Introduction

Patients with cardiac implantable electronic devices (CIEDs) often require magnetic resonance imaging (MRI), with an estimated 50–75% of patients requiring at least one MRI examination during CIED lifetime1,2. Currently, most CIED manufacturers offer MRI-conditional devices certified for scanning under specific conditions. The primary condition is that all components of the CIED system (generator and leads) come from the same manufacturer.

Due to the standardization of lead connectors, the implantation of system components from different manufacturers (“mixed-brand systems”) has become common practice for various reasons3. Current guidelines support performing MRI in patients with mixed-brand systems as an off-label use, based on the evidence of low risk of device failure or damage4,5. It is uncertain whether the evolution of MRI-conditional devices and emerging evidence of MRI safety even with non-MRI-conditional devices have influenced the prevalence of mixed-brand systems in clinical practice.

The aim of this analysis was to assess physicians’ preferences for single-brand or mixed-brand systems in normal clinical practice since 2012 in relation to the availability of MRI-conditional options and new evidence.

Methods

Sample selection

The first author conducted the investigation within the framework of the Italian Home Monitoring Expert Alliance (HMEA) project6. The HMEA project serves as an ongoing independent and permanent data repository from the routine care of patients with CIEDs monitored through the Home Monitoring system (BIOTRONIK SE & Co. KG, Berlin, Germany). Ethics approval for the HMEA project was obtained and all patients provided informed consent for data processing.

The present analysis included patients from the HMEA database who received a de novo conventional pacemaker, implantable cardioverter-defibrillator (ICD), or cardiac resynchronization therapy pacemaker/defibrillator (CRT-P/D). Patients with an unknown lead or generator manufacturer were excluded. Patients with devices requiring specific leads from the same manufacturer for technical reasons, such as single-lead ICDs with atrial sensing capacity (the DX ICD systems) were also excluded. Patient characteristics and electrical parameters of the implanted leads were recorded at the time of device implantation.

Analysis design and statistics

Patients were divided into two groups: a single-brand group (generator and leads from the same manufacturer) and a mixed-brand group (generator from one manufacturer and one or more leads from another manufacturer). Baseline clinical characteristics and electrical parameters of the implanted leads are presented as counts (percentages) or medians (interquartile ranges [IQRs]) and were compared between groups using the χ² test or Wilcoxon rank sum test, as appropriate.

The prevalence of mixed-brand systems was calculated with 95% confidence interval (CI) for the entire population and by device type (pacemaker, ICD, or CRT) and year of implantation. Differences among device types were assessed using the χ² test. Variations over time were tested using generalized estimating equations with mixed-brand implant as the dependent variable and a linear combination of periodic terms over years as the independent terms. A binomial distribution was assumed for the dependent variable.

Additionally, literature was reviewed to identify the most significant publications on MRI safety in patients with non-conditional CIEDs and then evaluate the potential impact of these publications on clinical practice.

Results

A total of 10,831 devices implanted in 81 Italian clinics between 2012 and 2022 had been registered in the HMEA project at the time of database freezing for the present study. After excluding device replacements or upgrades, DX ICD systems, and cases with incomplete device data, 5,853 patients with a de novo CIED were identified. One-quarter of them (25.6%) received a pacemaker, 37.7% a conventional ICD, and 36.6% a CRT device (Fig. 1). Most implants were single-brand systems (95.9%, n = 5,615). Mixed-brand systems (4.1%, n = 238) were implanted in 43 (53.1%) centers. Characteristics of the study cohort for the overall population and by group are compared in Table 1. The mixed-brand group had a significant higher prevalence of female sex, ischemic cardiomyopathy, and comorbidities (Table 1) compared to the single-brand group which may be related to the different prevalence of CRT systems for treating heart failure.

Fig. 1
figure 1

Sample selection. CRT, cardiac resynchronization therapy; CRT-D, CRT defibrillator; CRT-P, CRT pacemaker; DX ICD, single-lead ICD with atrial sensing capability and no atrial pacing capability; ICD, implantable cardioverter-defibrillator; PM, pacemaker.

Table 1 Patient characteristics at implantation.

In patients with 238 mixed-brand implants, a lead from a different manufacturer was connected to atrial (7.1%, n = 17), right ventricular (13.0%, n = 31), or left ventricular (40.3%, n = 96) port of the CIED, or to two or more of these three ports (39.5%, n = 94). A total of 129 mixed-brand leads were connected to the left ventricular port (see Table 2 for lead models).

Table 2 List of lead models not manufactured by Biotronik and connected to the left ventricular port in cardiac resynchronization therapy devices.

Table 3 summarizes the acute electrical parameters for non-mixed and mixed combinations of generator and lead. Statistically significant but clinically neglectable differences between the two groups were found in atrial pacing threshold and impedance, and in left ventricular sensing amplitude and pacing impedance.

Table 3 Electrical parameters for single-brand and mixed-brand combinations of generator-lead.

Mixed-brand prevalence by device type and over time

As illustrated in Fig. 2, the prevalence of mixed-brand CIED systems differed significantly by device type: 4.5% for pacemakers (95% CI, 3.5-5.7%), 1.1% for ICDs (95% CI, 0.7-1.7%), and 6.8% for CRT devices (95% CI, 5.7-7.9%) (p < 0.001). In addition, the prevalence of patients with mixed-brand implants varied significantly over the years (p < 0.001), with a notable decrease from 6.6% before 2015 to 1.3% in 2019, followed by a subsequent increase to 5.1% in 2022 (Fig. 3).

Fig. 2
figure 2

Proportion of patients with a mixed-brand implant by device type (χ² test p < 0.001). CRT, cardiac resynchronization therapy device; ICD, implantable cardioverter-defibrillator; PM, pacemaker.

Fig. 3
figure 3

Proportion of patients with mixed-brand implants over time.

Significant changes over time were observed for pacemaker and CRT mixed-brand implants (Fig. 4). For example, the proportion of mixed-brand pacemaker systems was ≥ 16% until 2015, to drop to 0% in 2019–2020 and increase again to 3.7% in 2022 (p < 0.001 for variation over time, using generalized estimating equations). The proportion of mixed-brand CRT systems ranged from 3.8 to 6.9% in the period 2014–2019 and increased to 9.3-11.9% in 2020–2022 (p = 0.001). The proportion of mixed-brand ICD systems was low during the whole decade, with a maximum of 3.7% before 2014 and a minimum of 0% in 2019–2020 and 2022 (p = 0.438).

Fig. 4
figure 4

Proportion of patients with mixed-brand implants over time by device model. CRT, cardiac resynchronization therapy device; ICD, implantable cardioverter-defibrillator; PM, pacemaker.

Timing of main publications on non-MRI-conditional CIEDs

A non-systematic literature review was conducted to identify main publications on MRI safety in non-conditional CIEDs. These publications, summarized in Table 4, consistently reported a low adverse event rate in recipients of non-MRI-conditional pacemakers and ICDs undergoing MRI. Already in 2017, the Heart Rhythm Society expert consensus paper made a class IIa recommendation for patients with a non-MRI-conditional CIED system to undergo MRI in the absence of fractured, epicardial, or abandoned leads if the MRI examination is the best test for the condition and there is an institutional protocol in which a designated responsible MRI physician and a CIED physician are involved in the procedure7. This recommendation was reinforced in the 2019 American College of Radiology guidance document on MRI safe practices (Class IIa recommendation)8. More recently, the 2021 European Society of Cardiology guidelines extended the endorsement of MRI to pacemaker patients with abandoned transvenous leads when no alternative imaging modality is available and assigned it a Class IIb recommendation4.

Table 4 Main publications on the safety of MRI in non-conditional pacemakers and ICDs.

Discussion

Main findings

This study analyzed 5,853 de-novo pacemaker, ICD, and CRT implantations from a large real-world database ongoing at 81 Italian sites, and revealed that the overall proportion of patients with a mixed-brand system (generator and at least one lead from different manufacturers) was 4.1%. Mixed-brand implants were more prevalent in CRT (6.8%) and pacemaker (4.5%) recipients than in patients with conventional single- or dual-chamber ICDs. The complexity of CRT procedures may lead operators to preferentially use left ventricular lead with which they feel more confident. For example, a left ventricular lead with an active fixation system, produced by a single manufacturer, might be preferred when a conventional lead with passive fixation is perceived to be unstable9. Furthermore, a minority of mixed-brand CRT implants (Table 2), distributed over years, involved a surgically implanted permanent epicardial lead, which may be necessary due to unsuitable coronary sinus anatomy for lead positioning. These systems are still not certified for MRI even in the case of a single-brand implant, but since their use is driven by clinical necessity it should not be influenced by the availability of MRI-conditional options10.

For pacemaker recipients, a higher proportion of mixed-brand implants in certain hospitals may be attributed to previous public tenders in Italy in which brady leads and generators were purchased from different lots.

In contrast, the lower percentage of mixed-brand implants for conventional ICDs can be explained by a simpler procedure than for CRT devices, which typically does not require a change in the lead model. Additionally, operators may have a ‘psychological habit’ of viewing a single-brand system as safer with respect to right ventricular sensing and ICD therapies. This perception could contribute to a preference for a unified brand in the context of ICD implantation.

MRI is often the preferred imaging modality for neurological, musculoskeletal, oncological, and cardiovascular disorders. However, there was a historical contra-indication for CIED recipients. This limitation has been overcome with the development of MRI-conditional devices, with the first adoption in 2010 for pacemakers and 2012 for ICD and CRT devices in Europe. The pivotal requirement for MRI conditionality is that the entire CIED system, encompassing the generator and leads, must come from the same manufacturer. Our study showed that technological advances have had a notable impact on the use of mixed-brand implants over the last decade. Indeed, the rate of mixed-brand systems significantly decreased from 6.6% in 2015 to 1.3% in 2019. This reduction may be due to operators’ reluctance to provide patients with implants without MRI certification. In recent years, the proportion of mixed-brand implants has increased again to 5.1% in 2022. The increase may be explained by recent reports on the safety of MRI with non-MRI-conditional devices. These reports have likely led to less attention being paid to MRI conditionality4,5,7. The growing confidence in the safety of MRI scans using non-MRI-conditional devices may influence the decision-making process and contribute to the observed increase in the rate of mixed-brand implants.

The literature review revealed key publications on MRI safety in patients with non-MRI-conditional CIEDs, with the first larger studies published in 2017. These studies, including the multicenter MagnaSafe Registry11, consistently demonstrated a low adverse event rate in non-MRI-conditional pacemaker and defibrillator recipients undergoing MRI. Recommendations evolved over time, with a class IIa recommendation for this procedure established in the 2017 Heart Rhythm Society expert consensus7. This recommendation was subsequently reinforced by the American College of Radiology in 20198 and the European Society of Cardiology in 20214. Recent studies, specifically investigating mixed-brand CIED recipients undergoing MRI, found no increased risk of adverse events compared to patients with MRI-conditional CIED systems12.

Clinical implications

The growing evidence supporting the safety of MRI in patients with non-conditional CIEDs challenges the historical requirement of single-brand systems for MRI access. The selection of the generator and leads should be made independently and based on clinical characteristics to give the operator flexibility in selecting lead models from different manufacturers. This flexibility, particularly relevant for left ventricular lead positioning, was reflected in our results, which showed no clinically relevant changes in lead parameters between non-mixed and mixed generator-lead combinations. The long-term safety of mixed-brand systems, even after device replacement, has been demonstrated by the recent Detect Long-term Complications After ICD Replacement (DECODE) Registry3, further supporting the interchangeability of the technology. These findings underscore the need for a nuanced approach to CIED component selection that focuses on individual patient characteristics and preferences.

Study limitations

Several limitations should be acknowledged in our analysis. First, the study relies on data from the HMEA project, where CIED generators were obtained from a single manufacturer, which inevitably affects the generalizability of the results. Second, data were collected exclusively from Italian sites, preventing extrapolation to other contexts. Healthcare practices, device preferences, and patient demographics can vary significantly in different regions. Finally, while the study examined trends in implantation, it did not assess long-term clinical outcomes related to mixed-brand implants, such as device complications, patient outcomes, or physician’s choice of device brand at the time of elective replacement. Future research is warranted to address this aspect.

Conclusions

Our study provides insights into the prevalence of mixed-brand CIED systems over the past decade and its relationship to the evolving landscape of MRI accessibility knowledge and options. Mixed-brand systems were more prevalent in patients receiving CRT devices and pacemakers than in patients with conventional ICDs. Following the introduction of MRI-conditional devices, a decline in the prevalence of mixed-brand systems was observed until 2019, with a resurgence in recent years.

The changing paradigm of MRI safety in CIEDs challenges conventional practices and opens opportunities for more patient-centered and adaptable approaches to generator and lead selection.