Table 5 Display error (%) for \(\:V_{T}\) across all test scenarios. It was estimated as: |(displayed Vt - measured Vt)| x 100%/ measured Vt.
From: Bench assessment of PC-CMVs modes in transport and emergency ventilators under ICU conditions
Display error (%) of Vt | |||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|
Ventilator test scenario | A | B | C | D | E | F | G | H | I | J | K |
1 | 3.214 | - | 5.882 | 11.200 | 20.750 | 1.667 | 5.102 | 8.824 | 9.556 | 17.000 | - |
2 | 1.967 | - | 0.806 | 1.500 | 3.0770 | 4.262 | 7.273 | 6.047 | 13.600 | 39.170 | - |
3 | 0.893 | - | 3.704 | 5.357 | 8.772 | 2.321 | 4.182 | 3.443 | 15.220 | 13.910 | - |
4 | 3.585 | - | 12.500 | 18.000 | 25.490 | 1.154 | 7.000 | 14.900 | 18.700 | 0.488 | - |