Abstract
This study presents the design and theoretical analysis of a pseudo parity-time symmetric photonic crystal biosensor for early oral cancer detection. Using the transfer matrix method, we numerically investigated the resonant transmission properties of a 1D photonic crystal structure with a central defect layer (D = 50 μm) containing healthy and cancer oral cells. The proposed \(\:{\left(Si{O}_{2}*Si\right)}^{N}\left(oral\:cells\right){\left(Si{O}_{2}*Si\right)}^{N}*substrate\) configuration exhibits exceptional sensitivity, quantified by two key metrics: a spectral sensitivity of 118 GHz/RIU and an extraordinary transmittance sensitivity of \(\:1.9\times\:{10}^{8}\:\%/RIU.\) These results demonstrate significant improvements over conventional photonic biosensors, enabled by the PT-symmetric design’s ability to operate near an exceptional point. The optimized structure achieves ultra-sharp resonances (\(\:9.1\times\:{10}^{-5}\) THz FWHM) with remarkable field enhancement, making it particularly suitable for detecting minute refractive index changes associated with malignant transformations. Our findings establish a foundation for developing high-performance, label-free diagnostic tools for early-stage oral cancer detection.
Similar content being viewed by others
Introduction
Photonic crystals (PhCs) are periodic dielectric structures that manipulate the propagation of electromagnetic waves through photonic bandgap effects (PBG), analogous to the way electronic bandgaps control electron flow in semiconductors1,2,3. These periodic variations in the refractive index (RI) allow for selective transmission or reflection of light at specific wavelengths, depending on the structural parameters and material composition of the crystal4,5. One-dimensional PhCs (1D PhCs), in particular, are widely used due to their ease of fabrication and tunable spectral properties6. Their layered design enables strong localization of light, which is highly desirable for applications in filtering, sensing, and lasing7,8,9,10. Introduction of a defect layer within a 1D PhC breaks the periodicity, creating localized defect modes within the PBG that are extremely sensitive to the optical properties of the defect medium11.
Recent advances in non-Hermitian photonics, particularly in systems exhibiting parity-time (PT) symmetry, have opened new paradigms in light manipulation. PT-symmetric systems are characterized by a complex RI distribution where the real part is even and the imaginary part is odd with respect to spatial inversion, i.e., \(\:{n}_{R}\left(x\right)={n}_{R}\left(-x\right)\) and \(\:{n}_{I}\left(x\right)=-{n}_{I}\left(-x\right)\)12,13. Such systems can support real eigenvalue spectra under certain conditions, despite being non-Hermitian, and can undergo phase transitions at so-called exceptional points where eigenvalues and eigenvectors coalesce14. In PhCs, introducing PT symmetry via spatially balanced gain and loss regions can lead to unique phenomena such as unidirectional invisibility, asymmetric transmission, and resonance enhancement15. These effects are particularly advantageous in sensing applications, as operation near exceptional points leads to drastic amplification of response to small perturbations16.
Recent developments in terahertz sensing have highlighted the potential of spin-controlled light-matter interactions, particularly through the photonic spin Hall effect (PSHE), which enables high-sensitivity and multitasking detection via spin-dependent field manipulation. Innovative designs, such as PSHE-based Janus sensors and non-reciprocal systems, have demonstrated exceptional performance by exploiting spin-orbit coupling and asymmetric light propagation17,18. Additionally, PT-symmetric structures have been shown to enhance PSHE responses through chirality modulation, further linking non-Hermitian concepts with spin-selective phenomena19. While these approaches rely on spin-momentum locking rather than non-Hermitian localization, they share a common goal of achieving ultra-sensitive detection through tailored wave control.
Photonic biosensors have emerged as promising platforms for label-free, real-time detection of biomolecular interactions20,21,22,23. Unlike traditional biochemical assays, photonic sensors detect changes in the RI of a sample region, enabling the identification of cellular or molecular transformations without chemical tagging24. In the context of early cancer diagnosis, the ability to detect minute changes in the RI of biological tissues is crucial, as early-stage malignant cells often exhibit slightly altered optical properties compared to healthy cells25,26. PhC-based biosensors, particularly those employing defect modes, are well suited to this task due to their sharp spectral features and high-quality factors. When coupled with PT-symmetric designs, these biosensors gain further enhancements in sensitivity and resolution.
This study presents a novel PT-symmetric 1D PhC biosensor designed for the early detection of oral cancer. The proposed structure, composed of alternating layers of silicon dioxide (\(\:Si{O}_{2}\)) and silicon (\(\:Si\)) surrounding a central defect layer containing biological samples, is engineered to operate near an exceptional point to exploit resonance enhancement effects. Through rigorous numerical analysis based on the transfer matrix method (TMM), we demonstrate that the biosensor achieves a spectral sensitivity of 118 GHz/RIU and an unprecedented transmittance sensitivity of\(\:1.9\times\:{10}^{8}\:\%/RIU.\) The structure exhibits ultra-narrow resonant linewidths (FWHM ≈ \(\:9.1\times\:{10}^{-5}\) THz), which allow for the detection of subtle RI shifts associated with the transition from healthy to malignant oral cells. Compared to conventional PhC biosensors, our PT-symmetric configuration offers superior sensitivity and specificity, laying the groundwork for advanced diagnostic platforms capable of early-stage cancer detection. The structure exhibits PT-like symmetry, with spatial parity symmetry and localized loss at the defect center, leading to resonance behaviors reminiscent of PT-symmetric systems. While gain and loss are not strictly balanced, the system supports non-Hermitian resonant modes characteristic of PT-symmetric or pseudo-PT regimes.
Equations and theory
The proposed sensor consists of a 1D PhC with a central defect layer designed to interact with oral cancer cells in the terahertz (THz) range, as clear in Fig. 1. The structure is fabricated on \(\:Si{O}_{2}\) substrate and follows a periodic arrangement of dielectric layers of \(\:Si{O}_{2}\) and \(\:Si\) (A and B) with alternating refractive indices (\(\:{n}_{A}\) and \(\:{n}_{B}\)) and thicknesses (\(\:{d}_{A}\) and \(\:{d}_{B}\)). The defect layer (\(\:{d}_{0}\)), embedded in the PhC, acts as a resonant cavity where the sample (oral cancer cells) is loaded. The choice of \(\:Si{O}_{2}\) and \(\:Si\) for the PhC structure is driven by their well-documented advantages in THz photonics. Si provides a high refractive index (~ 3.4) with low absorption loss in the THz regime, enabling strong light confinement essential for photonic bandgap formation27. Its compatibility with standard semiconductor fabrication techniques ensures cost-effective, large-scale production28. \(\:Si{O}_{2}\) serves as the low-index material (~ 1.9–2.1), creating the necessary refractive index contrast to widen the photonic bandgap while maintaining minimal THz absorption. Additionally, \(\:Si{O}_{2}\) offers excellent biocompatibility, preventing unwanted chemical interactions with biological samples29. The \(\:\text{S}\text{i}/Si{O}_{2}\) platform also facilitates precise layer thickness control through established deposition methods like PECVD and thermal oxidation, which is critical for optimizing defect-mode resonance30. Together, these materials provide an ideal balance of optical performance, fabrication scalability, and biosensing compatibility for the proposed PT-symmetric THz sensor.
The real part of RI \(\:n\left(f\right)\) and the absorption \(\:\alpha\:\left(f\right)\) of the normal and cancer oral samples at room temperature (20 ˚C) from 0.20 THz to 0.91 THz can be represented by fitting the experimental data as follows25:
For normal oral samples:
For cancered oral samples:
The TMM provides an effective framework for analyzing light propagation through PhC structures31,32. This computational approach breaks down the system into individual layers, with each layer represented by a transfer matrix that describes how electromagnetic waves propagate through it. For a 1D-PhC composed of alternating dielectric materials, the method calculates the cumulative effect of multiple layers by multiplying their respective transfer matrices. This allows researchers to determine the overall transmission and reflection properties of the complete PhC structure.
When applying TMM to PhCs, each layer’s transfer matrix incorporates key optical parameters, including refractive index, thickness, and propagation constants. The method accounts for both forward and backward propagating waves at each interface between different materials. By systematically multiplying these matrices for an entire periodic structure, researchers can obtain the total transfer matrix that characterizes the PhC’s optical behavior. This approach enables precise calculation of transmission spectra, including the identification of photonic band gaps where light propagation is forbidden.
TMM offers several significant advantages for PhC analysis and design. First, it provides an exact solution to Maxwell’s equations for stratified media, making it more accurate than approximate methods. Second, the method is computationally efficient, allowing rapid evaluation of different structural configurations. Third, TMM can easily incorporate complex material properties, including absorption and gain, which is particularly important for active photonic devices. These features make TMM an indispensable tool for optimizing PhC structures for specific applications such as optical filters, sensors, and waveguides33,34,35,36,37.
The total matrix of \(\:{\left(A*B\right)}^{N}\left(D\right){\left(A*B\right)}^{N}*substrate\) as oral cancer detector is:
where \(\:N\) is the number of (\(\:{a}_{A}{a}_{B})\:\)periods. The complete system behavior can be described by a composite transfer matrix \(\:A\) with components \(\:{A}_{11}\), \(\:{A}_{12}\), \(\:{A}_{21}\), and \(\:{A}_{22}\). Each individual layer within the PhC structure is represented by its own distinct transfer matrix, which captures its specific optical properties and propagation characteristics. These layer matrices incorporate the material’s refractive index (\(\:{n}_{i})\), thickness (\(\:{d}_{i})\), angle for each layer (\(\:{\theta\:}_{i}\)), and interaction with electromagnetic waves at the relevant frequencies as follows:
For TE polarized waves\(\:,\)
The total transmittance (T) is given by:
where \(\:{\varnothing\:}_{0}\) and \(\:{\varnothing\:}_{s}\) are for ambient and substrate mediums, and t is:
The experimental measurements demonstrated strong correlation with theoretical simulations performed using the TMM38,39. The PBG positions observed in practical testing aligned closely with calculated predictions. This close agreement between measured and simulated results confirms the reliability of TMM for modeling the optical behavior of PhC structures in the terahertz regime.
Results and discussions
Figure 2(a-b) represent the absorption and real RI of normal and cancered oral samples at room temperatures. At room temperature, water absorption in the THz range becomes more pronounced, potentially obscuring subtle differences between tissue types. The frequency-dependent dispersion of the refractive index would still follow a decreasing trend with increasing frequency, but with broader peaks due to enhanced molecular relaxation processes. Despite these challenges, the persistent contrast in refractive index between normal and cancerous tissues suggests that THz imaging could remain viable for ex vivo diagnostics at room temperature. The RI and absorption of normal oral samples has higher values than the cancered cells.
Experimental25 and fitted (a) real RI and (b) absorption of normal and cancered oral samples at room temperatures.
In the following, the transmittance of \(\:{\left(A*B\right)}^{N}{\left(A*B\right)}^{N}*substrate\) will be studied. The \(\:{n}_{A}\), \(\:{n}_{B}\), \(\:{d}_{A},\) and \(\:{d}_{B}\) are 2.1, 3.4, 65 \(\:\mu\:m\), and 35 \(\:\mu\:m\), respectively. Due to contrasting dielectric properties and periodic RI modulation between \(\:Si{O}_{2}\) and \(\:Si\), a PBG with central frequency of 0.58 THz, as clear in Fig. 3. \(\:Si\) has a high refractive index (~ 3.4 in the THz range), while \(\:Si{O}_{2}\) exhibits a lower index (~ 2.1), creating strong dielectric contrast. When arranged in a periodic lattice, this contrast leads to Bragg scattering, where certain THz frequencies are completely reflected due to destructive interference while others propagate. The PBG appears when the lattice periodicity matches the THz wavelength, preventing electromagnetic wave propagation in specific frequency ranges. The PBG’s position and width depend on the lattice geometry, filling ratio of \(\:\text{S}\text{i}/Si{O}_{2}\), and incident wave polarization. The PBG can be tuned by adjusting structural parameters. Additionally, the low THz absorption of \(\:Si{O}_{2}\) helps maintain sharp band edges, while silicon’s high index enhances light confinement, making \(\:\text{S}\text{i}/Si{O}_{2}\) PhCs efficient for THz filtering, sensing, and slow-light devices. The value of the central frequency of the PBG is compatible with the theoretical equation40:
where \(\:{c}_{0}\) is the speed of light in vacuum.
When a defect composed of normal oral cells with a thickness of 50 \(\:\mu\:m\) is introduced into a \(\:\text{S}\text{i}/Si{O}_{2}\) PhC, a resonant peak of \(\:{f}_{R}=\)0.61 THz, T= 5.3%, and bandwidth (FWHM) of 0.037 THz emerges within the PBG due to localized mode confinement. The PBG normally prohibits electromagnetic wave propagation at specific THz frequencies; however, the defect disrupts the periodic dielectric structure, creating a localized state that allows certain frequencies to resonate. Normal oral cells have distinct dielectric properties (refractive index and absorption) compared to the surrounding PhC lattice, forming a cavity-like region where THz waves become trapped. This defect-induced cavity supports standing waves at a particular resonant frequency, leading to a transmission peak within the forbidden band gap.
The position and intensity of the resonant peak depend on the defect’s thickness, RI contrast, and geometric placement within the PhC. Since normal oral cells exhibit different THz optical properties than \(\:\text{S}\text{i}/Si{O}_{2}\), their integration modifies the local electromagnetic field distribution, enhancing light-matter interaction at the defect site. This principle is exploited in biosensing applications, where the resonant frequency change can indicate variations in cell composition or pathology. The defect-mediated resonance thus enables precise detection of tissue abnormalities, leveraging the high sensitivity of PhCs to minute dielectric changes. Replacing the normal oral cells with cancered cells changes the resonant peak to \(\:{f}_{R}=\)0.61 THz, \(\:T=\:11.1\:\%,\) and FWHM of 0.0268 THz. As the RI contrast between cancered and normal cells is very small, the peak position doesn’t change. The transmittance of cancered peak cells is higher than the normal cell’s peak because the absorption of normal cells is higher than that of cancered cells, as clear in Fig. 2b.
To enhance the \(\:T\) and FWHM of peaks, \(\:Si{O}_{2}\) layers will be doped with quantum dots to get gain layers in the first PhC and loss layers in the second PhC41,42, as clear in Fig. 4. The proposed photonic structure incorporates an optically pumped gain layer to enhance its performance characteristics. When externally pumped with an optical source, the quantum dots embedded within the gain layer absorb photons at specific resonant frequencies. Through the process of stimulated emission, these excited quantum dots subsequently re-emit photons at the same characteristic frequencies. This coherent emission process effectively amplifies the propagating electromagnetic waves within the structure, leading to significant enhancement of the optical signal.
The amplification mechanism relies fundamentally on the precise coupling between the system’s inherent resonant peaks and the frequencies re-emitted by the quantum dots. This interaction creates constructive interference that magnifies the signal intensity at the desired operational frequencies. The gain layer’s ability to compensate for optical losses while simultaneously amplifying the signal stems from its carefully engineered complex refractive index, expressed as \(\:{n}_{gain}={n}_{R}-i\:0.01\:Q\), where \(\:{n}_{R}\) represents the real component, and \(\:Q\) denotes the gain factor. This formulation allows for controlled light amplification within the photonic structure.
Complementing the gain layer, the loss layer features a corresponding refractive index profile given by \(\:{n}_{loss}={n}_{R}+i\:0.01\:Q\). The balanced opposition between these gain and loss components creates a stable system where the imaginary parts of the refractive indices (\(\:\pm\:0.01Q\)) precisely govern the amplification and attenuation characteristics. This parity between gain and loss mechanisms enables exceptional control over light propagation dynamics while maintaining system stability.
The strategic implementation of quantum dot gain media combined with refractive index engineering offers several operational advantages. The stimulated emission from quantum dots provides frequency-selective amplification that can be precisely tuned through optical pumping parameters. Meanwhile, the complex refractive index modulation allows for dynamic control over light-matter interactions within the structure. Together, these features enable the development of advanced photonic devices with enhanced signal processing capabilities, improved signal-to-noise ratios, and tunable spectral response characteristics, making this approach particularly valuable for applications requiring controlled optical amplification and manipulation.
Transmittance of proposed oral cancer detector \(\:{\left(A*B\right)}^{N}\left(D\right){\left(A*B\right)}^{N}*substrate\) with normal cells at room temperature, N = 4, initial \(\:\theta\:\) of zerro degree, and \(\:{d}_{0}\) of 50 \(\:\mu\:m\) for (a) \(\:Q=1\), (b) \(\:Q=2\), (c) \(\:Q=3\), (d) \(\:Q=4\), (e) \(\:Q=5\), (f) \(\:Q=6\), (g) \(\:Q=7\), (h) \(\:Q=8\), (i) \(\:Q=9\), (j) \(\:Q=10\), (k) \(\:Q=11\), (l) \(\:Q=12\), (m) \(\:Q=13\), (n) \(\:Q=14\), and (o) \(\:Q=15\).
The transmittance spectra in Fig. 5(a-o) demonstrate the optical behavior of the proposed PhC structure \(\:{\left(A*B\right)}^{N}\left(D\right){\left(A*B\right)}^{N}*substrate\) under varying gain/loss conditions (\(\:Q\)). All simulations assumed normal incidence (\(\:\theta\:\:=\:0^\circ\:\)) at room temperature conditions. The spectra reveal a well-defined resonant peak centered near 0.61 THz that exhibits remarkable sensitivity to the loss parameter \(\:\kappa\:\). At low loss (\(\:\kappa\:\:=\:0.01-0.03,\:Q=1-3\)), the resonance appears broad with transmittance below 10%, indicating strong absorption. In moderate gain/loss conditions (\(\:\kappa\:\:=\:0.04-0.09,\) \(\:Q=4-9\)), the peak narrows significantly while achieving 40-100% transmittance, suggesting optimal balance between gain/loss factor and light localization. For higher gain/loss factor (\(\:\kappa\:\:=\:0.10-0.15,\:Q=10-15\)), the resonance becomes extremely sharp with transmittance amplification exceeding 2000%, demonstrating the structure’s ability to enhance specific frequencies despite increased absorption. The highest transmittance and lowest FWHM are achieved for healthy cells at \(\:Q=14\), as clear in Fig. 6.
By replacing the healthy cells with oral cancer sample, the spectra reveal a well-defined resonant peak nearly at the same position (0.61 THz) and exhibits remarkable sensitivity to the gain/loss parameter κ, as clear in Fig. 7(a-o). At low gain/loss factor (\(\:\kappa\:\:=\:0.01-0.06,\:Q=1-5\)), the resonance appears broad with transmittance below 100%, indicating strong absorption. In moderate gain/loss conditions (\(\:\kappa\:\:=\:0.06-0.08,\:Q=6-8\)), the peak narrows significantly while achieving 100-600% transmittance. For gain/loss factor (\(\:\kappa\:\:=\:0.09-0.10,\:Q=9-10\)), the resonance becomes extremely sharp with transmittance amplification exceeding 5000%, demonstrating the structure’s ability to enhance specific frequencies despite increased absorption. The highest transmittance and lowest FWHM are achieved for healthy cells at \(\:Q=9.6\), as clear in Fig. 8.
Transmittance of proposed oral cancer detector \(\:{\left(A*B\right)}^{N}\left(D\right){\left(A*B\right)}^{N}*substrate\) with cancer cells at room temperature, \(\:N=4\), initial \(\:\theta\:\) of zerro degree, and \(\:{d}_{0}\) of 50 \(\:\mu\:m\) for (a) \(\:Q=1\), (b) \(\:Q=2\), (c) \(\:Q=3\), (d) \(\:Q=4\), (e) \(\:Q=5\), (f) \(\:Q=6\), (g) \(\:Q=7\), (h) \(\:Q=8\), (i) \(\:Q=9\), (j) \(\:Q=10\), (k) \(\:Q=11\), (l) \(\:Q=12\), (m) \(\:Q=13\), (n) \(\:Q=14\), and (o) \(\:Q=15\).
The extraordinary performance of our PhC biosensor stems from its carefully designed parity-time (PT) symmetric configuration. This non-Hermitian system achieves perfect balance between gain and loss regions, creating unique optical properties not found in conventional structures. The periodic arrangement of layers with properly tuned imaginary refractive index components (\(\:\kappa\:\:=\:0.01Q\)) establishes the necessary PT-symmetric condition. This balanced gain-loss distribution enables the system to operate near an exceptional point, where eigenvalues and eigenvectors coalesce, producing the ultra-sharp resonant peaks observed in our simulations.
The PT-symmetric design transforms optical loss into a performance-enhancing feature through several interconnected mechanisms. First, the system’s operation near an exceptional point creates extreme sensitivity to perturbations while maintaining stable resonance conditions. This explains why increasing the loss parameter (\(\:\kappa\:\)) actually improves the resonant peak quality - a counterintuitive result impossible in ordinary photonic structures. Second, the non-Hermitian nature of the system enables unusual energy localization effects, where the electric field becomes strongly enhanced precisely at the defect layer containing biological samples. This field enhancement directly boosts the sensor’s sensitivity to minute changes in cell properties.
For oral cancer detection, the PT-symmetric design offers three crucial benefits. The exceptional point physics provides: (1) orders-of-magnitude better sensitivity to small refractive index changes between healthy and cancerous cells (Fig. 9), (2) inherent noise rejection by suppressing all frequencies except the sharp resonance, and (3) stability against fabrication imperfections. As our results show, these advantages become particularly pronounced at higher \(\:Q\) values (\(\:Q=9-14\)), where the system operates in the optimal PT-broken phase. This creates the ideal conditions for medical diagnostics - extremely narrow resonances for precise detection combined with strong transmission for reliable signal acquisition.
The proposed PT-symmetric PhC sensor demonstrates exceptional sensitivity to refractive index variations, as quantified by two complementary metrics. The spectral sensitivity \(\:S\left(f\right)=\frac{\varDelta\:{f}_{R}}{\varDelta\:{n}_{cells}}=118\:\text{G}\text{H}\text{z}/\text{R}\text{I}\text{U}\) characterizes the resonant peak shift per refractive index unit (RIU) change, indicating that minute variations in cellular dielectric properties produce measurable frequency displacements. This level of sensitivity enables detection of subtle biomolecular changes associated with early-stage oral cancer development.
The transmittance-based sensitivity \(\:S\left(T\right)=\frac{\varDelta\:T}{\varDelta\:{n}_{cells}}=1.9\times\:{10}^{8}\:\%/RIU\) reveals even more remarkable performance, representing an extraordinary amplification of optical response to refractive index perturbations. This exceptionally high value stems from the PT-symmetric design’s ability to convert small index variations into dramatic transmittance changes through its exceptional point-enhanced resonance. The dual-sensitivity mechanism provides robust detection capability - spectral shifts offer absolute measurements while transmittance changes provide amplified signals for low-concentration biomarkers.
These sensitivity values substantially surpass conventional photonic biosensors, particularly in the critical terahertz frequency range. The spectral sensitivity of 118 GHz/RIU represents a high improvement over typical PhC sensors, while the transmittance sensitivity of \(\:1.9\times\:{10}^{8}\:\%/RIU\) demonstrates unprecedented response amplification. This performance advantage originates from the PT-symmetric design’s unique ability to transform small refractive index perturbations into strongly amplified optical signals through its carefully balanced gain-loss configuration and exceptional point dynamics.
Table 1 presents a systematic performance comparison between our proposed PT-symmetric PhC sensor and previously reported biosensing devices. Our sensor demonstrates significant improvements across all key metrics when benchmarked against conventional approaches. The spectral sensitivity of 118 GHz/RIU represents a 2-fold enhancement over the best-performing PhC biosensor reported in literature50. More remarkably, the transmittance sensitivity of \(\:1.9\times\:{10}^{8}\:\%/RIU\) surpasses existing optical sensors, establishing a new benchmark for label-free detection. In addition, the quality factor (\(\:Q-factor=\frac{{f}_{R}}{FWHM}\)) and limit of detection (\(\:LoD=\frac{{f}_{R}}{20\:S\:Q-factor}\)) recorded high performance.
The comparison reveals two distinct advantages of our PT-symmetric design: superior sensitivity-to-footprint ratio enabled by exceptional point physics, and inherent noise rejection capability through the PT-symmetric filtering mechanism. These performance metrics position our sensor as a potentially transformative technology for early cancer diagnosis, particularly where conventional methods struggle with limited sensitivity or require complex labeling procedures.
While this study primarily correlates detection signals with refractive index (RI) changes, we recognize that real-world biological samples (e.g., cells and tissues) exhibit additional complexities such as variations in morphology, density, and heterogeneous biomolecule composition. These factors collectively contribute to the local RI distribution and can imprint distinct signatures on the resonance spectrum (e.g., peak broadening, asymmetric shifts, or secondary modes). Critically, our platform’s high sensitivity to RI gradients inherently captures these composite effects, as they manifest in the measured optical response. For practical applications, calibration protocols using control samples or complementary techniques (e.g., machine learning-aided deconvolution) could disentangle these intertwined contributions. Future work will systematically quantify such multifactorial interactions to enhance the sensor’s specificity in complex media.
While the proposed theoretical design demonstrates promising performance, practical implementation requires consideration of fabrication challenges and tolerances. The sensor structure, composed of Si, SiO₂-doped layers, may encounter deviations in layer thicknesses or refractive indices during experimental manufacturing due to deposition control, etching precision, or material inhomogeneity. Such variations can shift the resonant peak position and alter the optimal Q parameter. However, the underlying physical mechanism remains intact, enabling post-fabrication optimization through recalibration.
Conclusion
The theoretical investigation of our PT-symmetric PhC biosensor reveals exceptional performance characteristics for oral cancer detection. Through systematic TMM simulations, we demonstrated that the proposed structure achieves: (1) superior spectral sensitivity (118 GHz/RIU) through exceptional point-enhanced resonance, (2) unprecedented transmittance amplification (\(\:1.9\times\:{10}^{8}\:\%/RIU\)) via non-Hermitian field localization, and (3) effective noise rejection through PT-symmetric filtering. These capabilities stem from the careful balance of gain and loss in the PhC configuration, which transforms conventional limitations into performance-enhancing features. The sensor’s ability to resolve minute refractive index changes while maintaining strong signal transmission suggests strong potential for clinical applications where early detection is critical. This study advances the field of photonic biosensing by demonstrating how PT symmetry can overcome fundamental sensitivity limitations in conventional optical sensors. The structure demonstrates PT-like characteristics through spatial parity symmetry and a centrally localized loss layer, resulting in resonance behavior similar to that of PT-symmetric systems. Despite the absence of exact gain-loss balance, the system supports non-Hermitian resonant modes typical of PT-symmetric or quasi-PT regimes.
Data availability
Requests for materials or code should be addressed to Zaky A. Zaky.
References
Daher, M. G. et al. Design of a nano-sensor for cancer cell detection based on a ternary photonic crystal with high sensitivity and low detection limit. Chin. J. Phys. 77, 1168–1181. https://doi.org/10.1016/j.cjph.2022.03.032 (2022).
Zaky, Z. A., Al-Dossari, M., Hendy, A. S., Zayed, M. & Aly, A. H. Gamma radiation detector using cantor quasi-periodic photonic crystal based on porous silicon doped with polymer. Int. J. Mod. Phys. B. 38, 2450409. https://doi.org/10.1142/S0217979224504095 (2024).
Zaky, Z. A. et al. Theoretical study of doped porous silicon in cantor quasi periodic structure for gamma radiation detection. Sci. Rep. 15, 14995. https://doi.org/10.1038/s41598-025-94555-4 (2025).
Joannopoulos, J., Johnson, S., Winn, J. & Meade, R. Photonic Crystals: Molding the Flow of Light Princeton Univ Ed (University of, 2008).
Sakoda, K. & Sakoda, K. Optical Properties of Photonic Crystalsvol. 2 (Springer, 2005).
Zaky, Z. A. et al. Theoretical optimization of Tamm plasmon polariton structure for pressure sensing applications. Opt. Quant. Electron. 55, 738. https://doi.org/10.1007/s11082-023-05023-0 (2023).
Zaky, Z. A. et al. Theoretical analysis of porous silicon one-dimensional photonic crystal doped with magnetized cold plasma for hazardous gases sensing applications. Opt. Quant. Electron. 55, 584. https://doi.org/10.1007/s11082-023-04907-5 (2023).
Daher, M. G. & Trabelsi, Y. A tunable broadband-binary photonic crystal detector for the detection of organic chemical compounds. Phys. B: Condens. Matter. 707, 417158. https://doi.org/10.1016/j.physb.2025.417158 (2025).
Zaky, Z. A., Al-Dossari, M., Hendy, A. S. & Aly, A. H. Studying the impact of interface roughness on a layered photonic crystal as a sensor. Phys. Scr. 98, 105527. https://doi.org/10.1088/1402-4896/acfa4a (2023).
Zaky, Z. A., Alzahrani, A., Khedr, G. H., Elsharkawy, M. & Sallah, M. Radiation detector based on coupling between defect mode and topological edge state mode in photonic crystal. Sci. Rep. 15, 1–14. https://doi.org/10.1038/s41598-025-99332-x (2025).
Maldovan, M. & Thomas, E. L. Simultaneous localization of photons and phonons in two-dimensional periodic structures. Appl. Phys. Lett. 88 https://doi.org/10.1063/1.2216885 (2006).
Bender, C. M. & Boettcher, S. Real spectra in non-Hermitian hamiltonians having P T symmetry. Phys. Rev. Lett. 80, 5243. https://doi.org/10.1103/PhysRevLett.80.5243 (1998).
Rüter, C. E. et al. Observation of parity–time symmetry in optics. Nat. Phys. 6, 192–195. https://doi.org/10.1038/nphys1515 (2010).
El-Ganainy, R. et al. Non-Hermitian physics and PT symmetry. Nat. Phys. 14, 11–19. https://doi.org/10.1038/nphys4323 (2018).
Feng, L. et al. Experimental demonstration of a unidirectional reflectionless parity-time metamaterial at optical frequencies. Nat. Mater. 12, 108–113. https://doi.org/10.1038/nmat3495 (2013).
Hodaei, H. et al. Enhanced sensitivity at higher-order exceptional points. Nature 548, 187–191. https://doi.org/10.1038/nature23280 (2017).
Sui, J. Y., Zou, J. H., Liao, S. Y., Li, B. X. & Zhang, H. F. High sensitivity multiscale and multitasking Terahertz Janus sensor based on photonic spin hall effect. Appl. Phys. Lett. 122 https://doi.org/10.1063/5.0153342 (2023).
Sui, J. Y., Liao, S., Li, B. & Zhang, H. F. High sensitivity multitasking non-reciprocity sensor using the photonic spin hall effect. Opt. Lett. 47, 6065–6068. https://doi.org/10.1364/OL.476048 (2022).
Liang, C., Liu, D., Liu, R., Deng, D. & Wang, G. Chirality-modulated photonic spin Hall effect in PT-symmetry, Nanophotonics, vol. 11, pp. 3475–3484, (2022). https://doi.org/10.1515/nanoph-2022-0229
Daher, M. G. et al. ,., Detection of basal cancer cells using photodetector based on a novel surface plasmon resonance nanostructure employing perovskite layer with an ultra high sensitivity, Plasmonics, vol. 17, pp. 2365–2373. https://doi.org/10.1007/s11468-022-01727-3 (2022).
Daher, M. G., Taya, S. A., Colak, I. & Ramahi, O. M. Design of a novel optical sensor for the detection of waterborne bacteria based on a photonic crystal with an ultra-high sensitivity. Opt. Quant. Electron. 54, 108. https://doi.org/10.1007/s11082-021-03486-7 (2022).
Taya, S. A., Daher, M. G., Colak, I. & Ramahi, O. M. Highly sensitive nano-sensor based on a binary photonic crystal for the detection of Mycobacterium tuberculosis bacteria. J. Mater. Sci.: Mater. Electron. 32, 28406–28416. https://doi.org/10.1007/s10854-021-07220-7 (2021).
Daher, M. G. et al. Design of a novel detector based on photonic crystal nanostructure for ultra-high performance detection of cells with diabetes. Opt. Quant. Electron. 54, 701. https://doi.org/10.1007/s11082-022-04093-w (2022).
Fan, X. & White, I. M. Optofluidic microsystems for chemical and biological analysis. Nat. Photonics. 5, 591–597. https://doi.org/10.1038/nphoton.2011.206 (2011).
Sim, Y. C., Park, J. Y., Ahn, K. M., Park, C. & Son, J. H. Terahertz imaging of excised oral cancer at frozen temperature. Biomedical Opt. Express. 4, 1413–1421. https://doi.org/10.1364/BOE.4.001413 (2013).
Calado, G. Development of methodologies for Raman spectral analysis of human saliva for detection of oral cancer, 2020. https://doi.org/10.21427/6hty-gf29
Jepsen, P. U., Cooke, D. G. & Koch, M. Terahertz spectroscopy and imaging–Modern techniques and applications. Laser Photonics Rev. 5, 124–166. https://doi.org/10.1002/lpor.201000011 (2011).
Tonouchi, M. Cutting-edge Terahertz technology. Nat. Photonics. 1, 97–105. https://doi.org/10.1038/nphoton.2007.3 (2007).
He, Q., Zhang, Z., Gao, F., Li, Y. & Shi, J. In vivo biodistribution and urinary excretion of mesoporous silica nanoparticles: effects of particle size and PEGylation, small, vol. 7, pp. 271–280, (2011). https://doi.org/10.1002/smll.201001459
Sze, S. M., Li, Y. & Ng, K. K. Physics of Semiconductor Devices (Wiley, 2021).
Zaky, Z. A., Sallah, M., Zhaketov, V. & Aly, A. H. Topological edge state resonance as gamma dosimeter using Poly nanocomposite in symmetrical periodic structure. Sci. Rep. 15, 1–12 (2025).
Sayed, F. A. et al. Quasi periodic photonic crystal as gamma detector using Poly nanocomposite and porous silicon. Sci. Rep. 15, 1–17 (2025).
Yeh, P. Optical Waves in Layered Media (Wiley New York, 1988).
Zaky, Z. A., Mohaseb, M., Hendy, A. S. & Aly, A. H. Design of phononic crystal using open resonators as harmful gases sensor. Sci. Rep. 13, 9346. https://doi.org/10.1038/s41598-023-36216-y (2023).
Zaky, Z. A. et al. Photonic crystal with magnified resonant peak for biosensing applications. Phys. Scr. 98, 055108. https://doi.org/10.1088/1402-4896/accbf1 (2023).
Zaky, Z. A., Alamri, S., Zohny, E. I. & Aly, A. H. Simulation study of gas sensor using periodic phononic crystal tubes to detect hazardous greenhouse gases. Sci. Rep. 12, 21553. https://doi.org/10.1038/s41598-022-26079-0 (2022).
Zaky, Z. A., Panda, A., Pukhrambam, P. D. & Aly, A. H. The impact of magnetized cold plasma and its various properties in sensing applications. Sci. Rep. 12, 3754. https://doi.org/10.1038/s41598-022-07461-4 (2022).
Wang, Z. et al. 1D partially oxidized porous silicon photonic crystal reflector for mid-infrared application. J. Phys. D. 40, 4482. https://doi.org/10.1088/0022-3727/40/15/016 (2007).
Zhang, B. et al. ,., Bandwidth tunable optical bandpass filter based on parity-time symmetry, Micromachines, vol. 13, p. 89. https://doi.org/10.3390/mi13010089 (2022).
Lova, P., Manfredi, G. & Comoretto, D. Advances in functional solution processed planar 1D photonic crystals. Adv. Opt. Mater. 6, 1800730. https://doi.org/10.1002/adom.201800730 (2018).
Zaky, Z. A., Alamri, S., Zhaketov, V. & Aly, A. H. Refractive index sensor with magnified resonant signal. Sci. Rep. 12, 13777. https://doi.org/10.1038/s41598-022-17676-0 (2022).
Fang, Y., Zhang, Y. & Wang, J. J. Resonance-dependent extraordinary reflection and transmission in PC-symmetric layered structure. Opt. Commun. 407, 255–261. https://doi.org/10.1016/j.optcom.2017.09.049 (2018).
Ge, X. & He, S. Experimental realization of an open cavity. Sci. Rep. 4, 1–5. https://doi.org/10.1038/srep05965 (2014).
Andueza, Á., Pérez-Conde, J. & Sevilla, J. Differential refractive index sensor based on photonic molecules and defect cavities. Opt. Express. 24, 18807–18816. https://doi.org/10.1364/OE.24.018807 (2016).
Panghal, A. et al. Terahertz Chemical Sensor based on the Plasmonic Hexagonal Microstructured Holes Array in Aluminum, in 44th International Conference on Infrared, Millimeter, and Terahertz Waves (IRMMW-THz), 2019, pp. 1–2., 2019, pp. 1–2. (2019). https://doi.org/10.1109/IRMMW-THz.2019.8874063
Sun, B., Yu, Y. & Yang, W. Enhanced toroidal localized spoof surface plasmons in homolateral double-split ring resonators. Opt. Express. 28, 16605–16615. https://doi.org/10.1364/OE.395068 (2020).
Aly, A. H. et al. Photonic Crystal Enhanced by Metamaterial for Measuring Electric Permittivity in GHz Range, Photonics, vol. 8, p. 416, (2021). https://doi.org/10.3390/photonics8100416
Qin, P. et al. Angle-Insensitive toroidal metasurface for High-Efficiency sensing. IEEE Trans. Microwave Theory Tech. 69, 1511–1517. https://doi.org/10.1109/TMTT.2020.3027016 (2020).
Zaky, Z. A., Amer, H. A., Suthar, B. & Aly, A. H. Gas sensing applications using magnetized cold plasma multilayers. Opt. Quant. Electron. 54, 217. https://doi.org/10.1007/s11082-022-03594-y (2022).
Zaky, Z. A., Al-Dossari, M., Zhaketov, V. & Aly, A. H. Defected photonic crystal as propylene glycol THz sensor using parity-time symmetry. Sci. Rep. 14, 23209. https://doi.org/10.1038/s41598-024-73477-7 (2024).
Funding
This work was supported and funded by the Deanship of Scientific Research at Imam Mohammad Ibn Saud Islamic University (IMSIU) (grant number IMSIU-DDRSP2501).
Author information
Authors and Affiliations
Contributions
Z.A. Zaky invented the original idea of the study, implemented the computer code, performed the numerical simulations, analyzed the data, wrote and revised the main manuscript text, and was the team leader. A. Hennache analyzed the data and discussed the results. V. D. Zhaketov analyzed the data and discussed the results. Finally, all Authors developed the final manuscript.
Corresponding author
Ethics declarations
Competing interests
The authors declare no competing interests.
Ethics declarations
This article does not contain any studies involving animals or human participants performed by any authors.
Additional information
Publisher’s note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International License, which permits any non-commercial use, sharing, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if you modified the licensed material. You do not have permission under this licence to share adapted material derived from this article or parts of it. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by-nc-nd/4.0/.