Table 3 Main TAVI access-related complications of the overall population and by groups.

From: Crossover balloon occlusion technique for access closure during transcatheter aortic valve implantation: a multicenter observational study

 

Overall

(%)

CBOT

(%)

ProGlide

(%)

P

229

(100%)

135

(59%)

94

(41%)

Need for surgical closure

2

(0.9)

0

(0.0)

2

(2.1)

0.089

Perforation

8

(3.5)

2

(1.5)

6

(6.4)

0.047

Femoral artery occlusion

2

(0.9)

1

(0.7)

1

(1.1)

0.796

Pseudo-aneurysm

1

(0.4)

1

(0.7)

0

(0.0)

0.403

Retroperitoneal hematoma

1

(0.4)

0

(0.0)

1

(1.1)

0.230

Minor vasc. complications

11

(4.8)

3

(2.2)

8

(8.5)

0.029

Major vasc. complications

2

(0.9)

0

(0.0)

2

(2.1)

0.089