Abstract
Surgical safety has emerged as a significant public health concern. Ureteral injury (UI) is one of the most common forms of iatrogenic urological issues, lacking non-invasive prevention strategies. In this context, computer-assisted technologies offer a promising solution for enhancing intra-operative safety. This paper presents an in-vivo study focused on evaluating the feasibility of using an augmented reality (AR) surgical navigation system (EVA) for intra-operative ureteral identification on an animal model. An experienced surgeon performed a technical assessment of the system. The clinical evaluation was conducted by four general surgeons tasked with identifying the left or right ureter, both with and without EVA. The technical assessment highlighted that EVA is easily integrable with operating room instruments, achieving a calibration accuracy of \(1.7~\text {cm} \pm 0.002\), and the virtual ureter effectively overlapping with the real ureter. The questionnaires indicated that surgeons appreciated EVA, with a \(SUS_{score} = 73.75\). The perceived mental demand, task complexity, and distractions were lower when using the EVA system. Future work will focus on increasing the number of subjects and exploring the efficacy of the system on other clinical tasks.
Similar content being viewed by others
Introduction
According to the World Health Organization (WHO), unsafe surgical care causes substantial problems, such as major complications, occurring at an alarming rate of \(25\%\) during major surgeries and leading to a mortality rate of 0.5%–5%1. Minimally Invasive Surgery (MIS) and the introduction of robotic systems in the Operating Room (OR) have brought various advancements in surgical techniques in terms of manoeuvrability, precision and enhanced surgical field visualization. Despite that, a series of complications remain closely associated with the surgeon’s experience level.
Specifically, Ureteral Injury (UI) is one of the most common forms of iatrogenic urological injuries, arising in approximately 0.2–1% of pelvic surgeries. Among these cases, 50% are observed during obstetric and gynecological surgeries, with \(30\%\) specifically occurring during radical hysterectomies2,3. UI also occurs in colorectal surgery with an incidence ranging between 0.2 and 1.5%4. Other less common causes of UI include ovarian tumor removal, transabdominal urethropexy, and abdominal vascular surgery. More than \(50\%\) of UI that occurs intra-operatively are not immediately recognised and treated, especially during emergency surgery5. Consequently, these injuries are associated with increased morbidity, mortality, hospital stay, and healthcare costs that can exceed 30,000 US$ per case6. UI also represents a frequent reason for legal disputes with serious psychological, economic and professional burdens for both patients and doctors7; in fact, this complication represents \(17\%\) of non-obstetric lawsuits filed against gynecologists in the USA2.
Currently, there is not a single non-invasive prevention strategy that effectively reduces the risk of UI8. Surgical exposure of the ureter and prophylactic ureteral stenting can help in its identification, but are both invasive approaches, not free from risks and therefore not included in any management protocol8. Nonetheless, in case of high-risk patients, prophylactic placement of ureteral stents can be considered as an important strategy for the prevention of ureteral injury. Pokala et al. conducted a comparison between simultaneous intra-operative vs sequential ureteral stent placement. Their findings revealed that simultaneous stent placement significantly reduced the overall operative time by 19 minutes, without leading to an increase in morbidity9. Studies investigating the identification of ureters using ureter-specific vital dyes are still in the preclinical phase, showing promise for future applications10. Notably, indocyanine green, the most extensively studied vital dye, does not undergo renal metabolism. Consequently, direct instillation into the ureter is required, either via cystoscopy using a rigid cystoscope or by pre-placing a ureteral stent. However, these procedures are invasive and significantly elevate the risk of complications, along with prolonging operative times11.
The use of computer-assisted technologies during intra-operative surgical procedures offers a promising solution for enhancing safety in clinical settings. These technologies aim to improve the surgeon’s awareness and capabilities by providing additional information from pre-operative examinations, monitoring the surgical site, and alerting the surgeon to potential adverse events. Virtual Reality (VR) and Augmented Reality (AR) tools, particularly in laparoscopic assistive technologies, have shown significant potential in achieving these objectives over the past decade12,13,14,15. Atallah et al. demonstrated that stereotaxic surgical navigation in rectal cancer surgery is a valid technique for guiding dissection during surgery16,17. This technique enables the identification of the optimal dissection point using radiological imaging, thereby facilitating the pre-operative localization of anatomical structures before their visualization in the surgical field. However, it is important to note that while radiological anatomy remains static, surgical anatomy inherently exhibits deformable characteristics, which may lead to disparities between the two. On the other hand, Kiyomatsu et al. demonstrated that during minimally invasive surgery in a porcine model, the difference between radiological anatomy and surgical anatomy, especially concerning pelvic vessels, is minimal, being less than 2.1 mm18. These results support the idea that an intra-operative navigation system capable of integrating radiological and endoscopic images could serve as a significant support for surgeons.
This paper presents an in-vivo study focused on evaluating the feasibility of using the surgical navigation system EVA for intra-operative ureteral identification on an animal model. EVA is a system previously developed by the authors to enhance surgical navigation, aimed at improving the quality and safety of MIS19. In fact, acting like an intra-operative GPS for surgery, it allows the visualization of a dynamic VR and/or AR environment that shows 3D models of patient-specific organs in real-time, aligned and synchronized with the endoscopic view. The EVA-VR system was evaluated in a laboratory setting by 14 surgeons, as reported in previous publication19. Results showed a navigation precision lower than 5mm, a reduction in task execution time by \(15\%\) and a reduction in the size of the peritoneal incision (to reach the lesion) by \(46\%\) compared to the control group. Surgeon’s feedback highlighted how the use of EVA system improves the sense of orientation during surgery, therefore also helping in the identification of structures of interest.
Based on the preliminary findings, this study aims to further explore the potential of AR technology as a step forward from the VR GPS technology in surgical navigation. To this end, this paper reports the outcomes of a feasibility study focused on evaluating the EVA-AR system under conditions closely resembling those of real surgery, using a live porcine model during MIS (Fig. 1). The primary objective of this study is to evaluate the efficacy and usability of the system in identifying the ureter during MIS. The secondary endpoint involves assessing the system’s initialization and calibration in an OR setting, measuring both the time required for these processes and the calibration error. Here, the description of the in-vivo study follows the ARRIVE 2.0 guidelines20,21,22.
This paper is structured as follows: Section “Materials and methods” provides a brief overview of the EVA surgical navigation system and describes the study design and protocol. Technical and clinical results are presented in Section “Results”, followed by “Discussion and conclusions” in Section Discussion.
Materials and methods
EVA-AR surgical navigation system
EVA-AR is a surgical navigation system designed to offer an AR view of the surgical field. It provides a dynamic AR environment that shows in real-time 3D models of patient-specific organs overlaid and synchronized with the endoscopic view, showing Structures Of Interest (SOI) identified during the pre-operative planning (i.e., lesions, critical vessels, etc.). This overlay is obtained by manually registering the 3D organ models with the actual ones, using fiducial landmarks selected on rigid structures, such as bones. This process will be referred to as patient calibration. The alignment is then continuously updated during the surgical intervention based on the movements of the endoscope, monitored by an external Optical Tracking System. The idea behind EVA is to make the pre-operative planning easily accessible in the surgical room, relieving the surgeons’ cognitive load required to mentally remember and fuse the plan with the intra-operative endoscopic images. To this end, EVA allows intuitive navigation and guidance, supporting the surgeon’s decision-making process. The system capabilities are enhanced by advanced functionalities, such as changing the transparency of organs to visualize internal structures (e.g., tissue abnormalities, vessel structures, etc.). As shown in Fig. 2, the system is configured as a medical cart that integrates (i) an Optical Tracking Camera to track the endoscope motions, (ii) a full-HD monitor for the visualization of the AR functionalities, and (iii) a computer for the generation of the AR view. The tracking of the endoscope is achieved through a maker tool positioned on the body of the surgical endoscope.
Patient-specific 3D models reconstruction
The EVA system requires patient-specific 3D models as input, which can be obtained from diagnostic images such as CT scans or MRI scans. For this study, the HA3D™ service (Medics Srl, Italy) was used to generate the 3D models using CT scans, as illustrated in Fig. 1. HA3D™ technology integrates engineering modeling knowledge with clinical know-how in order to obtain 3D patient-specific medical-grade models based on 2D images23. HA3D™ reconstructions can be manipulated by the surgeon, generating 3D model views that can be fitted, rotated, zoomed and virtually dissected, allowing a clearer study of the lesion itself. The reconstruction can also reveal possible anomalies near the neoplasm, such as inflammation, lymph nodes or smaller adjacent lesions . This service has already been verified in a pilot study aimed at evaluating the efficacy of high-resolution preoperative 3D reconstructions for lesion localization in oncological colorectal surgery24.
Study protocol
The assessment study of EVA-AR was conducted at the Mini Invasive Advanced Surgery (MIAS) Academy with approval from the Italian Ministry of Health n°654/2023-PR (Risp. a prot. AC827.8). All experiments involving a live animal were performed in accordance with the ARRIVE guidelines and relevant regulations21,22. Experiments involving human participants were conducted in compliance with the Declaration of Helsinki and all applicable ethical standards and regulations. Informed consent was obtained from all participants for study participation and publication of any identifying information/images in an open-access format.
A single male Yorkshire pig weighing between 20 to 30 kg, without genetic modification and in good health status, was used for the study. The animal was housed at the MIAS site and allowed a 12-hour acclimatization period before undergoing an abdominal CT with contrast under general anesthesia. The CT scan included an arterial phase, a late venous phase, and a urographic phase with contrast medium bolus to simultaneously enhance all anatomical structures, including arterial, venous, and urological components.
Before image acquisition, the animal was positioned supine in the same orientation it would assume during surgery, with the goal of minimizing position-induced anatomical deformation. The radiological images were acquired with a slice thickness of 1.5 mm. Subsequently, the animal was returned to the MIAS animal facility and remained there until the day of the experiment. The radiological images were transmitted via the Medics platform and used to generate 3D graphical reconstructions within three working days. The reconstructed structures were the retroperitoneal and pelvic organs, which are minimally affected by position-induced changes and pneumoperitoneum.
The experiment took place in the MIAS operating room, where the animal received general anesthesia administered by a trained veterinarian. The animal was positioned supine and in Trendelenburg position. Three trocars were used at the level of the transverse supra-umbilical line, and the bladder was emptied and anchored to the upper abdominal wall with a fixed trans point. At the conclusion of the study, euthanasia was performed in accordance with approved veterinary protocols. Standard minimally invasive surgery instrumentation, including a laparoscopic column equipped with TIPCAM 1 S 3D LAP, 0° optics (KARL STORZ - ENDOSKOPE, Tuttlingen, Germany), and dissecting and grasping instruments were used.
All anesthetic and euthanasia procedures followed current veterinary best practices. The animal received an intramuscular administration of tiletamine at 4 mg/kg, zolazepam 4 mg/kg (Zoletil), along with dexmedetomidine at 10 mcg/kg (Dexdomitor). Intravenous access was established in both ear veins using 20-gauge catheters. Subsequently, after IV administration of propofol at 4 mg/kg (Propovet) the animal was intubated and mechanically ventilated with a tidal volume of 10–13 mL/kg. General anesthesia was maintained with 2–3% sevoflorane (Sevflo) throughout the procedure. Analgesia was provided by an IV constant rate infusion of fentanyl at 30 mcg/kg/h (Fentadon). At the end of the procedure, the animal was euthanized using the pharmacological procedure of intravenous administration of pentobarbital sodium (150 mg/kg). This method ensures a painless procedure, as it rapidly depresses the central nervous system, leading to loss of consciousness and cessation of respiratory and cardiac functions.
The experimental evaluation was divided in two phases. The first phase was focused on assessing the system’s initialization and calibration in an OR setting, later referred to as Technical evaluation. The second phase was focused on evaluating the efficacy and usability of the system in identifying the ureter during MIS, later referred to as Clinical evaluation.
Technical evaluation
A trained surgeon, with 10 years of experience in MIS and RMIS, and 5 years of experience with animal surgical experimentation, was asked to conduct a technical evaluation of the EVA system. The evaluation consisted in:
OR feasibility Qualitative assessment of the feasibility of integrating EVA’s cart in an operating room;
Configuration assessment Qualitative assessment was performed for the system start-up configuration, to evaluate its proper initialization and ease of use; while a quantitative assessment of the patient calibration accuracy (\(patientCalib_{acc}\)) was performed by computing the re-projection error between the corresponding virtual and real points selected on the models and on the pig, respectively; Specifically, the reprojection error was measured as the Euclidean distance between corresponding virtual points (selected on the preoperative 3D model) and their real counterparts (manually identified on the pig’s anatomy), which were re-projected onto the virtual 3D model using the rigid transformation identified during the calibration process. These points were selected based on anatomical landmarks visible in both the virtual and real environments. The real points were acquired by pointing the tip of a pointer tracked by the tracking system on the skin of the animal model.
Clinical evaluation
Four surgeons participated to the study, including 2 residents and 2 expert surgeons, all male with an age range of 31 to 43 years and an average of 36 years. All of them were from the field of general surgery. Each experimental session included two trials, during which participants performed the same task with and without the support of the EVA-AR system.
To optimize the use of the animal, non-destructive tasks were selected to prevent alterations in its anatomy that could affect the task’s validity for subsequent participants. The chosen task was the identification of the ureter. The anatomy of the pig’s ureter, like all retroperitoneal organs, closely resembles that of humans. However, a notable difference lies in the representation of retroperitoneal adipose tissue, which is minimal in pigs. Consequently, retroperitoneal structures are easily visible through the peritoneum. For this reason, the task was designed to be performed by residents rather than experienced surgeons, who would have found it too easy. This setup allowed for a better assessment of how the EVA system supports less experienced surgeons in identifying key anatomical structures while ensuring validation from an experienced supervisor.
The participants were divided into two groups (Group A and Group B), each comprising an experienced surgeon and a resident. Group A was assigned the task of identifying the left ureter using the EVA-AR system and the right ureter without its assistance, while Group B performed the opposite. The task was considered complete when the participant confidently identified the ureter. For each group, the resident conducted the identification task, while the expert surgeon manipulated the endoscope and confirmed the position of the ureter. The task required the participant to move the intestinal loops and localize the ureter. For each task, the completion time \(t_{task}\) was measured, and the correspondence between the ureter 3D model and the real structure was qualitatively evaluated. In addition, the subjects were asked to fill the following questionnaires:
-
Face validity A custom face validity questionnaire was developed to assess the relevance of the setup, the system and the tasks executed. The specific questions are reported in Fig. 6.
-
System usability scale (SUS) The SUS is a widely used questionnaire-based method for assessing the perceived usability of a system25. It consists of ten statements that participants can evaluate on a Likert scale ranging from “strongly disagree” to “strongly agree”. The statements are designed to cover aspects of usability, efficiency, and learnability. After collecting individual responses, scores are calculated and transformed to provide a global usability score (\(SUS_{score}\)) on a scale from 0 to 100.
-
User experience questionnaire (UEQ) The UEQ is a fast and reliable questionnaire to measure the User Experience of interactive products26. The questionnaire investigates different aspects of the user experience: classical usability aspects, such as efficiency, perspicuity, dependability, and user experience aspects, such as originality and stimulation. The questionnaire uses pairs of contrasting attributes, and participants rate their experience on a scale from 0 to 7.
-
The surgical task load index (Surg-TLX) The Surg-TLX questionnaire is designed to assess the perceived workload experienced by individuals performing surgical tasks. It takes into account various factors contributing to the overall workload, including mental, physical, and temporal demands, as well as factors like task complexity, situational stress and distractions, on a scale from 0 to 100.
While Face Validity, SUS, and UEQ were completed by both residents and surgeons, SurgTLX was filled out only by the residents who directly performed the tasks. Moreover, the surgeons completed the questionnaire only once, after completing both tasks. The results from these questionnaires were analyzed by computing the median and interquartile values, which are represented using boxplots, except for the UEQ results that were analyzed using the standard data analysis tool provided for the questionnaire.
Results and discussion
Technical evaluation results
-
OR feasibility
The dimensions of the EVA system cart are similar to those of a conventional laparoscopic column. Fig. 3 demonstrates its compatibility with the placement of other instruments, indicating its potential for seamless integration into conventional ORs. The cart must be positioned such as to place the Optical Tracking Camera at a distance between 1 m and 2.4 m from the operating field, enabling installation outside the sterile area. During our experiment, we ensured that the tracking camera maintained a clear line of sight to the surgical field by instructing the surgeon to orient it toward the center of the abdomen while keeping the camera as high as possible. This positioning was chosen to optimize visibility and tracking accuracy while minimizing occlusions. No significant tracking occlusions were observed, suggesting that the selected placement was effective in maintaining a continuous and stable tracking signal. From an ergonomic point of view, the monitor for visualizing the AR environment is positioned at eye level when standing. The endoscope marker tool, attached on the laparoscopic endoscope body, is small in size and did not interfere or impede the surgical procedure, as indicated also by the results of the face validy questionnaire. Due to its position, this tool should be made of a sterilizable material.
-
Configuration assessment
The Graphical User Interface (GUI) of the system was considered simple and intuitive. In order to perform the patient calibration it was necessary to choose at least 3 anatomical landmarks that are fixed (i.e. located on the ribcage), easily palpable and recognisable from outside by the surgeon. We chose the pubic symphysis and the medial margin of the last ribs, bilaterally. Regarding the choice of ribcage landmarks, we acknowledge that respiratory motion affects the ribcage. However, we selected the ribcage because it is a relatively rigid structure that maintains a stable spatial relationship with the preoperative 3D model and remains consistent throughout the intervention. While the ribcage moves due to breathing, this motion was minimal compared to other anatomical regions in the pig model. In human procedures, iliac crests would be the preferred landmarks for patient calibration due to their stability and independence from respiratory motion. However, in pigs, this was not feasible due to anatomical differences, making the ribcage the most viable alternative for calibration. Fig. 4 shows the process of landmarks selection. The calibration process took on average 1.04 min. A \(patientCalib_{acc}\) of \(1.7~\text{cm} \pm 0.002\) was measured. The sources for this calibration error can be attributed partially to the pneumoperitoneum-induced changes and to the landmarks detection error. In fact, the virtual landmarks chosen on the 3D models did not take into account the skin-fat-muscle layers covering the bones.
Clinical evaluation results
The average task execution time \(t_{task}\) for all the tasks conducted with and without the support of the EVA-AR system were respectively \(41.41s \pm 11.11s\) and \(81.41s \pm 62.10s\); Table 1 shows raw values for task execution time for all the trials.
The surgeon had access to both the traditional laparoscopic column screen and EVA’s screen. The EVA screen displayed AR images, where the 3D models were continuously updated in real time to align with the endoscope view. The AR views were presented without noticeable delays compared to the endoscopic images and without any inconsistencies. Following the initial patient calibration, the 3D models demonstrated a high level of congruence with the anatomical structures of the pig. In both tasks conducted with the support of EVA, manual adjustment using the mouse was necessary at the beginning of the trial to refine the registration between the 3D models and the target structures. The reference points used for alignment included the kidneys and some features of the visible pelvic anatomy. In both cases, surgeons confirmed that the projected 3D model of the ureter accurately corresponded to the real anatomical structure, as shown in Fig. 5.
Figure 6 reports the boxplot of the results from the Face Validity questionnaire, indicating the median values and interquartile range for each question, with responses separated between surgeons (in blue) and residents (in orange). An analysis of the results is provided below:
-
1.
The navigation system cart’s disturbance to the intra-operative setup received a relatively low score, indicating that it may not significantly disrupt the surgical environment;
-
2.
Both residents and surgeons reported that the markers placed on the endoscope did not interfere with intra-operative practice;
-
3.
Participants felt that the task performed was not too simple compared to a real surgical operation, as indicated by the relatively high score.
-
4.
Participants considered the system to be useful in MIS, in particular for the surgeons, while performing complex surgeries.
-
5.
The system’s impact on task execution time compared to traditional methods received a moderately high score, suggesting that it may increase the time required for certain tasks; This opinion however is in contrast with the measure execution time, which appears to be lower when using the EVA system. We hypothesize this refers to the time to configure the system together with its usage during the surgery.
-
6.
During the experimental session, participants’ attention was divided between the traditional laparoscopy screen and EVA’s augmented reality screen. A difference in preference was observed between residents and surgeons, the latter focusing more on the AR display than on the endoscopic video.
While the Face Validity questionnaire provided valuable qualitative feedback on the execution task, system usability and intraoperative integration of the EVA system, we acknowledge the potential for social desirability bias in custom-made questionnaire. This limitation is particularly relevant given the small number of participants and the exploratory nature of the study. To mitigate this, the questionnaire was designed to include both positive and critical evaluation aspects. Additionally, future studies will incorporate standardized usability assessments with a larger and more diverse participant pool to further validate these findings.
Results of the SUS questionnaire are presented in Fig. 7, where boxplots are reported with responses separated between surgeons and residents. The overall score obtained with the SUS questionnaire is \(SUS_{score} = 73.75\), with very similar results from residents (72.5) and surgeons (75). This indicates a general positive perception of the system’s usability, as a SUS score above 68 is considered above the average. Focusing on the single questions, the subjects expressed a high likelihood of frequent system usage and found the system easy to use. While there were mixed feelings regarding the system’s complexity, subjects generally perceived good integration of system functions and low levels of inconsistency. The majority of respondents felt confident in using the system and expressed confidence in its learnability, expecting most people to learn to use it quickly (more for surgeons). While there was some indication of a moderate need for learning before becoming proficient with the system, overall, the findings suggest a favorable perception of the system’s usability with room for improvement in addressing perceived complexity and learning needs.
The UEQ questionnaire results, shown in Fig. 8, indicate that the EVA-AR system was rated highly across several user experience attributes. The highest mean score was for Attractiveness (\(2.250 \pm 0,68\) for both residents and surgeons), suggesting a positive overall impression of the system. Perspicuity and Efficiency were evaluated quite differently from residents and surgeons, with generally positive scores from the residents (\(2.25 \pm 0.5\) and \(2.135 \pm 0.28\) respectively) and lower scores from the surgeons (\(0.875 \pm 0.28\) and \(0.625 \pm 0.28\) respectively). This suggests that, in general, participants found the system relatively easy to learn and use, although there is room for improvement in these areas. Dependability and Stimulation also received favorable evaluations both groups, with residents providing slightly higher scores (\(2.0 \pm 2.0\) and \(2.215 \pm 1.53\) respectively) than surgeons (\(1.38 \pm 1.53\) and \(1.38 \pm 0.03\) respectively), highlighting the system’s ability to engage and interest users. Finally, Novelty was the only metric evaluated more positively by surgeons (\(2.25 \pm 0.0\)) than residents (\(1.5 \pm 0.0\)), but it still remained positive overall.
The results obtained from the Surg-TLX questionnaire are presented in Fig. 9. These results reflect feedback exclusively from the residents who directly performed the task. Due to the small sample size, no statistical analysis was performed. However, the box plots shows an increased median and interquartile range in particular for the Mental Demand, Task Complexity and Distractions for the task executed without the support of EVA. These findings suggest that EVA may alleviate the challenges associated with surgical navigation and identification of critical structures.
Conclusion
This paper presented an in-vivo study on an animal model performed to explore the feasibility of the EVA-AR surgical navigation system for uretral identification. Based on an optical tracking system, EVA aligns 3D reconstructions of patient-specific organs, generated from preoperative radiological images, with intra-operative images captured by the endoscope. EVA-AR continuously updates the view of the 3D model in real-time, enabling a consistent and congruent alignment between virtual anatomical structures and those observed by the surgeon.
Being aware of the challenges associated with using an AR navigation system in general surgery, particularly due to the deformable nature of tissue organs and their susceptibility to positional changes with respect to the preoperative planning, this study was focused on retroperitoneal and pelvic organs. Despite being soft, these organs exhibit relative stability in position, with less susceptibility to modification by factors such as patient positioning and pneumoperitoneum-induced stimuli. This decision was influenced by findings from studies conducted on a porcine model by Kiyomatsu18.
This study demonstrated how EVA-AR can enable intra-operative surgical navigation. The EVA cart integrated well into the operating room and did not hinder the use of other equipment. The system was fully calibrated and operational in less than 15 minutes, requiring the involvement of two individuals, one at the cart and one at the operating table. According to questionnaire results, the participants who used EVA-AR believe that the system is functional and efficient, easy to learn, and user-friendly. The participants observed both the endoscopic video on the traditional laparoscopic display and on EVA’s AR display during task execution, defining EVA as a supportive tool. This research focused on the problem of ureter identification, which is a crucial aspect in numerous surgical procedures. In this context, the support of EVA-AR navigation system appears to have positively influenced the task, facilitating the localization of the actual ureter.
However, further research is still necessary to validate this initial finding. Limitations of this study include: (i) The limited number of subjects who tested EVA; (ii) The fact that experiments were performed on a single animal; (iii) The exclusion of more invasive actions during the ureter identification trials; (iv) The absence of retroperitoneal fat in pigs, which facilitates ureter identification through the parietal peritoneum. Further studies are already planned to address the shortcomings identified in this feasibility study before experimenting with EVA in the operating room on humans.
Data availibility
The data that support the findings of this study are available from the corresponding author on reasonable request.
References
Safety, W.P., Organization, W.H. et al. WHO guidelines for safe surgery 2009: safe surgery saves lives. WHO/IER/PSP/2008.08-1E (World Health Organization, 2009).
Jha, S., Coomarasamy, A. & Chan, K. K. Ureteric injury in obstetric and gynaecological surgery. Obstetr. Gynaecol. 6, 203–208 (2004).
Kumari Rani, A. Urological injury in obstetric and gynecological surgery: A retrospective analysis of five years. Indian J. Obstet. Gynecol. 89, 18–22 (2020).
Croghan, S. M. et al. The sentinel stent? A systematic review of the role of prophylactic ureteric stenting prior to colorectal resections. Int. J. Colorectal Dis. 34, 1161–1178 (2019).
Burks, F. N. & Santucci, R. A. Management of iatrogenic ureteral injury. Ther. Adv. Urol. 6, 115–124 (2014).
Halabi, W. J. et al. Ureteral injuries in colorectal surgery: An analysis of trends, outcomes, and risk factors over a 10-year period in the united states. Dis. Colon Rectum 57, 179–186 (2014).
Jha, S. Ureteric injury: Always a guilty verdict?. BJOG Int. J. Obstet. Gynaecol. 122, 499–499 (2015).
Brollo, P. P. et al. Preventing iatrogenic ureteral injury in colorectal surgery: A comprehensive and systematic review of the last 2 decades of literature and future perspectives. Surg. Today 54, 1–19 (2023).
Pokala, N. et al. A randomized controlled trial comparing simultaneous intra-operative vs sequential prophylactic ureteric catheter insertion in re-operative and complicated colorectal surgery. Int. J. Colorectal Dis. 22, 683–687 (2007).
Lau, L. W., Luciano, M., Schnermann, M. & Cha, J. Ureter identification in an inflammatory pig model using a novel near-infrared fluorescent dye. Lasers Surg. Med. 52, 537–542 (2020).
White, L. A. et al. Intraureteral indocyanine green augments ureteral identification and avoidance during complex robotic-assisted colorectal surgery. Colorectal Dis. 23, 718–723 (2021).
Bernhardt, S., Nicolau, S. A., Soler, L. & Doignon, C. The status of augmented reality in laparoscopic surgery as of 2016. Med. Image Anal. 37, 66–90 (2017).
Penza, V. et al. Envisors: Enhanced vision system for robotic surgery. A user-defined safety volume tracking to minimize the risk of intraoperative bleeding. Front. Robot. AI 4, 15 (2017).
Penza, V., Moccia, S., De Momi, E. & Mattos, L. S. Enhanced vision to improve safety in robotic surgery. In Handbook of Robotic and Image-Guided Surgery, 223–237 (Elsevier, 2020).
Amparore, D. et al. Indocyanine green drives computer vision based 3d augmented reality robot assisted partial nephrectomy: The beginning of “automatic’’ overlapping era. Urology 164, e312–e316 (2022).
Atallah, S. et al. Robotic-assisted stereotactic real-time navigation: Initial clinical experience and feasibility for rectal cancer surgery. Tech. Coloproctol. 23, 53–63 (2019).
Atallah, S., Larach, S. W. & Monson, J. R. Stereotactic navigation for TAMIS-TME. Minim. Invas. Ther. Allied Technol. 25, 271–277 (2016).
Kiyomatsu, H. et al. Deformation of the pelvic arteries caused by pneumoperitoneum and postural changes in an animal model. In vivo 35, 275–281 (2021).
Penza, V. et al. The GPS for surgery: A user-centered evaluation of a navigation system for laparoscopic surgery. Int. J. Med. Robot. Comput. Assist. Surg. 16, 1–13 (2020).
Percie du Sert, N. et al. The arrive guidelines 2.0. Updated guidelines for reporting animal research. J. Cereb. Blood Flow Metab. 40, 1769–1777 (2020).
Kilkenny, C., Browne, W. J., Cuthill, I. C., Emerson, M. & Altman, D. G. Improving bioscience research reporting: The arrive guidelines for reporting animal research. J. Pharmacol. Pharmacother. 1, 94–99 (2010).
Kilkenny, C., Browne, W., Cuthill, I. C., Emerson, M. & Altman, D. G. Animal research: Reporting in vivo experiments: The arrive guidelines. Br. J. Pharmacol. 160, 1577 (2010).
Medics 3D. Medics 3d official website (2024). Accessed: 2024-08-12.
Soriero, D. et al. Efficacy of high-resolution preoperative 3d reconstructions for lesion localization in oncological colorectal surgery-first pilot study. Healthcare 10, 900 (2022).
Brooke, J. et al. SUS-A quick and dirty usability scale. Usab. Evaluat. Ind. 189, 4–7 (1996).
Laugwitz, B., Held, T. & Schrepp, M. Construction and evaluation of a user experience questionnaire. In HCI and Usability for Education and Work: 4th Symposium of the Workgroup Human-Computer Interaction and Usability Engineering of the Austrian Computer Society, USAB 2008, Graz, Austria, November 20-21, 2008. Proceedings 4, 63–76 (Springer, 2008).
Acknowledgements
This work was funded by the 2022-2024 Triennial Current Research Program of IRCCS San Martino Hospital under the following project: Project Code N705A, Line 3, ’Study of an Intraoperative Navigation System to Improve Safety in Minimally Invasive Oncological Surgery.’ Additionally, this work was funded by the European Union - NextGenerationEU and carried out within the framework of the project ’RAISE - Robotics and AI for Socio-economic Empowerment’.
Author information
Authors and Affiliations
Contributions
V.P., L.S.M., A.N. and J.O. conceived the experiment, D.S., B.S., D.P., L.E., G.C. and M.A. conducted the experiment, D.S. provided the technical assessment, V.P. analyzed the results, V.P., A.N. and L.S.M. wrote the manuscript. All authors reviewed the manuscript.
Corresponding author
Ethics declarations
Competing interest
V.P., D.S., B.S., A.N., J.O., D.P., L.E., F.C., M.A., S.S, D.G.C., and L.S.M. have no conflicts of interest or financial ties to disclose.
Informed consent
Informed consent for publication of identifying images in this online open-access publication has been obtained from all the subjects depicted.
Additional information
Publisher’s note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International License, which permits any non-commercial use, sharing, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if you modified the licensed material. You do not have permission under this licence to share adapted material derived from this article or parts of it. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by-nc-nd/4.0/.
About this article
Cite this article
Penza, V., Soriero, D., Sperotto, B. et al. Evaluating the EVA surgical navigation system for ureteral identification in an in vivo porcine model. Sci Rep 15, 16976 (2025). https://doi.org/10.1038/s41598-025-00138-8
Received:
Accepted:
Published:
DOI: https://doi.org/10.1038/s41598-025-00138-8