Table 3 \(\:Inspiratory pressure\) and \(\:{T}_{90}\) measurements for all test scenarios.

From: Bench assessment of PC-CMVs modes in transport and emergency ventilators under ICU conditions

Test

scenario

Lung

ventilator

\(\:Inspiratory\:pressure\)(cmH2O)

\(\:{T}_{90}\)(s)

1

2

3

4

1

2

3

4

A

14.41 ± 0.05HJ

22.51 ± 0.10

29.22 ± 0.08I

35.50 ± 0.34CIJK

0.51 ± 0.01DGK

0.41 ± 0.01

0.42 ± 0.01JK

0.50 ± 0.07

B

14.31 ± 0.03

22.36 ± 0.19I

28.16 ± 0.12EH

32.94 ± 0.05

0.74 ± 0.01

0.79 ± 0.02

0.49 ± 0.01

0.41 ± 0.01 JK

C

13.32 ± 0.03E

21.63 ± 0.17

28.50 ± 0.02H

32.91 ± 0.01A

NR

NR

0.49 ± 0.01

0.59 ± 0.01

D

14.05 ± 0.06

22.26 ± 0.30HI

29.31 ± 0.03IJ

33.91 ± 0.10

0.78 ± 0.01AE

0.74 ± 0.01

0.58 ± 0.01

0.62 ± 0.01E

E

14.70 ± 0.05CHIJ

23.10 ± 0.06HIJK

29.69 ± 0.04BIJ

34.55 ± 0.08IJ

0.55 ± 0.01DG

0.33 ± 0.02

0.42 ± 0.00 JK

0.26 ± 0.01DJK

F

13.94 ± 0.06

21.90 ± 0.11

28.93 ± 0.03

33.82 ± 0.03

0.69 ± 0.01

0.50 ± 0.11

0.38 ± 0.03HJK

0.39 ± 0.03JK

G

13.63 ± 0.02

22.00 ± 0.03I

28.89 ± 0.06

33.48 ± 0.02

0.86 ± 0.01AE

0.86 ± 0.01

0.59 ± 0.01

0.59 ± 0.01

H

11.70 ± 0.03AEK

19.54 ± 0.11BDE

32.68 ± 0.03IBJ

35.01 ± 0.07CIK

NR

NR

0.66 ± 0.01F

0.53 ± 0.01

I

12.06 ± 0.28EK

17.53 ± 0.44BDEG

25.29 ± 0.45ADEH

29.89 ± 0.39AEHJ

NR

NR

NR

NR

J

10.14 ± 0.03AEK

20.67 ± 0.10E

27.67 ± 0.02DEH

32.06 ± 0.03AEH

NR

NR

0.77 ± 0.01AEF

0.89 ± 0.01BEF

K

14.47 ± 0.01HIJ

20.80 ± 0.03E

28.59 ± 0.11

32.37 ± 0.01AH

0.72 ± 0.01A

NR

0.72 ± 0.01AEF

0.78 ± 0.01BEF

  1. Data is presented as mean ± standard deviation (SD). A Differed (ρ<.05) from mechanical ventilator A, B Differed from mechanical ventilator B, C Differed from mechanical ventilator C, D Differed from mechanical ventilator D, E Differed from mechanical ventilator E, F Differed from mechanical ventilator F, G Differed from mechanical ventilator G, H Differed from mechanical ventilator H, I Differed from mechanical ventilator I, J Differed from mechanical ventilator J, K Differed from mechanical ventilator K and NR = no reach the set pressure within set inspiratory time.