 3 (β = 0.227, p = 0.008). The BMI was not significant factors in any regression analysis models. According to the estimate in the multiple regression analysis, a 1-mm increase in the SCD resulted in a 5.26- to 6.11 dB/m increase in the CAP.
Adjusted CAP and its diagnostic performance
According to the partial regression coefficient in the multiple regression analysis of Model 1 and 3, we devised the following formulas to calculate the adjusted CAP: Model 1-adjusted CAP (dB/m) = CAP − (5.26 × SCD); Model 3-adjusted CAP (dB/m) = CAP − (5.35 × SCD) − (25.77 × serum albumin concentration). The AUROC of the Model 1-adjusted CAP for detecting patients with a steatosis score ≥ 2 was 0.678, which was significantly greater than the CAP without adjustment (0.621 vs. 0.678, p = 0.030) (Fig. 2a). According to Youden’s index, the cut-off value of the Model 1-adjusted CAP for the diagnosis of steatosis with a score of ≥ 2 was 165 dB/m with 85.0% sensitivity and 47.9% specificity, which was greater than that of the original CAP (Table 5). The positive predictive value and negative predictive value of the adjusted CAP were greater than those of the original CAP. Like the Model 1-adjusted CAP, the AUROC of the Model 3-adjusted CAP was also significantly greater than the CAP without adjustment (0.621 vs. 0.684, p = 0.024) (Fig. 2b). The cut-off value according to Youden’s index was 64 dB/m with 85.0% sensitivity and 52.1% specificity, which was greater than that of the original CAP (Table 5). There were no significant differences in the AUROC between the Model 1-adjusted CAP and the Model 3-adjusted CAP (0.678 vs. 0.684, p = 0.673).
Discussion
The present study showed that the SCD was independently associated with the CAP as measured by the M probe of the FibroScan® for the evaluation of liver steatosis in patients with NAFLD. The SCD was positively correlated with the CAP, suggesting that evaluation based on the CAP can result in overdiagnosis of liver steatosis in patients with NAFLD who have a greater SCD. Moreover, we generated a formula to calculate the adjusted CAP using the SCD, and the diagnostic performance of the CAP was improved. The diagnostic performance of the CAP varies among previous studies; the AUROC for the diagnosis of a steatosis score of ≥ 2 reported in individual studies ranges from 0.64 to 0.86 for the M probe16. The result of the present study was similar with these previous reports. Recent studies have shown that the diagnostic performance of the MRI-PDFF is superior to that of the CAP in detecting and grading liver steatosis in patients with NAFLD35,36. However, because of the high costs and long time required for the examination as well as the low availability of resources, it is difficult to measure the MRI-PDFF for screening and evaluation of disease progression in all patients with NAFLD, even in the hospital setting. Therefore, the diagnostic performance of the CAP must be improved for more accurate evaluation of steatosis.
The BMI has been considered to be a factor associated with the CAP in various chronic liver diseases, including NAFLD11,16,23,24,25,26,27,28. In the present study, however, the independent association between the BMI and CAP was negated when the SCD was included as an explanatory variable for the multiple regression analysis. Furthermore, the SCD showed a more significant correlation with the CAP than BMI. The present study raises the following question: Why is the SCD, not the BMI, independently associated with the CAP? We measured the SCD in the right intercostal region and defined it as the total distance of the tissue layers between the skin and liver surface. Because these tissue layers mainly comprise subcutaneous fat, the SCD can be used to represent the amount of subcutaneous fat. Adiposity is one of the major findings of obesity. Subcutaneous fat accumulation, as well as visceral fat accumulation, is associated with the pathogenesis of NALFD37,38,39,40,41. The distance of the abdominal subcutaneous fat measured by ultrasound is reportedly associated with the severity of liver steatosis in adults with NAFLD and is a more accurate predictor of severe liver steatosis than is the BMI40. Similarly, in obese children, the abdominal subcutaneous fat thickness measured by ultrasound is positively correlated with the severity of liver steatosis41. Like the visceral adipose tissue, the deep layer of the subcutaneous adipose tissue shows higher expression of inflammatory genes than does the superficial subcutaneous adipose tissue, and such gene expression is correlated with hepatic steatosis and fibrosis in patients with NAFLD42. These findings suggest that the subcutaneous fat is significantly associated with liver steatosis and that the SCD, including the subcutaneous fat thickness, might be more closely related to liver steatosis than the BMI, which is always affected by the lean mass including the skeletal muscle and bone.
Because the CAP is calculated by the attenuation of the ultrasound pulse, the SCD can directly affect the CAP. The FibroScan® manufacturer, Echosens, recommends that the FibroScan® probe selection be based on the SCD; specifically, the M probe should be used for patients with an SCD of ≤ 25 mm, and the XL probe for should be used for those with an SCD of > 25 mm and ≤ 35 mm. In 2015, Shen et al.43 first reported that the SCD affected the CAP in patients with NAFLD and chronic hepatitis B (n = 381). At that time, the CAP was available only for the M probe, and its diagnostic accuracy and association with the SCD had been tested using only the M probe. The association of the SCD with the CAP independent of pathological steatosis and CAP was overestimated in patients with an SCD of ≥ 25 mm. Today, the CAP is available for both the M probe and XL probe, and CAP measurement using the M probe in patients with an SCD of ≥ 25 mm is outside the indication for the M probe; such patients should be examined using the XL probe. Interestingly, the SCD affected the CAP even in patients whose SCD was within the indications of the individual probes. Our data suggest for the first time that the effect of the SCD on the CAP is not determined by a threshold such as the upper range of the probe indication of the SCD but is consecutively observed within the indication of the SCD in patients with NAFLD. More interestingly, the standard partial regression coefficient of SCD is higher than that of steatosis in any multiple regression analysis models, which means that the effect of SCD on CAP is greater than that of steatosis. Therefore, it makes sense that the diagnostic performance of CAP improves by using the adjusted CAP.
In the present study, the effect of the SCD on patients with an SCD exceeding the probe indication remains unclear; this can be considered a study limitation. According to a previous report, optimization of the cut-off value might be needed for CAP measurement in patients with an SCD greater than the probe indication, at least for the M probe measurement43. The adjusted CAP in the present study should be validated in further multicenter studies that include patients with an SCD exceeding the probe indication. Additionally, the adjusted CAP should be tested in a study cohort that includes subjects without NAFLD (healthy subjects) to determine its detecting capability in the primary care setting and during health check-ups. The effect of inter- and intra-operator error on the CAP measurement and the diagnostic performance of the adjusted CAP remains unclear in the present study. Recently, high inter- and intra-operator reproducibility of the FibroScan was demonstrated44. Further studies are required to test the effects of inter- and intra-operator error on the CAP and the adjusted CAP.
In conclusion, the SCD is positively correlated with the CAP and is associated with the CAP independent of steatosis. Adjustment of the CAP using the SCD improves the diagnostic performance of the CAP in patients with NAFLD.
References
Younossi, Z. et al. Global perspectives on non-alcoholic fatty liver disease and non-alcoholic steatohepatitis. Hepatology 69, 2672–2682 (2019).
Bedossa, P. et al. Histopathological algorithm and scoring system for evaluation of liver lesions in morbidly obese patients. Hepatology 56, 1751–1759 (2012).
Sanyal, A. J. AGA technical review on nonalcoholic fatty liver disease. Gastroenterology 123, 1705–1725 (2002).
Joseph, A. E., Dewbury, K. C. & McGuire, P. G. Ultrasound in the detection of chronic liver disease (the “bright liver”). Br. J. Radiol. 52(615), 184–188 (1979).
Dasarathy, S. et al. Validity of real time ultrasound in the diagnosis of hepatic steatosis: A prospective study. J. Hepatol. 51, 1061–1067 (2009).
Strauss, S. et al. Interobserver and intraobserver variability in the sonographic assessment of fatty liver. Am. J. Roentgenol. 189(6), W320–W323 (2007).
Piccinino, F. et al. Complications following percutaneous liver biopsy. J. Hepatol. 2, 165–173 (1986).
Ratziu, V. et al. Sampling variability of liver biopsy in nonalcoholic fatty liver disease. Gastroenterology 128, 1898–1906 (2005).
Kuwashiro, T. et al. Discordant pathological diagnosis of non-alcoholic fatty liver disease: A prospective multicenter study. JGH Open 4, 497–502 (2019).
Yoneda, M. et al. Transient elastography in patients with non-alcoholic fatty liver disease (NAFLD). Gut 56, 1330–1331 (2007).
Chan, W. K., Nik Mustapha, N. R. & Mahadeva, S. Controlled attenuation parameter for the detection and quantification of hepatic steatosis in nonalcoholic fatty liver disease. J. Gastroenterol. Hepatol. 29, 1470–1476 (2014).
Sasso, M. et al. Controlled attenuation parameter (CAP): A Novel VCTETM guided ultrasonic attenuation measurement for the evaluation of hepatic steatosis: Preliminary study and validation in a cohort of patients with chronic liver disease from various causes. Ultrasound Med. Biol. 36, 1825–1835 (2010).
Oeda, S. et al. Accuracy of liver stiffness measurement and controlled attenuation parameter using FibroScan®M/XL probes to diagnose liver fibrosis and steatosis in patients with nonalcoholic fatty liver disease: A multicenter prospective study. J. Gastroenterol. 55, 428–440 (2019).
Shi, K. Q. et al. Controlled attenuation parameter for the detection of steatosis severity in chronic liver disease: A meta-analysis of diagnostic accuracy. J. Gastroenterol. Hepatol. 29, 1149–1158 (2014).
Caussy, C. et al. Optimal threshold of controlled attenuation parameter with MRI-PDFF as the gold standard for the detection of hepatic steatosis. Hepatology 67(4), 1348–1359 (2018).
Oeda, S. et al. Diagnostic accuracy of FibroScan and factors affecting measurements. Diagnostics (Basel). 10(11), 940 (2020).
Coco, B. et al. Transient elastography: A new surrogate marker of liver fibrosis influenced by major changes of transaminases. J. Viral Hepat. 14, 360–369 (2007).
Arena, U. et al. Acute viral hepatitis increases liver stiffness values measured by transient elastography. Hepatology 47, 380–384 (2007).
Sagir, A. et al. Transient elastography is unreliable for detection of cirrhosis in patients with acute liver damage. Hepatology 47, 592–595 (2007).
Millonig, G. et al. Liver stiffness is directly influenced by central venous pressure. J. Hepatol. 52, 206–210 (2010).
Millonig, G. et al. Extrahepatic cholestasis increases liver stiffness (Fibroscan®) irrespective of fibrosis. Gastroenterology 48(5), 1718–1723 (2008).
Loustaud-Ratti, V. et al. Non-invasive detection of hepatic amyloidosis: FibroScan, a new tool. Amyloid 18, 19–24 (2011).
Jung, K. S. et al. Factors affecting the accuracy of controlled attenuation parameter (CAP) in assessing hepatic steatosis in patients with chronic liver disease. PLoS ONE 9, e98689 (2014).
Kumar, M. et al. Controlled attenuation parameter for non-invasive assessment of hepatic steatosis: Does etiology affect performance?. J. Gastroenterol. Hepatol. 28, 1194–1201 (2013).
Chon, Y. E. et al. Controlled attenuation parameter (CAP) for detection of hepatic steatosis in patients with chronic liver diseases: A prospective study of a native Korean population. Liver Int. 34, 102–109 (2013).
Shen, F. et al. Controlled attenuation parameter for non-invasive assessment of hepatic steatosis in Chinese patients. World J. Gastroenterol. 20, 4702–4711 (2014).
Mi, Y. et al. Controlled attenuation parameter for noninvasive assessment of hepatic steatosis using Fibroscan®: Validation in chronic hepatitis B. Dig. Dis. Sci. 60, 243–251 (2014).
Fujimori, N. et al. Controlled attenuation parameter is correlated with actual hepatic fat content in patients with non-alcoholic fatty liver disease with none-to-mild obesity and liver fibrosis. Hepatol. Res. 46(10), 1019–1027 (2016).
Kumagai, E. et al. Appropriate use of virtual touch quantification and FibroScan M and XL probes according to the skin capsular distance. J. Gastroenterol. 51(5), 496–505 (2016).
Kleiner, D. E. et al. Design and validation of a histological scoring system for nonalcoholic fatty liver disease. Hepatology 41, 1313–1321 (2005).
Brunt, E. M. et al. Nonalcoholic steatohepatitis: a proposal for grading and staging the histological lesions. Am. J. Gastroenterol. 94, 2467–2474 (1999).
Sirli, R., Sporea, I., Bota, S. & Jurchiş, A. Factors influencing reliability of liver stiffness measurements using transient elastography (M-probe)-monocentric experience. Eur. J. Radiol. 82(8), e313–e316 (2013).
Youden, W. J. Index for rating diagnostic tests. Cancer 3(1), 32–35 (1950).
DeLong, E. R., DeLong, D. M. & Clarke-Pearson, D. L. Comparing the areas under two or more correlated receiver operating characteristic curves: a nonparametric approach. Biometrics 44(3), 837–845 (1988).
Cardoso, A. C. et al. The performance of M and XL probes of FibroScan for the diagnosis of steatosis and fibrosis on a Brazilian nonalcoholic fatty liver disease cohort. Eur. J. Gastroenterol. Hepatol. 32(2), 231–238 (2020).
Park, C. C. et al. Magnetic resonance elastography vs transient elastography in detection of fibrosis and noninvasive measurement of steatosis in patients with biopsy-proven nonalcoholic fatty liver disease. Gastroenterology 152(3), 598-607.e2 (2017).
Sogabe, M. et al. Usefulness of differentiating metabolic syndrome into visceral fat type and subcutaneous fat type using ultrasonography in Japanese males. J. Gastroenterol. 47(3), 293–299 (2012).
Mager, D. R. et al. Anthropometric measures of visceral and subcutaneous fat are important in the determination of metabolic dysregulation in boys and girls at risk for nonalcoholic fatty liver disease. Nutr. Clin. Pract. 28(1), 101–111 (2013).
Kure, T. et al. Nonalcoholic fatty liver disease is associated with both subcutaneous and visceral adiposity: A cross-sectional study. Medicine (Baltimore) 98(46), e17879 (2019).
Hegazy, M. A. et al. Abdominal subcutaneous fat thickness and homeostasis model assessment of insulin resistance as simple predictors of nonalcoholic steatohepatitis. Diabetes Metab. Syndr. Obes. 12, 1105–1111 (2019).
Lee, S. H. et al. Abdominal subcutaneous fat thickness measured by ultrasonography correlates with hyperlipidemia and steatohepatitis in obese children. Pediatr. Gastroenterol. Hepatol. Nutr. 18(2), 108–114 (2015).
Tordjman, J. et al. Structural and inflammatory heterogeneity in subcutaneous adipose tissue: Relation with liver histopathology in morbid obesity. J. Hepatol. 56(5), 1152–1158 (2012).
Shen, F. et al. Impact of skin capsular distance on the performance of controlled attenuation parameter in patients with chronic liver disease. Liver Int. 35(11), 2392–2400 (2015).
Mandelia, C. et al. Performance characteristics, intra- and inter-operator agreement of transient elastography in pediatric nonalcoholic fatty liver disease. J. Pediatr. Gastroenterol. Nutr. 72(3), 430–435 (2021).
Acknowledgements
We sincerely thank Ms. Kozue Tashiro, Ms. Maki Miyahara, Ms. Tomomi Yada, and all the medical staff and processing personnel at the research facilities who collected the clinical data. We also thank Angela Morben, DVM, ELS, and H. Nikki March, PhD, from Edanz Group (https://en-author-services.edanz.com/ac), for editing a draft of this manuscript.
Author information
Authors and Affiliations
Contributions
Conception and design: K.T., S.O., K.A., E.S. and H.T.; acquisition of data: S.K., K.T., S.O., K.I., C.I., Y,K., W.Y., M.O., H.I., T.K., T.A., A.K., A.O., M.O., Y.M. and S.A; analysis and interpretation of data: S.K., K.T., S.O. and A.K.; drafting of the manuscript: S.K., K.T., S.O. and H.T.; statistical analysis: S.O. and H.T; study supervision: S.O. and H.T; critical revision of the manuscript; S.K. K.T, S.O. and H.T.; assistance for image interpretation: K.T., S.O. and H.T.
Corresponding author
Ethics declarations
Competing interests
The authors declare no competing interests.
Additional information
Publisher's note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Supplementary Information
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Kimura, S., Tanaka, K., Oeda, S. et al. Effect of skin–capsular distance on controlled attenuation parameter for diagnosing liver steatosis in patients with nonalcoholic fatty liver disease. Sci Rep 11, 15641 (2021). https://doi.org/10.1038/s41598-021-94970-3
Received:
Accepted:
Published: